Faisal M Merchant1, Paul Jones2, Scott Wehrenberg2, Michael S Lloyd1, Leslie A Saxon3.
Abstract
BACKGROUND: A significant number of implantable cardioverter-defibrillator (ICD) patients do not experience shocks after ICD implant. Elective generator exchange (GE) has been associated with increased risk of infection and ICD lead failure. There is a paucity of contemporary data reporting on shock incidence with replacement devices. METHODS ANDEntities:
Keywords: defibrillation; electrophysiology; sudden death
Mesh:
Year: 2014 PMID: 25385346 PMCID: PMC4338723 DOI: 10.1161/JAHA.114.001289
Source DB: PubMed Journal: J Am Heart Assoc ISSN: 2047-9980 Impact factor: 5.501
Baseline Characteristics
| Group A (n=16 230) | Group B (n=7973) | ||
|---|---|---|---|
| Age at GE, y | 71.3±11.6 | 68.8±12.5 | <0.001 |
| Male sex, % | 70.9 | 76.9 | <0.001 |
| Time of first battery implant, y | 5.4±1.3 | 5.3±1.2 | <0.001 |
| Device type, % | <0.001 | ||
| Single‐chamber ICD | 21.4 | 25.5 | |
| Dual‐chamber ICD | 34.0 | 37.4 | |
| Biventricular ICD | 44.6 | 37.1 | |
| Tachytherapy device programming, % | <0.001 | ||
| 1 zone | 27.7 | 20.6 | |
| 2 zones | 51.5 | 53.4 | |
| 3 zones | 20.8 | 26.0 | |
| VT dectection rate, bpm | 174±14 | 172±16 | <0.001 |
| VF detection rate, bpm | 202±17 | 203±18 | <0.001 |
| Afib burden >5% | 24.5 | 31.9 | <0.001 |
| RV pacing >5%, % | 42.9 | 47.0 | <0.001 |
| LV pacing >95% | 75.9 | 61.8 | <0.001 |
| DFT performed at GE, % | 69.7 | 68.3 | 0.025 |
Afib indicates atrial fibrillation; bpm, beats per minute; DFT, defibrillation threshold testing; GE, generator exchange; ICD, implantable cardioverter‐defibrillator; LV, left ventricle; RV, right ventricle; VF, ventricular fibrillation; VT, ventricular tachycardia.
Among patients with atrial leads.
Among patients with cardiac resynchronization devices.
Figure 1.Cumulative rate of ICD shocks following generator exchange in groups A and B. Cumulative rate and number at risk are specified beneath each time point. ICD indicates implantable cardioverter‐defibrillator.
Figure 2.Cumulative rate of ICD shocks for detection rates >200 bpm following generator exchange in groups A and B. Cumulative rate and number at risk are specified beneath each time point. bpm indicates beats per minute; ICD, implantable cardioverter‐defibrillator.
Multivariate Cox Proportional Hazards Model of Time to First ICD Shock After Generator Exchange (Top Panel) and Time to First ICD Shock For Detection Rate >200 bpm (Bottom Panel)
| Covariate | Hazard Ratio (95% CI) | |
|---|---|---|
| Group A (relative to group B) | 0.31 (0.29 to 0.33) | <0.001 |
| Age | 0.99 (0.99 to 1.00) | 0.001 |
| Male sex | 1.43 (1.32 to 1.55) | <0.001 |
| Dual‐chamber ICD (relative to single chamber) | 0.93 (0.86 to 1.01) | 0.088 |
| CRT‐D (relative to single chamber) | 1.01 (0.93 to 1.10) | 0.743 |
| Implant DFT performed | 0.31 (0.29 to 0.33) | 0.020 |
| Group A (relative to group B) | 0.33 (0.31 to 0.36) | <0.001 |
| Age | 0.99 (0.99 to 1.00) | <0.001 |
| Male sex | 1.55 (1.40 to 1.73) | <0.001 |
| Dual‐chamber ICD (relative to single chamber) | 0.84 (0.75 to 0.93) | 0.001 |
| CRT‐D (relative to single chamber) | 0.94 (0.85 to 1.05) | 0.282 |
| Implant DFT performed | 0.88 (0.80 to 0.96) | 0.006 |
bpm indicates beats per minute; CRT‐D, cardiac resynchronization therapy‐defibrillator; DFT, defibrillation threshold test; ICD, implantable cardioverter‐defibrillator.