Effects of the reservoir bag disconnection on inspired gases during general anesthesia: a simulator-based study.

Miljenko Križmarić1,2, Uroš Maver3, Marko Zdravković3,4, Dušan Mekiš3,4.   

Abstract

BACKGROUND: Fresh gas decoupling is a feature of the modern anesthesia workstation, where the fresh gas flow (FGF) is diverted into the reservoir bag and is not added to the delivered tidal volume, which thus remains constant. The present study aimed to investigate the entraining of the atmospheric air into the anesthesia breathing circuit in case the reservoir bag was disconnected.
METHODS: We conducted a simulator-based study, where the METI HPS simulator was connected to the anesthesia workstation. The effect of the disconnected reservoir bag was evaluated using oxygen (O2) and air or oxygen and nitrous oxide (N2O) as a carrier gas at different FGF rates. We disconnected the reservoir bag for 10 min during the maintenance phase. We recorded values for inspiratory O2, N2O, and sevoflurane. The time constant of the exponential process was estimated during reservoir bag disconnection.
RESULTS: The difference of O2, N2O and sevoflurane concentrations, before, during, and after reservoir bag disconnection was statistically significant at 0.5, 1, and 2 L/min of FGF (p < 0.001). The largest decrease of the inspired O2 concentrations (FIO2) was detected in the case of oxygen and air as the carrier gas and an FGF of 1 L/min, when oxygen decreased from median [25th-75th percentile] 55.00% [54.00-56.00] to median 39.50% [38.00-42.50] (p < 0.001). The time constant for FIO2 during reservoir bag disconnection in oxygen and air as the carrier gas, were median 2.5, 2.5, and 1.5 min in FGF of 0.5, 1.0, and 2 L/min respectively.
CONCLUSIONS: During the disconnection of the anesthesia reservoir bag, the process of pharmacokinetics takes place faster compared to the wash-in and wash-out pharmacokinetic properties in the circle breathing system. The time constant was affected by the FGF rate, as well as the gradient of anesthetic gases between the anesthesia circle system and atmospheric air.

Entities:  

Keywords:  Anesthesia ventilators; Equipment design; Fresh gas decoupling; Low-flow anesthesia; Patient safety; Ventilation

Mesh:

Substances:

Year:  2021        PMID: 33522905      PMCID: PMC7848043          DOI: 10.1186/s12871-021-01256-2

Source DB:  PubMed          Journal:  BMC Anesthesiol        ISSN: 1471-2253            Impact factor:   2.217


  13 in total

1.  Novel breathing circuit architecture: new consequences of old problems.

Authors:  Warren S Sandberg; Sheila Kaiser
Journal:  Anesthesiology       Date:  2004-03       Impact factor: 7.892

2.  [A study of the parameters of the delivered tidal volume. Ventilation on a lung model using the CICERO anesthetic ventilator].

Authors:  U Schirmer; J Villwock; T Fösel; M Schreiber; A Goertz; M Georgieff
Journal:  Anaesthesist       Date:  1992-12       Impact factor: 1.041

3.  Fall in inspired oxygen and anaesthetic agent concentrations during change of soda lime absorber.

Authors:  Byrappa Vinay; Kadarapura Nanjundaiah Gopalakrishna; Ganne S Umamaheswara Rao
Journal:  J Clin Monit Comput       Date:  2014-09-27       Impact factor: 2.502

4.  Closed and low flow systems. Theoretical considerations.

Authors:  C M Conway
Journal:  Acta Anaesthesiol Belg       Date:  1984

5.  [The effect of fresh-gas decoupling on respiratory volume. Draegar Sulla 808V anesthesia ventilator].

Authors:  P Biro
Journal:  Anaesthesist       Date:  1997-11       Impact factor: 1.041

6.  Tidal volume changes due to the interaction of anesthesia machine and anesthesia ventilator.

Authors:  N Gravenstein; M J Banner; G McLaughlin
Journal:  J Clin Monit       Date:  1987-07

7.  [Mechanical ventilation in an anesthetic circle system using the lowest tidal volume--studies of 3 anesthesia ventilators in a lung model and an animal experiment].

Authors:  T Fösel; U Schirmer; C Wick; E Pfenninger; W Siegler; K H Altemeyer
Journal:  Anaesthesist       Date:  1991-11       Impact factor: 1.041

8.  Automated, real-time fresh gas flow recommendations alter isoflurane consumption during the maintenance phase of anesthesia in a simulator-based study.

Authors:  Isaac Luria; Samsun Lampotang; Wilhelm Schwab; Lou Ann Cooper; David Lizdas; Nikolaus Gravenstein
Journal:  Anesth Analg       Date:  2013-11       Impact factor: 5.108

9.  Teaching first-year medical students physiology: does the human patient simulator allow for more effective teaching?

Authors:  G M Tan; L K Ti; S Suresh; B S Ho; T L Lee
Journal:  Singapore Med J       Date:  2002-05       Impact factor: 1.858

Review 10.  Simulation-based assessments in health professional education: a systematic review.

Authors:  Tayne Ryall; Belinda K Judd; Christopher J Gordon
Journal:  J Multidiscip Healthc       Date:  2016-02-22
View more
  1 in total

1.  Programmed multi-level ventilation in COVID-19-related acute respiratory distress syndrome: a multi-center retrospective observational study.

Authors:  Filip Depta; Pavol Török; Andrew G Miller; Peter Firment; Jozef Leškanič; Adam Porubän; Pavol Halaš; Stanislav Mandinec; Vladimír Filka; Henryk Zajac; Michael A Gentile; Marko Zdravkovic
Journal:  J Int Med Res       Date:  2022-05       Impact factor: 1.573

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.