Portal imaging to assess set-up errors, tumor motion and tumor shrinkage during conformal radiotherapy of non-small cell lung cancer.

Sara C Erridge1, Yvette Seppenwoolde, Sara H Muller, Marcel van Herk, Katrien De Jaeger, José S A Belderbos, Liesbeth J Boersma, Joos V Lebesque.   

Abstract

PURPOSE: To investigate patient set-up, tumor movement and shrinkage during 3D conformal radiotherapy for non-small cell lung cancer.
MATERIALS AND METHODS: In 97 patients, electronic portal images (EPIs) were acquired and corrected for set-up using an off-line correction protocol based on a shrinking action level. For 25 selected patients, the orthogonal EPIs (taken at random points in the breathing cycle) throughout the 6-7 week course of treatment were assessed to establish the tumor position in each image using both an overlay and a delineation technique. The range of movement in each direction was calculated. The position of the tumor in the digitally reconstructed radiograph (DRR) was compared to the average position of the lesion in the EPIs. In addition, tumor shrinkage was assessed.
RESULTS: The mean overall set-up errors after correction were 0, 0.6 and 0.2 mm in the x (left-right), y (cranial-caudal) and z (anterior-posterior) directions, respectively. After correction, the standard deviations (SDs) of systematic errors were 1.4, 1.5 and 1.3 mm and the SDs of random errors were 2.9, 3.1 and 2.0 mm in the x-, y- and z-directions, respectively. Without correction, 41% of patients had a set-up error of more than 5 mm vector length, but with the set-up correction protocol this percentage was reduced to 1%. The mean amplitude of tumor motion was 7.3 (SD 2.7), 12.5 (SD 7.3) and 9.4 mm (SD 5.2) in the x-, y- and z-directions, respectively. Tumor motion was greatest in the y-direction and in particular for lower lobe tumors. In 40% of the patients, the projected area of the tumor regressed by more than 20% during treatment in at least one projection. In 16 patients it was possible to define the position of the center of the tumor in the DRR. There was a mean difference of 6 mm vector length between the tumor position in the DRR and the average position in the portal images.
CONCLUSIONS: The application of the correction protocol resulted in a significant improvement in the set-up accuracy. There was wide variation in the observed tumor motion with more movement of lower lobe lesions. Tumor shrinkage was observed. The position of the tumor on the planning CT scan did not always coincide with the average position as measured during treatment.

Entities:  

Mesh:

Year:  2003        PMID: 12559524     DOI: 10.1016/s0167-8140(02)00287-6

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  40 in total

1.  Localization accuracy from automatic and semi-automatic rigid registration of locally-advanced lung cancer targets during image-guided radiation therapy.

Authors:  Scott P Robertson; Elisabeth Weiss; Geoffrey D Hugo
Journal:  Med Phys       Date:  2012-01       Impact factor: 4.071

2.  A novel technique for markerless, self-sorted 4D-CBCT: feasibility study.

Authors:  Irina Vergalasova; Jing Cai; Fang-Fang Yin
Journal:  Med Phys       Date:  2012-03       Impact factor: 4.071

3.  Design of a compensating bolus by use of exhalation CT data for covering residual motion in respiratory-gated charged-particle lung therapy: four-dimensional carbon beam dose calculation.

Authors:  Shinichiro Mori; Hiroshi Asakura; Shuhei Komatsu; Tomoyasu Yashiro; Motoki Kumagai; Susumu Kandatsu; Masayuki Baba; Masahiro Endo
Journal:  Radiol Phys Technol       Date:  2007-11-27

4.  Practical approaches to four-dimensional heavy-charged-particle lung therapy.

Authors:  Shinichiro Mori; Ziji Wu; Michael R Folkert; Motoki Kumagai; Suguru Dobashi; Toshio Sugane; Masayuki Baba
Journal:  Radiol Phys Technol       Date:  2009-10-14

5.  A simulation technique for computation of the dosimetric effects of setup, organ motion and delineation uncertainties in radiotherapy.

Authors:  Bongile Mzenda; Mir Hosseini-Ashrafi; Antony Palmer; Honghai Liu; David J Brown
Journal:  Med Biol Eng Comput       Date:  2010-04-23       Impact factor: 2.602

6.  Predictive modeling of lung motion over the entire respiratory cycle using measured pressure-volume data, 4DCT images, and finite-element analysis.

Authors:  Jaesung Eom; Xie George Xu; Suvranu De; Chengyu Shi
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

Review 7.  Anatomic, functional and molecular imaging in lung cancer precision radiation therapy: treatment response assessment and radiation therapy personalization.

Authors:  Michael MacManus; Sarah Everitt; Tanja Schimek-Jasch; X Allen Li; Ursula Nestle; Feng-Ming Spring Kong
Journal:  Transl Lung Cancer Res       Date:  2017-12

8.  A simplified method of four-dimensional dose accumulation using the mean patient density representation.

Authors:  Carri K Glide-Hurst; Geoffrey D Hugo; Jian Liang; Di Yan
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

9.  Dose-guided radiotherapy for lung tumors.

Authors:  Angelo Piermattei; Andrea Fidanzio; Savino Cilla; Francesca Greco; Luigi Azario; Domenico Sabatino; Mattia Grusio; Mariella Cozzolino; Vincenzo Fusco
Journal:  Med Biol Eng Comput       Date:  2009-12-10       Impact factor: 2.602

10.  An evaluation of planning techniques for stereotactic body radiation therapy in lung tumors.

Authors:  Jianzhou Wu; Huiling Li; Raj Shekhar; Mohan Suntharalingam; Warren D'Souza
Journal:  Radiother Oncol       Date:  2008-03-24       Impact factor: 6.280

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.