PURPOSE: New-generation stented bioprostheses coupled with better imaging modalities are expanding the clinical utility of transcatheter aortic valve replacement (TAVR). This study aimed at evaluating the feasibility of real-time cardiovascular magnetic resonance (rtCMR) -guided TAVR using the Medtronic Engager aortic valve system in a preclinical model. DESCRIPTION: The Engager delivery device was slightly modified to make it CMR-compatible. Ten Yucatan swine underwent rtCMR-guided transapical TAVR. Postplacement phase-contrast and first-pass perfusion CMR sequences were used to evaluate for aortic regurgitation and myocardial perfusion, respectively. EVALUATION: Real-time CMR provided excellent visualization of cardiac anatomy during TAVR. Nine of 10 animals had proper valve placement in the aortic annulus as determined by CMR and confirmed by necropsy inspection. Postplacement phase-contrast scans confirmed no intravalvular or paravalvular leaks. Perfusion scans demonstrated sufficient coronary flow. Roentgenographs confirmed proper placement of the prostheses.
CONCLUSIONS: The Engager valve can be implanted transapically under rtCMR guidance with a modified, CMR-compatible delivery device in a preclinical model. Cardiovascular magnetic resonance allowed for accurate preplacement evaluation, real-time guidance, and postplacement functional assessment.
Copyright © 2014 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2014        PMID: 25468087      PMCID: PMC6524139          DOI: 10.1016/j.athoracsur.2014.09.017

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  2 in total

1.  Real-time magnetic resonance imaging-guided transcatheter aortic valve replacement.

Authors:  Justin G Miller; Ming Li; Dumitru Mazilu; Tim Hunt; Keith A Horvath
Journal:  J Thorac Cardiovasc Surg       Date:  2015-11-22       Impact factor: 5.209

2.  Robot-assisted real-time magnetic resonance image-guided transcatheter aortic valve replacement.

Authors:  Justin G Miller; Ming Li; Dumitru Mazilu; Tim Hunt; Keith A Horvath
Journal:  J Thorac Cardiovasc Surg       Date:  2015-12-10       Impact factor: 5.209

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.