Literature DB >> 17213746

Bipolar radiofrequency ablation using internally cooled electrodes in ex vivo bovine liver: prediction of coagulation volume from applied energy.

Stephan Clasen1, Diethard Schmidt, Klaus Dietz, Andreas Boss, Stefan M Kröber, Christina Schraml, Jan Fritz, Claus D Claussen, Philippe L Pereira.   

Abstract

OBJECTIVE: We sought to evaluate the relationship between parameters of bipolar radiofrequency (RF) ablation using internally cooled electrodes.
MATERIALS AND METHODS: Bipolar RF ablations (n = 24) were performed in ex vivo bovine liver using an internally cooled applicator with 2 electrodes located on the same shaft. The power-output was systematically varied (20-75 W). On the basis of our experimental data, mathematical functions were fitted and the goodness-of-fit was assessed by the parameter R.
RESULTS: The duration to induce an increase of tissue resistance and the amount of applied energy increased with a decreased power-output. The maximum short-axis was 4.5 cm (20 W) and required an application of 64 kilojoules (kJ). The volume of coagulation can be determined as a function of the duration of energy application (R = 0.954) and the amount of applied energy (R = 0.945).
CONCLUSION: The amount of applied energy and the duration of energy application can predict the volume of induced coagulation and may be useful to control internally cooled bipolar RF ablation.

Entities:  

Mesh:

Year:  2007        PMID: 17213746     DOI: 10.1097/01.rli.0000248973.95949.eb

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  3 in total

1.  Clinical effectiveness of bipolar radiofrequency ablation for small liver cancers.

Authors:  Yukio Osaki; Kenji Ikeda; Namiki Izumi; Satoyoshi Yamashita; Hiromitsu Kumada; Shinji Hatta; Kiwamu Okita
Journal:  J Gastroenterol       Date:  2012-10-10       Impact factor: 7.527

2.  Image-guided tumor ablation: emerging technologies and future directions.

Authors:  Justin P McWilliams; Edward W Lee; Shota Yamamoto; Christopher T Loh; Stephen T Kee
Journal:  Semin Intervent Radiol       Date:  2010-09       Impact factor: 1.513

3.  Moving-shot versus fixed electrode techniques for radiofrequency ablation: comparison in an ex-vivo bovine liver tissue model.

Authors:  Eun Ju Ha; Jung Hwan Baek; Jeong Hyun Lee
Journal:  Korean J Radiol       Date:  2014-11-07       Impact factor: 3.500

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.