Real-time magnetic resonance imaging-guided transcatheter aortic valve replacement.

Justin G Miller1, Ming Li1, Dumitru Mazilu1, Tim Hunt1, Keith A Horvath2.   

Abstract

OBJECTIVES: To demonstrate the feasibility of Real-time magnetic resonance imaging (rtMRI) guided transcatheter aortic valve replacement (TAVR) with an active guidewire and an MRI compatible valve delivery catheter system in a swine model.
METHODS: The CoreValve system was minimally modified to be MRI-compatible by replacing the stainless steel components with fluoroplastic resin and high-density polyethylene components. Eight swine weighing 60-90 kg underwent rtMRI-guided TAVR with an active guidewire through a left subclavian approach.
RESULTS: Two imaging planes (long-axis view and short-axis view) were used simultaneously for real-time imaging during implantation. Successful deployment was performed without rapid ventricular pacing or cardiopulmonary bypass. Postdeployment images were acquired to evaluate the final valve position in addition to valvular and cardiac function.
CONCLUSIONS: Our results show that the CoreValve can be easily and effectively deployed through a left subclavian approach using rtMRI guidance, a minimally modified valve delivery catheter system, and an active guidewire. This method allows superior visualization before deployment, thereby allowing placement of the valve with pinpoint accuracy. rtMRI has the added benefit of the ability to perform immediate postprocedural functional assessment, while eliminating the morbidity associated with radiation exposure, rapid ventricular pacing, contrast media renal toxicity, and a more invasive procedure. Use of a commercially available device brings this rtMRI-guided approach closer to clinical reality.
Copyright © 2016 The American Association for Thoracic Surgery. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Medtronic CoreValve; real-time magnetic resonance image; swine animal model; transcatheter aortic valve replacement

Mesh:

Year:  2015        PMID: 26725711      PMCID: PMC4834228          DOI: 10.1016/j.jtcvs.2015.11.024

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  50 in total

Review 1.  Timing of aortic valve surgery.

Authors:  C M Otto
Journal:  Heart       Date:  2000-08       Impact factor: 5.994

2.  Editor's page: is CMR safe?

Authors:  G M Pohost
Journal:  J Cardiovasc Magn Reson       Date:  2001       Impact factor: 5.364

3.  CoreValve® transcatheter self-expandable aortic bioprosthesis.

Authors:  Giuseppe Bruschi; Federico De Marco; Luigi Martinelli; Silvio Klugmann
Journal:  Expert Rev Med Devices       Date:  2013-01       Impact factor: 3.166

4.  Acute renal failure after coronary intervention: incidence, risk factors, and relationship to mortality.

Authors:  P A McCullough; R Wolyn; L L Rocher; R N Levin; W W O'Neill
Journal:  Am J Med       Date:  1997-11       Impact factor: 4.965

5.  Transapical sutureless aortic valve implantation under magnetic resonance imaging guidance: Acute and short-term results.

Authors:  Keith A Horvath; Dumitru Mazilu; Junfeng Cai; Bogdan Kindzelski; Ming Li
Journal:  J Thorac Cardiovasc Surg       Date:  2014-11-01       Impact factor: 5.209

6.  Real-time magnetic resonance-guided aortic valve replacement using Engager valve.

Authors:  Bogdan A Kindzelski; Ming Li; Dumitru Mazilu; Timothy Hunt; Keith A Horvath
Journal:  Ann Thorac Surg       Date:  2014-12-01       Impact factor: 4.330

7.  Magnetic resonance fluoroscopy allows targeted delivery of mesenchymal stem cells to infarct borders in Swine.

Authors:  Alexander J Dick; Michael A Guttman; Venkatesh K Raman; Dana C Peters; Breno S S Pessanha; Jonathan M Hill; Scott Smith; Greig Scott; Elliot R McVeigh; Robert J Lederman
Journal:  Circulation       Date:  2003-12-01       Impact factor: 29.690

8.  Active two-channel 0.035'' guidewire for interventional cardiovascular MRI.

Authors:  Ozgur Kocaturk; Ann H Kim; Christina E Saikus; Michael A Guttman; Anthony Z Faranesh; Cengizhan Ozturk; Robert J Lederman
Journal:  J Magn Reson Imaging       Date:  2009-08       Impact factor: 4.813

9.  Transcatheter aortic valve implantation without balloon predilation: a single-center pilot experience.

Authors:  Oscar A Mendiz; Hugo Fraguas; Gustavo A Lev; Leon R Valdivieso; Roberto R Favaloro
Journal:  Catheter Cardiovasc Interv       Date:  2013-04-18       Impact factor: 2.692

Review 10.  Transmyocardial revascularization devices: technology update.

Authors:  Bogdan A Kindzelski; Yifu Zhou; Keith A Horvath
Journal:  Med Devices (Auckl)       Date:  2014-12-18
View more
  3 in total

1.  Improving the cardiac cath-lab interventional imaging eco-system.

Authors:  Benjamin R Ciske; Michael A Speidel; Amish N Raval
Journal:  Transl Pediatr       Date:  2018-01

2.  Improved passive catheter tracking with positive contrast for CMR-guided cardiac catheterization using partial saturation (pSAT).

Authors:  Mari Nieves Velasco Forte; Kuberan Pushparajah; Tobias Schaeffter; Israel Valverde Perez; Kawal Rhode; Bram Ruijsink; Mazen Alhrishy; Nicholas Byrne; Amedeo Chiribiri; Tevfik Ismail; Tarique Hussain; Reza Razavi; Sébastien Roujol
Journal:  J Cardiovasc Magn Reson       Date:  2017-08-15       Impact factor: 5.364

3.  Interventional cardiac magnetic resonance imaging: current applications, technology readiness level, and future perspectives.

Authors:  Sophie C Rier; Suzan Vreemann; Wouter H Nijhof; Vincent J H M van Driel; Ivo A C van der Bilt
Journal:  Ther Adv Cardiovasc Dis       Date:  2022 Jan-Dec
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.