Nasrul Anuar Abd Razak1, Noor Azuan Abu Osman, Hossein Gholizadeh, Sadeeq Ali.
Abstract
BACKGROUND: Understanding of kinematics force applied at the elbow is important in many fields, including biomechanics, biomedical engineering and rehabilitation. This paper provides a comparison of a mathematical model of elbow joint using three different types of prosthetics for transhumeral user, and characterizes the forces required to overcome the passive mechanical of the prosthetics at the residual limb.Entities:
Mesh:
Year: 2014 PMID: 25208636 PMCID: PMC4228154 DOI: 10.1186/1475-925X-13-134
Source DB: PubMed Journal: Biomed Eng Online ISSN: 1475-925X Impact factor: 2.819
Figure 1Force exist as elbow become the origin.
Age, weight, transradial mass of each subjects and theoretical force calculated
| Subjects | Age | Weight, W = mg, (g = 9.81) | Transradial mass, m (W × 0.0160) | Force (N) F = mg (Theoretical) |
|---|---|---|---|---|
| 1 | 42 | 62 | 0.992 | 9.73152 |
| 2 | 45 | 73 | 1.168 | 11.45808 |
| 3 | 30 | 59 | 0.944 | 9.26064 |
| 4 | 33 | 77 | 1.232 | 12.08592 |
| 5 | 30 | 80 | 1.28 | 12.5568 |
| 6 | 40 | 72 | 1.152 | 11.30112 |
| 7 | 37 | 62 | 0.992 | 9.73152 |
Figure 2Comparison of free body diagram from three different types of prostheses; B. Body-powered prosthetic, C. Myoelectric prosthetic, D. Air splint prosthetic and A. normal human hand. (S= shoulder, E= elbow, A= arm and P= prosthetic). The forces direction react referring to the x-axis and y-axis.
Result pressure profile applied for each type of prostheses
| Subjects | Body-powered socket pressure, kPa | Myoelectric socket pressure, kPa | Air splint socket pressure, kPa |
|---|---|---|---|
| Area (0.03±0.005 m) | Area (0.045±0.005 m) | Area (0.045±0.005 m) | |
| 1 | 5.78 | 6.02 | 3.22 |
| 2 | 7.47 | 7.45 | 5.24 |
| 3 | 4.32 | 5.72 | 2.91 |
| 4 | 7.35 | 7.62 | 5.61 |
| 5 | 8.21 | 8.01 | 5.97 |
| 6 | 6.58 | 6.48 | 4.93 |
| 7 | 6.21 | 6.09 | 4.23 |
Comparison of force applied by each type of prostheses with the normal hand force
| Subjects | Theoretical force F joint (N) | Body-powered F tension +F shoulder (N) | Myoelectric F socket (N) | Myoelectric F socket (N) |
|---|---|---|---|---|
| (static area) | (dynamic area) | |||
| 1 | 9.73152 | 17.34 | 27.09 | 14.49 |
| 2 | 11.45808 | 22.41 | 33.525 | 23.58 |
| 3 | 9.26064 | 12.96 | 25.74 | 13.095 |
| 4 | 12.08592 | 22.05 | 34.29 | 25.245 |
| 5 | 12.5568 | 24.63 | 36.045 | 26.865 |
| 6 | 11.30112 | 19.74 | 29.16 | 22.185 |
| 7 | 9.73152 | 18.63 | 27.405 | 19.035 |