Literature DB >> 22341084

Buckling resistance of pathfinding endodontic instruments.

Hélio P Lopes1, Carlos N Elias, Marcelo Mangelli, Weber S P Lopes, Giorgiana Amaral, Letícia C Souza, José F Siqueira.   

Abstract

INTRODUCTION: This study compared the buckling resistance of the following endodontic pathfinding instruments: C-Pilot file (VDW, Munich, Germany), C+ file (Maillefer/Dentsply, Ballaigues, Switzerland), and PathFile (Maillefer/Dentsply).
METHODS: The test instruments were subjected to a devised buckling resistance test, which consisted of the application of an increasing load in the axial direction of the instrument by using a universal testing machine. The maximum load required to generate a lateral elastic displacement of 1 mm was recorded for each instrument.
RESULTS: The results indicated that the buckling resistance decreased in the following order: C+ file > C-Pilot file > PathFile. The difference was statistically significant (P < .05).
CONCLUSIONS: The stainless steel instruments (C+ and C-Pilot) were more resistant to buckling than the nickel-titanium instrument (PathFile). Considering that buckling resistance may influence the performance of instruments during the negotiation of constricted canals, the C+ files showed significantly better results than the other instruments tested. Copyright Â
© 2012 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22341084     DOI: 10.1016/j.joen.2011.10.029

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  3 in total

1.  Buckling resistance, bending stiffness, and torsional resistance of various instruments for canal exploration and glide path preparation.

Authors:  Sang-Won Kwak; Jung-Hong Ha; WooCheol Lee; Sung-Kyo Kim; Hyeon-Cheol Kim
Journal:  Restor Dent Endod       Date:  2014-07-16

2.  Evaluation of the Canal Transportation Following Glide Path Preparation with Different Rotary Systems: A Comparative Study.

Authors:  Mohsen Aminsobhani; Fatemeh Hamidzadeh; Arvin Rezaei Avval; Farid Merrikhi; Ehsan Sadri
Journal:  ScientificWorldJournal       Date:  2022-06-30

3.  Shaping Ability and Buckling Resistance of TruNatomy, WaveOne gold, and XP-Endo Shaper Single-File Systems.

Authors:  Neveen Ali Shaheen; Nahla Gamal Eldin Elhelbawy
Journal:  Contemp Clin Dent       Date:  2022-09-24
  3 in total