Literature DB >> 25216133

Settling of abutments into implants and changes in removal torque in five different implant-abutment connections. Part 1: Cyclic loading.

Ki-Seong Kim, Jung-Suk Han, Young-Jun Lim.   

Abstract

PURPOSE: The aim of this study was to evaluate and compare the settling of abutments into implants and the removal torque values (RTVs) before and after cyclic loading.
MATERIALS AND METHODS: Five different implant-abutment connections were tested: Ext = external butt joint + two-piece abutment; Int-H2 = internal hexagon + two-piece abutment; Int-H1 = internal hexagon + one-piece abutment; Int-O2 = internal octagon + two-piece abutment; and Int-O1 = internal octagon + one-piece abutment. Ten abutments from each group were secured to their corresponding implants (total n = 50). All samples were tested in a universal testing machine with a vertical load of 250 N for 100,000 cycles of 14 Hz. The amount of settling of the abutment into the implant was calculated from the change in the total length of the implant-abutment sample before and after loading, as measured with an electronic digital micrometer. The RTV after cyclic loading was compared to the initial RTV with a digital torque gauge. Statistical analysis was performed at a 5% significance level.
RESULTS: A multiple-comparison test showed specific significant differences in settling values in each group after 250 N cyclic loading (Int-H1, Ext < Int-H2 < Int-O2 < Int-O1). There were statistically significant decreases in RTVs after loading compared to the initial RTVs in the Int-H2 and Int-O2 groups. No statistically significant differences were found in the Ext, Int-H1, and Int-O1 groups.
CONCLUSION: The results of this study demonstrated that the settling amount and RTV (loss of preload) after cyclic loading were specific to the abutment type and related to the design characteristics of the implant-abutment connection.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25216133     DOI: 10.11607/jomi.3383

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  10 in total

1.  Evaluation the loosening of abutment screws in fluid contamination: an in vitro study.

Authors:  Fei Sun; Wei Cheng; Bao-Hong Zhao; Gui-Qiu Song; Zeng Lin
Journal:  Sci Rep       Date:  2022-06-24       Impact factor: 4.996

2.  A comparative biomechanical study of original and compatible titanium bases: evaluation of screw loosening and 3D-crown displacement following cyclic loading analysis.

Authors:  Rimantas Ožiūnas; Jurgina Sakalauskienė; Darius Jegelevičius; Gintaras Janužis
Journal:  J Adv Prosthodont       Date:  2022-04-27       Impact factor: 1.989

3.  The Mechanical Behavior of a Screwless Morse Taper Implant-Abutment Connection: An In Vitro Study.

Authors:  Aimen Bagegni; Vincent Weihrauch; Kirstin Vach; Ralf Kohal
Journal:  Materials (Basel)       Date:  2022-05-08       Impact factor: 3.748

4.  Effect of abutment screw length and cyclic loading on removal torque in external and internal hex implants.

Authors:  Hnd Hadi Mohammed; Jin-Han Lee; Ji-Myung Bae; Hye-Won Cho
Journal:  J Adv Prosthodont       Date:  2016-02-23       Impact factor: 1.904

Review 5.  Preload loss of abutment screws after dynamic fatigue in single implant-supported restorations. A systematic review.

Authors:  Beatriz Pardal-Peláez; Javier Montero
Journal:  J Clin Exp Dent       Date:  2017-11-01

6.  Axial Displacements and Removal Torque Changes of Five Different Implant-Abutment Connections under Static Vertical Loading.

Authors:  Ki-Seong Kim; Young-Jun Lim
Journal:  Materials (Basel)       Date:  2020-02-04       Impact factor: 3.623

Review 7.  Review of the Mechanical Behavior of Different Implant-Abutment Connections.

Authors:  Ana Sofia Vinhas; Carlos Aroso; Filomena Salazar; Paula López-Jarana; José Vicente Ríos-Santos; Mariano Herrero-Climent
Journal:  Int J Environ Res Public Health       Date:  2020-11-23       Impact factor: 3.390

8.  Mechanical Pull-Out Test of a New Hybrid Fixture-Abutment Connection: An In Vitro Study.

Authors:  Gianmaria D'Addazio; Bruna Sinjari; Lorenzo Arcuri; Beatrice Femminella; Giovanna Murmura; Manlio Santilli; Sergio Caputi
Journal:  Materials (Basel)       Date:  2021-03-22       Impact factor: 3.623

9.  Analysis of Torque Maintenance and Fracture Resistance after Fatigue in Retention Screws Made of Different Metals for Screw-Retained Implant-Borne Prosthesis Joints.

Authors:  Maria Beatriz Bello Taborda; Gabriela Sumie Yaguinuma Gonçalves; Cecília Alves de Sousa; Wirley Gonçalves Assunção
Journal:  Int J Dent       Date:  2021-11-18

10.  Effect of lateral oblique cyclic loading on microleakage and screw loosening of implants with different connections.

Authors:  Hakimeh Siadat; Hossain Najafi; Marzieh Alikhasi; Babak Falahi; Elaheh Beyabanaki; Farid Zayeri
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2018-09-18