Literature DB >> 20821115

Measurement of thermal neutron fluence distribution with use of 23Na radioactivation around a medical compact cyclotron.

Toshioh Fujibuchi1, Ichiro Yamaguchi, Tetsuharu Kasahara, Takashi Iimori, Yoshitada Masuda, Ken-ichi Kimura, Hiroshi Watanabe, Tomonori Isobe, Takeji Sakae.   

Abstract

A medical compact cyclotron produces about 10(15) neutrons per day along with 100 GBq of (18)F. Therefore, it is important to establish radiation safety guidelines on residual radioactivity for routine operation, maintenance work, and decommissioning. Thus, we developed a simple method for measuring the thermal neutrons in a cyclotron room. In order to verify the feasibility of our proposed method, we measured the thermal neutron distribution around a cyclotron by using the activation of (23)Na in salt. We installed 78 salt dosimeters in the cyclotron room with a 50 cm mesh. The photopeak of (24)Na was measured, and the neutron flux distribution was estimated. Monitoring the neutron flux distribution in a cyclotron room appears to be useful for not only obtaining an accurate estimate of the distribution of induced radioactivity, but also optimizing the shield design for radiation safety in preparation for the decommissioning process.

Mesh:

Substances:

Year:  2009        PMID: 20821115     DOI: 10.1007/s12194-009-0060-7

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  4 in total

1.  Applicability of self-activation of an NaI scintillator for measurement of photo-neutrons around a high-energy X-ray radiotherapy machine.

Authors:  Genichiro Wakabayashi; Akihiro Nohtomi; Eriko Yahiro; Toshioh Fujibuchi; Junichi Fukunaga; Yoshiyuki Umezu; Yasuhiko Nakamura; Katsumasa Nakamura; Makoto Hosono; Tetsuo Itoh
Journal:  Radiol Phys Technol       Date:  2014-11-18

2.  Comparison of neutron fluxes in an 18-MeV unshielded cyclotron room and a 16.5-MeV self-shielded cyclotron room.

Authors:  Toshioh Fujibuchi; Genki Horitsugi; Ichiro Yamaguchi; Akihisa Eto; Yasuo Iwamoto; Satoshi Obara; Takashi Iimori; Yoshitada Masuda; Hiroshi Watanabe; Jun Hatazawa
Journal:  Radiol Phys Technol       Date:  2012-03-24

3.  Radiologic assessment of a self-shield with boron-containing water for a compact medical cyclotron.

Authors:  Genki Horitsugi; Toshioh Fujibuchi; Ichiro Yamaguchi; Akihisa Eto; Yasuo Iwamoto; Hiromi Hashimoto; Seiki Hamada; Satoshi Obara; Hiroshi Watanabe; Jun Hatazawa
Journal:  Radiol Phys Technol       Date:  2012-02-28

4.  Solution Radioactivated by Hadron Radiation Can Increase Sister Chromatid Exchanges.

Authors:  Junko Maeda; Charles R Yurkon; Yoshihiro Fujii; Hiroshi Fujisawa; Sayaka Kato; Colleen A Brents; Mitsuru Uesaka; Akira Fujimori; Hisashi Kitamura; Takamitsu A Kato
Journal:  PLoS One       Date:  2015-12-14       Impact factor: 3.240

  4 in total