Flow-regulated extracorporeal arteriovenous tubing loop for cardiac output measurements by ultrasound velocity dilution: validation in post-cardiac surgery intensive care unit patients.

Alexsandr A Eremenko1, Perviz N Safarov.   

Abstract

Assessment of cardiac output (CO) is crucial in the management of the critically ill, especially in post cardiac surgery intensive care unit (ICU) patients. In this study, we validated CO measured by the novel ultrasound dilution (COUD) with those measured by pulmonary artery (PA) thermodilution (COTD) in 26 adult post cardiac surgery patients. For COUD, blood was circulated through an extracorporeal arteriovenous (AV) loop from the radial artery catheter to the introducer of PA catheter for 5-8 minutes. Three to four injections of 25 ml body temperature isotonic saline were performed into the venous limb of the AV loop. For COTD, five injections of 10 ml ice cold saline were performed. A total of 77 COUD and COTD measurement sets were compared. Cardiac output measured by thermodilution ranged from 3.28 to 9.4 L/min, whereas COUD ranged from 2.85 to 10.1 L/min. The correlation between the methods was found to be r = 0.91, COUD = 0.93(COTD) + 0.42 L/min. Bias and precision (mean difference ± 2SDs) was -0.004 ± 1.34 L/min between the two methods. The percentage error (2SD/mean) was 22.2%, which is below the clinically acceptable limit (<30%). Cardiac output measured by ultrasound dilution and thermodilution methods agreed well in post cardiac surgery ICU patients and hence can be interchangeably used.

Entities:  

Mesh:

Year:  2010        PMID: 21245798     DOI: 10.1097/MAT.0b013e3181effdf8

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872


  6 in total

1.  Ultrasound dilution: an accurate means of determining cardiac output in children.

Authors:  Ivory Crittendon; William J Dreyer; Jamie A Decker; Jeffrey J Kim
Journal:  Pediatr Crit Care Med       Date:  2012-01       Impact factor: 3.624

2.  Validation of a new method based on ultrasound velocity dilution to measure cardiac output in paediatric patients.

Authors:  Alejandro A Floh; Gustavo La Rotta; Julius Z Wermelt; Patricia Bastero-Miñón; V Ben Sivarajan; Tilman Humpl
Journal:  Intensive Care Med       Date:  2013-02-22       Impact factor: 17.440

3.  Hemodynamic volumetry using transpulmonary ultrasound dilution (TPUD) technology in a neonatal animal model.

Authors:  Sabine L Vrancken; Arno F van Heijst; Jeroen C Hopman; Kian D Liem; Johannes G van der Hoeven; Willem P de Boode
Journal:  J Clin Monit Comput       Date:  2014-12-14       Impact factor: 2.502

4.  Validation of cardiac output measurement by ultrasound dilution technique with pulmonary artery thermodilution in a pediatric animal model.

Authors:  Edward Darling; Naveen Thuramalla; Bruce Searles
Journal:  Pediatr Cardiol       Date:  2011-02-27       Impact factor: 1.655

5.  Accuracy of the transpulmonary ultrasound dilution method for detection of small anatomic shunts.

Authors:  R Saxena; N Krivitski; K Peacock; A Durward; J M Simpson; S M Tibby
Journal:  J Clin Monit Comput       Date:  2014-09-21       Impact factor: 2.502

6.  Cardiac output measurements via echocardiography versus thermodilution: A systematic review and meta-analysis.

Authors:  Yun Zhang; Yan Wang; Jing Shi; Zhiqiang Hua; Jinyu Xu
Journal:  PLoS One       Date:  2019-10-03       Impact factor: 3.240

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.