| Literature DB >> 25308947 |
Christiane Krämer1, Natalie Jaspers2, Dirk Nierhoff2, Kathrin Kuhr3, Andrea Bowe2, Tobias Goeser2, Guido Michels4.
Abstract
The present study was conducted to assess the diagnostic accuracy of Acoustic Structure Quantification (ASQ) ultrasound software in liver biopsy of patients with liver fibrosis and cirrhosis. Eighty patients (47 ± 14 y, 41 men) with chronic liver diseases underwent ultrasound examination of the liver and liver biopsy. In addition to the standard-care ultrasound examination, three valid gray-scale images were obtained for each patient. With the ASQ software, the average and peak values (Cm(2)) of each ultrasound gray-scale image were calculated and then compared with histologic fibrosis staging (F0-F4). No correlation was found between ASQ values and histologic fibrosis stage (p > 0.05). Areas under the curve for the diagnosis of no or mild fibrosis (F0 and F1), moderate/severe fibrosis (F2 and F3) and cirrhosis (F4) using average/peak Cm(2) values of small regions of interest were 0.46/0.43, 0.62/0.68 and 0.38/0.33. Determination of liver fibrosis with ASQ in its present form as an alternative approach to liver biopsy is too imprecise.Entities:
Keywords: Acoustic structure quantification; Liver disease; Liver fibrosis; Ultrasound
Mesh:
Year: 2014 PMID: 25308947 DOI: 10.1016/j.ultrasmedbio.2014.07.020
Source DB: PubMed Journal: Ultrasound Med Biol ISSN: 0301-5629 Impact factor: 2.998