Applicability of self-activation of an NaI scintillator for measurement of photo-neutrons around a high-energy X-ray radiotherapy machine.

Genichiro Wakabayashi1, Akihiro Nohtomi, Eriko Yahiro, Toshioh Fujibuchi, Junichi Fukunaga, Yoshiyuki Umezu, Yasuhiko Nakamura, Katsumasa Nakamura, Makoto Hosono, Tetsuo Itoh.   

Abstract

The applicability of the activation of an NaI scintillator for neutron monitoring at a clinical linac was investigated experimentally. Thermal neutron fluence rates are derived by measurement of the I-128 activity generated in an NaI scintillator irradiated by neutrons; β-rays from I-128 are detected efficiently by the NaI scintillator. In order to verify the validity of this method for neutron measurement, we irradiated an NaI scintillator at a research reactor, and the neutron fluence rate was estimated. The method was then applied to neutron measurement at a 10-MV linac (Varian Clinac 21EX), and the neutron fluence rate was estimated at the isocenter and at 30 cm from the isocenter. When the scintillator was irradiated directly by high-energy X-rays, the production of I-126 was observed due to photo-nuclear reactions, in addition to the generation of I-128 and Na-24. From the results obtained by these measurements, it was found that the neutron measurement by activation of an NaI scintillator has a great advantage in estimates of a low neutron fluence rate by use of a quick measurement following a short-time irradiation. Also, the future application of this method to quasi real-time monitoring of neutrons during patient treatments at a radiotherapy facility is discussed, as well as the method of evaluation of the neutron dose.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25404493     DOI: 10.1007/s12194-014-0300-3

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  6 in total

1.  Dose to radiation therapists from activation at high-energy accelerators used for conventional and intensity-modulated radiation therapy.

Authors:  J Alan Rawlinson; Mohammad K Islam; Duncan M Galbraith
Journal:  Med Phys       Date:  2002-04       Impact factor: 4.071

2.  Measurement of thermal neutron fluence distribution with use of 23Na radioactivation around a medical compact cyclotron.

Authors:  Toshioh Fujibuchi; Ichiro Yamaguchi; Tetsuharu Kasahara; Takashi Iimori; Yoshitada Masuda; Ken-ichi Kimura; Hiroshi Watanabe; Tomonori Isobe; Takeji Sakae
Journal:  Radiol Phys Technol       Date:  2009-05-16

3.  Measurements of secondary neutron dose from 15 MV and 18 MV IMRT.

Authors:  Rebecca M Howell; Michele S Ferenci; Nolan E Hertel; Gary D Fullerton; Timothy Fox; Lawrence W Davis
Journal:  Radiat Prot Dosimetry       Date:  2005       Impact factor: 0.972

4.  Neutron spectra and dose equivalents calculated in tissue for high-energy radiation therapy.

Authors:  Stephen F Kry; Rebecca M Howell; Mohammad Salehpour; David S Followill
Journal:  Med Phys       Date:  2009-04       Impact factor: 4.071

5.  ICRP Publication 107. Nuclear decay data for dosimetric calculations.

Authors:  K Eckerman; A Endo
Journal:  Ann ICRP       Date:  2008

6.  Photoneutrons from medical linear accelerators--radiobiological measurements and risk estimates.

Authors:  E J Hall; S G Martin; H Amols; T K Hei
Journal:  Int J Radiat Oncol Biol Phys       Date:  1995-08-30       Impact factor: 7.038

  6 in total
  1 in total

Review 1.  Medical application of particle and heavy ion transport code system PHITS.

Authors:  Takuya Furuta; Tatsuhiko Sato
Journal:  Radiol Phys Technol       Date:  2021-06-30
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.