Tetsuo Saito1, Tomohiko Matsuyama1, Ryo Toya1, Yoshiyuki Fukugawa1, Takamasa Toyofuku1, Akiko Semba1, Natsuo Oya1.
Abstract
PURPOSE: We evaluated the effects of respiratory gating on treatment accuracy in lung cancer patients undergoing lung stereotactic body radiotherapy by using electronic portal imaging device (EPID) images.Entities:
Mesh:
Year: 2014 PMID: 25379729 PMCID: PMC4224502 DOI: 10.1371/journal.pone.0112824
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Figure 1Tumor displacement measured with reference to the first EPID image acquired at the first session.
EPID = electronic portal imaging device.
Figure 2Variability assessment for each patient.
Variability was assessed by determining the standard deviation (SD) and range of the tumor position during all sessions.
Figure 3Three-dimensional tumor respiratory motion during treatment simulation evaluated with four-dimensional CT.
Figure 4Standard deviation (SD) of the tumor position on electronic portal imaging device (EPID) images.
Each point represents a single patient's SD of the tumor position during all treatment sessions. CC = craniocaudal; ML = mediolateral.
Figure 5Range of the tumor position on electronic portal imaging device (EPID) images.
Each point represents a single patient's range of the tumor position during all treatment sessions. CC = craniocaudal; ML = mediolateral.
Earlier reports on the variability of the lung tumor position evaluated on EPID images.
| Authors (year) [ref] | Number of patients | Respiratory motion during simulation | Respiratory gating | Intra/interfractional variability | Variability assessed by SD | Variability assessed by range |
| Ford EC (2002) | 2 | 6.9 mm | yes | intrafractional | 2.6–5.7 mm (range) | - |
| Gaede S (2008) | 4 | - | yes | intrafractional | 0.94 mm (not specified) | - |
| 1 | - | no | 2.1 mm (not specified) | - | ||
| Spoelstra FOB (2008) | 11 | 5.0–19.5 mm | yes | Interfractional | 1.3–2.1 mm (range) | - |
| Ueda (2012) | 28 | 3.1 mm | no | intrafractional | - | 4.0 mm (mean) |
| Present study | 14 | 9 mm | yes | Interfractional | 1.1 mm (median) | 4.0 mm (median) |
| 16 | 2 mm | no | 0.9 mm (median) | 3.0 mm (median) |
EPID = electronic portal imaging device, SD = standard deviation.
Tumor position variability during one treatment session.
Tumor position variability during more than one treatment session.
SD of the tumor position in the craniocaudal direction.
Range of the tumor position in the craniocaudal direction.