Buckling resistance, bending stiffness, and torsional resistance of various instruments for canal exploration and glide path preparation.

Sang-Won Kwak1, Jung-Hong Ha2, WooCheol Lee3, Sung-Kyo Kim2, Hyeon-Cheol Kim1.   

Abstract

OBJECTIVES: This study compared the mechanical properties of various instruments for canal exploration and glide-path preparations.
MATERIALS AND METHODS: The buckling resistance, bending stiffness, ultimate torsional strength, and fracture angle under torsional load were compared for C+ file (CP, Dentsply Maillefer), M access K-file (MA, Dentsply Maillefer), Mani K-file (MN, Mani), and NiTiFlex K-file (NT, Dentsply Maillefer). The files of ISO size #15 and a shaft length of 25 mm were selected. For measuring buckling resistance (n = 10), the files were loaded in the axial direction of the shaft, and the maximum load was measured during the files' deflection. The files (n = 10) were fixed at 3-mm from the tip and then bent 45° with respect to their long axis, while the bending force was recorded by a load cell. For measuring the torsional properties, the files (n = 10) were also fixed at 3-mm, and clockwise rotations (2-rpm) were applied to the files in a straight state. The torsional load and the distortion angle were recorded until the files succumbed to the torque.
RESULTS: The CP was shown to require the highest load to buckle and bend the files, and the NT showed the least. While MA and MN showed similar buckling resistances, MN showed higher bending stiffness than MA. The NT had the lowest bending stiffness and ultimate torsional strength (p < 0.05).
CONCLUSIONS: The tested instruments showed different mechanical properties depending on the evaluated parameters. CP and NT files were revealed to be the stiffest and the most flexible instruments, respectively.

Entities:  

Keywords:  Bending stiffness; Buckling resistance; Glide path; Nickel-titanium rotary file; Stainless steel file; Torsional resistance

Year:  2014        PMID: 25383345      PMCID: PMC4223096          DOI: 10.5395/rde.2014.39.4.270

Source DB:  PubMed          Journal:  Restor Dent Endod        ISSN: 2234-7658


  21 in total

1.  Analysis of mechanical preparations in extracted teeth using ProTaper rotary instruments: value of the safety quotient.

Authors:  J Y Blum; P Machtou; C Ruddle; J P Micallef
Journal:  J Endod       Date:  2003-09       Impact factor: 4.171

2.  Buckling resistance of pathfinding endodontic instruments.

Authors:  Hélio P Lopes; Carlos N Elias; Marcelo Mangelli; Weber S P Lopes; Giorgiana Amaral; Letícia C Souza; José F Siqueira
Journal:  J Endod       Date:  2011-12-22       Impact factor: 4.171

3.  The endodontic Glidepath: "Secret to rotary safety".

Authors:  John D West
Journal:  Dent Today       Date:  2010-09

4.  Comparative analysis of endodontic pathfinders.

Authors:  Michael J Allen; Gerald N Glickman; Jason A Griggs
Journal:  J Endod       Date:  2007-04-09       Impact factor: 4.171

5.  Use of nickel-titanium rotary PathFile to create the glide path: comparison with manual preflaring in simulated root canals.

Authors:  Elio Berutti; Giuseppe Cantatore; Arnaldo Castellucci; Giorgio Chiandussi; Francesco Pera; Giuseppe Migliaretti; Damiano Pasqualini
Journal:  J Endod       Date:  2009-03       Impact factor: 4.171

6.  Flexibility and torsional behaviour of rotary nickel-titanium PathFile, RaCe ISO 10, Scout RaCe and stainless steel K-File hand instruments.

Authors:  R K L Nakagawa; J L Alves; V T L Buono; M G A Bahia
Journal:  Int Endod J       Date:  2013-07-02       Impact factor: 5.264

7.  A retrospective study of endodontic treatment outcome between nickel-titanium rotary and stainless steel hand filing techniques.

Authors:  Gary S P Cheung; Christopher S Y Liu
Journal:  J Endod       Date:  2009-07       Impact factor: 4.171

8.  Correlation between experimental cyclic fatigue resistance and numerical stress analysis for nickel-titanium rotary files.

Authors:  Min-Ho Lee; Antheunis Versluis; Byung-Min Kim; Chan-Joo Lee; Bock Hur; Hyeon-Cheol Kim
Journal:  J Endod       Date:  2011-05-17       Impact factor: 4.171

9.  An in vitro comparison of torsional stress and cyclic fatigue resistance of ProFile GT and ProFile GT Series X rotary nickel-titanium files.

Authors:  Thomas R Kramkowski; James Bahcall
Journal:  J Endod       Date:  2009-03       Impact factor: 4.171

10.  Cyclic fatigue resistance tests of Nickel-Titanium rotary files using simulated canal and weight loading conditions.

Authors:  Ok-In Cho; Antheunis Versluis; Gary Sp Cheung; Jung-Hong Ha; Bock Hur; Hyeon-Cheol Kim
Journal:  Restor Dent Endod       Date:  2013-02-26
View more
  2 in total

1.  Debris extrusion by glide-path establishing endodontic instruments with different geometries.

Authors:  Jung-Hong Ha; Sung Kyo Kim; Sang Won Kwak; Rashid El Abed; Yong Chul Bae; Hyeon-Cheol Kim
Journal:  J Dent Sci       Date:  2016-04-18       Impact factor: 2.080

2.  Shaping Ability and Buckling Resistance of TruNatomy, WaveOne gold, and XP-Endo Shaper Single-File Systems.

Authors:  Neveen Ali Shaheen; Nahla Gamal Eldin Elhelbawy
Journal:  Contemp Clin Dent       Date:  2022-09-24
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.