Three-dimensional biplanar radiography as a new means of accessing femoral version: a comparitive study of EOS three-dimensional radiography versus computed tomography.

M Lucius Pomerantz1, Diana Glaser, Josh Doan, Sita Kumar, Eric W Edmonds.   

Abstract

OBJECTIVE: To validate femoral version measurements made from biplanar radiography (BR), three-dimensional (3D) reconstructions (EOS imaging, France) were made in differing rotational positions against the gold standard of computed tomography (CT).
MATERIALS AND METHODS: Two cadaveric femurs were scanned with CT and BR in five different femoral versions creating ten total phantoms. The native version was modified by rotating through a mid-diaphyseal hinge twice into increasing anteversion and twice into increased retroversion. For each biplanar scan, the phantom itself was rotated -10, -5, 0, +5 and +10°. Three-dimensional CT reconstructions were designated the true value for femoral version. Two independent observers measured the femoral version on CT axial slices and BR 3D reconstructions twice. The mean error (upper bound of the 95% confidence interval), inter- and intraobserver reliability, and the error compared to the true version were determined for both imaging techniques.
RESULTS: Interobserver intraclass correlation for CT axial images ranged from 0.981 to 0.991, and the intraobserver intraclass correlation ranged from 0.994 to 0.996. For the BR 3D reconstructions these values ranged from 0.983 to 0.998 and 0.982 to 0.998, respectively. For the CT measurements the upper bound of error from the true value was 5.4-7.5°, whereas for BR 3D reconstructions it was 4.0-10.1°. There was no statistical difference in the mean error from the true values for any of the measurements done with axial CT or BR 3D reconstructions.
CONCLUSION: BR 3D reconstructions accurately and reliably provide clinical data on femoral version compared to CT even with rotation of the patient of up to 10° from neutral.

Entities:  

Mesh:

Year:  2014        PMID: 25319563     DOI: 10.1007/s00256-014-2031-2

Source DB:  PubMed          Journal:  Skeletal Radiol        ISSN: 0364-2348            Impact factor:   2.199


  24 in total

Review 1.  Computed tomography--an increasing source of radiation exposure.

Authors:  David J Brenner; Eric J Hall
Journal:  N Engl J Med       Date:  2007-11-29       Impact factor: 91.245

2.  Femoral version in acute slipped capital femoral epiphysis.

Authors:  C L Stanitski; R Woo; D F Stanitski
Journal:  J Pediatr Orthop B       Date:  1996       Impact factor: 1.041

3.  Fast 3D reconstruction of the lower limb using a parametric model and statistical inferences and clinical measurements calculation from biplanar X-rays.

Authors:  Y Chaibi; T Cresson; B Aubert; J Hausselle; P Neyret; O Hauger; J A de Guise; W Skalli
Journal:  Comput Methods Biomech Biomed Engin       Date:  2011-05-24       Impact factor: 1.763

4.  Femoral and tibial torsion measurements with 3D models based on low-dose biplanar radiographs in comparison with standard CT measurements.

Authors:  Florian M Buck; Roman Guggenberger; Peter P Koch; Christian W A Pfirrmann
Journal:  AJR Am J Roentgenol       Date:  2012-11       Impact factor: 3.959

5.  Estimates of effective dose for CT scans of the lower extremities.

Authors:  Natalia Saltybaeva; Mary Ellen Jafari; Martin Hupfer; Willi A Kalender
Journal:  Radiology       Date:  2014-06-16       Impact factor: 11.105

6.  A simple biplanar method of measuring femoral anteversion and neck-shaft angle.

Authors:  K Ogata; E M Goldsand
Journal:  J Bone Joint Surg Am       Date:  1979-09       Impact factor: 5.284

7.  Diagnostic imaging of spinal deformities: reducing patients radiation dose with a new slot-scanning X-ray imager.

Authors:  Sylvain Deschênes; Guy Charron; Gilles Beaudoin; Hubert Labelle; Josée Dubois; Marie-Claude Miron; Stefan Parent
Journal:  Spine (Phila Pa 1976)       Date:  2010-04-20       Impact factor: 3.468

8.  Femoral and tibial torsion measurement in children and adolescents: comparison of 3D models based on low-dose biplanar radiography and low-dose CT.

Authors:  Andrea B Rosskopf; Leonhard E Ramseier; Reto Sutter; Christian W A Pfirrmann; Florian M Buck
Journal:  AJR Am J Roentgenol       Date:  2014-03       Impact factor: 3.959

9.  Femoral anteversion in developmental dysplasia of the hip.

Authors:  Wudbhav N Sankar; Christopher O Neubuerger; Colin F Moseley
Journal:  J Pediatr Orthop       Date:  2009-12       Impact factor: 2.324

10.  Bilateral double osteotomy in severe torsional malalignment syndrome: 16 years follow-up.

Authors:  Francesco Leonardi; Fabrizio Rivera; Alessandra Zorzan; Syed Mohsin Ali
Journal:  J Orthop Traumatol       Date:  2013-08-29
View more
  5 in total

1.  Measuring physiological and pathological femoral anteversion using a biplanar low-dose X-ray system: validity, reliability, and discriminative ability in cerebral palsy.

Authors:  Matthias Thépaut; Sylvain Brochard; Julien Leboucher; Mathieu Lempereur; Eric Stindel; Valentin Tissot; Bhushan S Borotikar
Journal:  Skeletal Radiol       Date:  2015-11-27       Impact factor: 2.199

2.  Development and Validation of a Mobile Application for Measuring Femoral Anteversion in Patients With Cerebral Palsy.

Authors:  Ki Hyuk Sung; Kibeom Youn; Chin Youb Chung; Muhammad I Kitta; Hendra C Kumara; Jae Jung Min; Jehee Lee; Moon Seok Park
Journal:  J Pediatr Orthop       Date:  2019-11-01       Impact factor: 2.324

3.  Using the Axial Oblique View of Computed Tomography (CT) in Evaluating Femoral Anteversion: A Comparative Cadaveric Study.

Authors:  Kwang-Soon Song; Chang-Jin Yon; Yu-Ran Heo; Jae-Ho Lee; Seung-Bo Lee; Yeon-Kyoung Ko; Kyung-Jae Lee; Si-Wook Lee
Journal:  Diagnostics (Basel)       Date:  2022-07-28

4.  EOS(®) biplanar X-ray imaging: concept, developments, benefits, and limitations.

Authors:  Elias Melhem; Ayman Assi; Rami El Rachkidi; Ismat Ghanem
Journal:  J Child Orthop       Date:  2016-02-16       Impact factor: 1.548

5.  Development and Validation of a Mobile Application for Measuring Femoral Anteversion in Patients With Cerebral Palsy.

Authors:  Ki Hyuk Sung; Kibeom Youn; Chin Youb Chung; Muhammad I Kitta; Hendra C Kumara; Jae Jung Min; Jehee Lee; Moon Seok Park
Journal:  J Pediatr Orthop       Date:  2020-07       Impact factor: 2.537

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.