[Intra-operative drop in sevoflurane and oxygen concentrations. Leakage of the ventilation system].

M Roiss1, K Bremer, G N Schmidt.   

Abstract

After problem-free induction of narcosis in an 84-year-old female patient an intra-operative drop in sevoflurane and oxygen concentrations occurred during low-flow anesthesia. Although the concentrations of sevoflurane and oxygen in the fresh gas flow were increased no adequate elevation of the inspiratory concentrations could be achieved. Disconnection of the Dräger Primus IE manual bag-valve-mask could be identified as the cause of the drop in concentrations. Interestingly no error alarm function was initiated. This case demonstrates how important knowledge of the function, set-up and alarm conditions of respiratory machines is. This should be an important component of training in anesthesiology as well as securely established algorithms for difficult ventilation to ensure safe anesthesia despite technical failures.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 20012245     DOI: 10.1007/s00101-009-1638-5

Source DB:  PubMed          Journal:  Anaesthesist        ISSN: 0003-2417            Impact factor:   1.041


  2 in total

1.  [Anaesthesia tube constriction. Small defect--serious effect].

Authors:  T Steinke; J Radke
Journal:  Anaesthesist       Date:  2008-04       Impact factor: 1.041

2.  [Accidentally opened adjustable pressure-limiting valve. Failure of manual ventilation].

Authors:  I Eggemann; B W Böttiger; F Spöhr
Journal:  Anaesthesist       Date:  2009-03       Impact factor: 1.041

  2 in total
  2 in total

1.  Fall in inspired oxygen and anaesthetic agent concentrations during change of soda lime absorber.

Authors:  Byrappa Vinay; Kadarapura Nanjundaiah Gopalakrishna; Ganne S Umamaheswara Rao
Journal:  J Clin Monit Comput       Date:  2014-09-27       Impact factor: 2.502

2.  Effects of the reservoir bag disconnection on inspired gases during general anesthesia: a simulator-based study.

Authors:  Miljenko Križmarić; Uroš Maver; Marko Zdravković; Dušan Mekiš
Journal:  BMC Anesthesiol       Date:  2021-02-01       Impact factor: 2.217

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.