Literature DB >> 25204662

Comparison between radiation exposure levels using an image intensifier and a flat-panel detector-based system in image-guided central venous catheter placement in children weighing less than 10 kg.

Roberto Miraglia1, Luigi Maruzzelli, Kelvin Cortis, Marcello Piazza, Roberta Gerasia, Simona Maggio, Fabio Tuzzolino, Angelo Luca.   

Abstract

BACKGROUND: Ultrasound-guided central venous puncture and fluoroscopic guidance during central venous catheter (CVC) positioning optimizes technical success and lowers the complication rates in children, and is therefore considered standard practice.
OBJECTIVE: The purpose of this study was to compare the radiation exposure levels recorded during CVC placement in children weighing less than 10 kg in procedures performed using an image intensifier-based angiographic system (IIDS) to those performed in a flat-panel detector-based interventional suite (FPDS).
MATERIALS AND METHODS: A retrospective review of 96 image-guided CVC placements, between January 2008 and October 2013, in 49 children weighing less than 10 kg was performed. Mean age was 8.2 ± 4.4 months (range: 1-22 months). Mean weight was 7.1 ± 2.7 kg (range: 2.5-9.8 kg). The procedures were classified into two categories: non-tunneled and tunneled CVC placement.
RESULTS: Thirty-five procedures were performed with the IIDS (21 non-tunneled CVC, 14 tunneled CVC); 61 procedures were performed with the FPDS (47 non-tunneled CVC, 14 tunneled CVC). For non-tunneled CVC, mean DAP was 113.5 ± 126.7 cGy cm(2) with the IIDS and 15.9 ± 44.6 cGy · cm(2) with the FPDS (P < 0.001). For tunneled CVC, mean DAP was 84.6 ± 81.2 cGy · cm(2) with the IIDS and 37.1 ± 33.5 cGy cm(2) with the FPDS (P = 0.02).
CONCLUSION: The use of flat-panel angiographic equipment reduces radiation exposure in small children undergoing image-guided CVC placement.

Entities:  

Mesh:

Year:  2014        PMID: 25204662     DOI: 10.1007/s00247-014-3119-5

Source DB:  PubMed          Journal:  Pediatr Radiol        ISSN: 0301-0449


  20 in total

1.  Comparison of digital flat-panel detector and conventional angiography machines: evaluation of stent detection rates, visibility scores, and dose-area products.

Authors:  Benjamin Wiesinger; Alexander Stütz; Jörg Schmehl; Claus Detlef Claussen; Jakub Wiskirchen
Journal:  AJR Am J Roentgenol       Date:  2012-04       Impact factor: 3.959

2.  Percutaneous placement of Hickman catheters: comparison of sonographically guided and blind techniques.

Authors:  J S Laméris; P J Post; H M Zonderland; P G Gerritsen; M C Kappers-Klunne; H E Schütte
Journal:  AJR Am J Roentgenol       Date:  1990-11       Impact factor: 3.959

3.  US guidance for vascular access. Technical note.

Authors:  P F Jaques; M A Mauro; B Keefe
Journal:  J Vasc Interv Radiol       Date:  1992-05       Impact factor: 3.464

4.  Radiation doses from venous access procedures.

Authors:  Erik S Storm; Donald L Miller; Laurie Jean Hoover; Jeffrey D Georgia; Tara Bivens
Journal:  Radiology       Date:  2006-01-19       Impact factor: 11.105

Review 5.  Central venous access.

Authors:  Arul Ganeshan; Dinuke R Warakaulle; Raman Uberoi
Journal:  Cardiovasc Intervent Radiol       Date:  2007 Jan-Feb       Impact factor: 2.740

Review 6.  Image Gently, Step Lightly: increasing radiation dose awareness in pediatric interventional radiology.

Authors:  Manrita Sidhu; Brian D Coley; Marilyn J Goske; Bairbre Connolly; John Racadio; Terry T Yoshizumi; Tara Utley; Keith J Strauss
Journal:  Pediatr Radiol       Date:  2009-08-20

Review 7.  Radiation safety in pediatric interventional radiology: Step Lightly.

Authors:  Manrita Sidhu
Journal:  Pediatr Radiol       Date:  2010-04

8.  AAPM/RSNA physics tutorial for residents: physics of flat-panel fluoroscopy systems: Survey of modern fluoroscopy imaging: flat-panel detectors versus image intensifiers and more.

Authors:  Edward Lee Nickoloff
Journal:  Radiographics       Date:  2011 Mar-Apr       Impact factor: 5.333

9.  Radiologically placed tunneled hemodialysis catheters: a single pediatric institutional experience of 120 patients.

Authors:  Melkamu Adeb; Kevin M Baskin; Marc S Keller; Ganesh Krishnamurthy; Els Nijs; Kevin Meyers; Madhura Pradhan; Anne Marie Cahill
Journal:  J Vasc Interv Radiol       Date:  2012-03-27       Impact factor: 3.464

10.  Are routine chest radiographs needed after fluoroscopically guided percutaneous insertion of central venous catheters in children?

Authors:  Hussamuddin Adwan; Hannah Gordon; Eric Nicholls
Journal:  J Pediatr Surg       Date:  2008-02       Impact factor: 2.545

View more
  2 in total

1.  Comparison of image quality and radiation dose between an image-intensifier system and a newer-generation flat-panel detector system — technical phantom measurements and evaluation of clinical imaging in children.

Authors:  Meike Weis; Claudia Hagelstein; Theo Diehm; Stefan O Schoenberg; K Wolfgang Neff
Journal:  Pediatr Radiol       Date:  2016-02

2.  Radiation dose reduction with frame rate conversion in X-ray fluoroscopic imaging systems with flat panel detector: basic study and clinical retrospective analysis.

Authors:  Noriyuki Sakai; Katsuyuki Tabei; Jiro Sato; Toshikazu Imae; Yuichi Suzuki; Shigeharu Takenaka; Keiichi Yano; Osamu Abe
Journal:  Eur Radiol       Date:  2018-07-09       Impact factor: 5.315