Kazunori Hirasawa1, Nobuyuki Shoji1, Yukako Yoshii1, Shota Haraguchi1.
Abstract
PURPOSE: To determine the axial length requiring adjustment of measured circumpapillary retinal nerve fiber layer (cpRNFL) thickness to account for ocular magnification during spectral-domain optical coherence tomography (SD-OCT).Entities:
Mesh:
Year: 2014 PMID: 25215521 PMCID: PMC4162641 DOI: 10.1371/journal.pone.0107553
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Demographic and Ocular Characteristics of the Participants.
| Parameter | Mean ± SD | Range |
| Participants (men/women) | 132 (41/91) | |
| Assessed eye (right/left) | 132 (65/67) | |
| Age (years) | 21.7±1.6 | 20 to 28 |
| Spherical refraction (diopters) | −3.02±3.08 | −13.00 to 4.00 |
| Astigmatism (diopters) | −0.68±0.83 | −6.00 to 0.00 |
| Corneal radius (mm) | 7.79±0.27 | 6.93 to 8.42 |
| Visual acuity (logMAR) | −0.23±0.07 | −0.30 to 0.00 |
| Axial length (mm) | 24.78±1.51 | 21.25 to 28.35 |
| Corneal thickness (µm) | 539.3±24.5 | 472.0 to 595.0 |
| Intraocular pressure (mmHg) | 14.4±2.2 | 7.7 to 20.0 |
SD = standard deviation; logMAR = logarithm of the minimum angle resolution.
Figure 1Correlation of axial length and total cpRNFL thickness measured with and without adjustment for ocular magnification.
(A) Uncorrected total cpRNFL thickness decreased as axial length increased (y = −2.8015x+174.45; R2 = 0.208; p<0.001) whereas (B) the corrected one increased as axial length increased (y = 1.1597x+77.068; R2 = 0.039; p = 0.024).
Measurement Error Values of Total cpRNFL Thickness.
| Adjusted for ocularmagnification | Examiner A (µm) | Examiner B (µm) | Inter-examiner (µm) |
| Yes | 5.2 | 4.4 | 5.3 |
| No | 4.9 | 4.8 | 4.7 |
cpRNFL = circumpapillary retinal nerve fiber layer; Sw = within-subject variation.
The values were calculated as 2.77*Sw.
Results of the ROC Analysis.
| Comparison | Axial lengthcutoff (mm) | AUC | Sensitivity | Specificity | Youden Index |
| Thinner vs. control group | 23.60 | 0.959 | 0.889 | 0.809 | 0.698 |
| Control vs. thicker group | 25.55 | 0.944 | 0.943 | 0.790 | 0.733 |
ROC = receiver operating characteristic; AUC = area under the curve.