Literature DB >> 23096205

Femoral and tibial torsion measurements with 3D models based on low-dose biplanar radiographs in comparison with standard CT measurements.

Florian M Buck1, Roman Guggenberger, Peter P Koch, Christian W A Pfirrmann.   

Abstract

OBJECTIVE: The purpose of this study was to evaluate the interchangeability of femoral and tibial torsion measurements obtained with 3D models based on low-dose biplanar radiographs and standard CT measurements by testing the following two hypotheses: that there is excellent agreement between the two methods and that there is excellent interreader agreement.
MATERIALS AND METHODS: Two independent readers used 3D models based on low-dose simultaneous biplanar radiographs and axial CT images to measure femoral and tibial torsion in 35 patients (mean age, 65 years; range, 46-89 years) with osteoarthritis of the knee who were to undergo prosthesis insertion. The two measurements were compared by means of Bland-Altman plots and descriptive statistics. Interreader agreement was quantified with intraclass correlation coefficients.
RESULTS: The average differences between readers on the CT measurements were 1.3° (range, 0°-11°) for the femur and 1.5° (range, 0°-12°) for the tibia. The average differences for the measurements obtained with the 3D model were 0.1° (range, 0°-9°) for the femur and 0.8° (range, 0°-10°) for the tibia. The average differences between the two methods were 0° (range, -5° to 7°) for the femoral measurements and 3° (range, -12° to 5°) for the tibial measurements. Bland-Altman plots showed no relevant differences between the results of the two measurement modalities. Except for one measurement of femoral torsion and one measurement of tibial torsion, all results based on the 3D models were within the 95% limit of agreement (mean ± 1.96 SD). Interreader agreement was statistically significant (p < 0.001) for all measurements with high intraclass correlation coefficients (> 0.9).
CONCLUSION: Femoral and tibial torsion measurements obtained with 3D models based on biplanar radiographs are interchangeable with standard CT measurements in patients with osteoarthritis of the knee.

Entities:  

Mesh:

Year:  2012        PMID: 23096205     DOI: 10.2214/AJR.11.8295

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  30 in total

1.  Assessment of two-dimensional (2D) and three-dimensional (3D) lower limb measurements in adults: Comparison of micro-dose and low-dose biplanar radiographs.

Authors:  Andrea B Rosskopf; Christian W A Pfirrmann; Florian M Buck
Journal:  Eur Radiol       Date:  2016-01-06       Impact factor: 5.315

2.  Measuring physiological and pathological femoral anteversion using a biplanar low-dose X-ray system: validity, reliability, and discriminative ability in cerebral palsy.

Authors:  Matthias Thépaut; Sylvain Brochard; Julien Leboucher; Mathieu Lempereur; Eric Stindel; Valentin Tissot; Bhushan S Borotikar
Journal:  Skeletal Radiol       Date:  2015-11-27       Impact factor: 2.199

3.  Three-dimensional biplanar radiography as a new means of accessing femoral version: a comparitive study of EOS three-dimensional radiography versus computed tomography.

Authors:  M Lucius Pomerantz; Diana Glaser; Josh Doan; Sita Kumar; Eric W Edmonds
Journal:  Skeletal Radiol       Date:  2014-10-17       Impact factor: 2.199

4.  Acetabular coverage differs between standing and supine positions: model-based assessment of low-dose biplanar radiographs and comparison with CT.

Authors:  Benjamin Fritz; Christoph A Agten; Franca K Boldt; Patrick O Zingg; Christian W A Pfirrmann; Reto Sutter
Journal:  Eur Radiol       Date:  2019-03-22       Impact factor: 5.315

5.  Femoral and tibial torsion measurements in children and adolescents: comparison of MRI and 3D models based on low-dose biplanar radiographs.

Authors:  Andrea B Rosskopf; Florian M Buck; Christian W A Pfirrmann; Leonhard E Ramseier
Journal:  Skeletal Radiol       Date:  2017-02-03       Impact factor: 2.199

6.  Development and Validation of a Mobile Application for Measuring Femoral Anteversion in Patients With Cerebral Palsy.

Authors:  Ki Hyuk Sung; Kibeom Youn; Chin Youb Chung; Muhammad I Kitta; Hendra C Kumara; Jae Jung Min; Jehee Lee; Moon Seok Park
Journal:  J Pediatr Orthop       Date:  2019-11-01       Impact factor: 2.324

7.  Evaluation of the 3-dimensional, weight-bearing orientation of the normal adult knee.

Authors:  Denis Nam; Ritesh R Shah; Ryan M Nunley; Robert L Barrack
Journal:  J Arthroplasty       Date:  2013-10-31       Impact factor: 4.757

8.  Offset and anteversion reconstruction after cemented and uncemented total hip arthroplasty: an evaluation with the low-dose EOS system comparing two- and three-dimensional imaging.

Authors:  Jean Yves Lazennec; Adrien Brusson; Folinais Dominique; Marc-Antoine Rousseau; Aidin Eslam Pour
Journal:  Int Orthop       Date:  2014-12-20       Impact factor: 3.075

9.  Comparison of radiation dose, workflow, patient comfort and financial break-even of standard digital radiography and a novel biplanar low-dose X-ray system for upright full-length lower limb and whole spine radiography.

Authors:  Tobias J Dietrich; Christian W A Pfirrmann; Alexander Schwab; Katja Pankalla; Florian M Buck
Journal:  Skeletal Radiol       Date:  2013-03-28       Impact factor: 2.199

Review 10.  How to assess femoral and tibial component rotation after total knee arthroplasty with computed tomography: a systematic review.

Authors:  Eduard J De Valk; Julia C A Noorduyn; Eduard L A R Mutsaerts
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-09-21       Impact factor: 4.342