Literature DB >> 25260538

Fall in inspired oxygen and anaesthetic agent concentrations during change of soda lime absorber.

Byrappa Vinay1, Kadarapura Nanjundaiah Gopalakrishna, Ganne S Umamaheswara Rao.   

Abstract

Following an episode of reduction in inspired oxygen concentration (FiO(2)) and inhalational agent concentration (Fi agent) during the changing of a soda lime absorber, We conducted an in vitro experiment to understand the impact of disconnection of the absorber on inspired gas dilution at different fresh gas flows. We found that both in Dräger Fabius GS and Primus anaesthesia work stations, disconnection of the absorber caused progressive reduction in FiO(2) and Fi agent as the FGF was decreased. The operating principle of fresh gas decoupling (FGD) valve is a potential source of this complication, which must be kept in mind while changing the soda lime during the course of surgery where an anaesthetic work stations utilizing FGD valves are used.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25260538     DOI: 10.1007/s10877-014-9614-2

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  2 in total

1.  Awareness and hypoxia risk with Dräger Cato and Fabius anaesthesia machines.

Authors:  E Murphy; S Willis
Journal:  Anaesth Intensive Care       Date:  2004-10       Impact factor: 1.669

2.  [Intra-operative drop in sevoflurane and oxygen concentrations. Leakage of the ventilation system].

Authors:  M Roiss; K Bremer; G N Schmidt
Journal:  Anaesthesist       Date:  2009-12       Impact factor: 1.041

  2 in total
  3 in total

Review 1.  Journal of Clinical Monitoring and Computing 2015 end of year summary: anesthesia.

Authors:  Jan F A Hendrickx; Andre De Wolf; Stanley Skinner
Journal:  J Clin Monit Comput       Date:  2015-12-26       Impact factor: 2.502

2.  In response: B Vinay et al.: Fall in inspired oxygen and anaesthetic agent concentrations during change of soda lime absorber, J Clin Monit Comput (2015) Jun; 29 (3):373-376.

Authors:  Axel Bischoff; Alexander Herden
Journal:  J Clin Monit Comput       Date:  2016-01-13       Impact factor: 2.502

3.  Effects of the reservoir bag disconnection on inspired gases during general anesthesia: a simulator-based study.

Authors:  Miljenko Križmarić; Uroš Maver; Marko Zdravković; Dušan Mekiš
Journal:  BMC Anesthesiol       Date:  2021-02-01       Impact factor: 2.217

  3 in total