Comparison of digital flat-panel detector and conventional angiography machines: evaluation of stent detection rates, visibility scores, and dose-area products.

Benjamin Wiesinger1, Alexander Stütz, Jörg Schmehl, Claus Detlef Claussen, Jakub Wiskirchen.   

Abstract

OBJECTIVE: The objective of this study was to compare the performance and radiation doses of a flat-panel detector (FPD) angiography machine with an image intensifier (II) angiography machine.
MATERIALS AND METHODS: Images of four nitinol stents (Sinus-SuperFlex, SMART, Luminexx, and Zilver stents) in a phantom of a human pelvis were acquired on an FPD system (Axiom Artis) and an II system (Fluorospot TOP) using the following modes: spot-film, continuous fluoroscopy (4, 7.5, 15, and 30 pulses/s), and three amplification modes. Objective stent detection rates and subjective radiopacity scores (scale: 0 [not visible] to 4 [excellent visibility]) were calculated. The radiation doses evaluated by the respective machines were compared.
RESULTS: Over all modes and stents, the mean objective correct stent detection rates and mean subjective radiopacity scores were 89.49% and 1.81, respectively, for the Axiom Artis and 91.00% and 2.26 for the Fluorospot TOP. The stent detection rates over all modes for the SMART and Luminexx stents were better using the Axiom Artis machine (97.61% vs 93.55% and 98.28% vs 90.41%, respectively) and those for the Sinus-SuperFlex and Zilver stents were better using the Fluorospot TOP machine (90.83% vs 83.56% and 89.29% vs 80.50%). The subjective radiopacity scores of stent visibility were worse for the Axiom Artis than the Fluorospot TOP for all stents except the Luminexx stent (mean score, 2.34 vs 2.21, respectively). The objective stent detection rates and subjective radiopacity scores improved using the spot-film mode and with raising amplification, whereas increases in the fluoroscopy pulsing frequency did not improve stent detection rates or radiopacity scores for either machine. The radiation doses at continuous fluoroscopy were approximately 90% higher for the Axiom Artis than for the Fluorospot TOP (2.60 vs 1.41 μGy/m(2) at 30 pulses/s, respectively).
CONCLUSION: The objective correct stent detection rates were similar for both machines with differences in detection for the respective stents. The subjective radiopacity scores were almost always better for the Fluorospot TOP machine. Also, the Axiom Artis machine generated approximately 90% higher radiation doses in fluoroscopy. For both machines, using a higher fluoroscopy pulsing frequency had no positive effect on objective correct stent detection rates or subjective radiopacity scores.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22451565     DOI: 10.2214/AJR.11.7245

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  5 in total

1.  Comparison of image quality and radiation dose between an image-intensifier system and a newer-generation flat-panel detector system — technical phantom measurements and evaluation of clinical imaging in children.

Authors:  Meike Weis; Claudia Hagelstein; Theo Diehm; Stefan O Schoenberg; K Wolfgang Neff
Journal:  Pediatr Radiol       Date:  2016-02

2.  Comparison between radiation exposure levels using an image intensifier and a flat-panel detector-based system in image-guided central venous catheter placement in children weighing less than 10 kg.

Authors:  Roberto Miraglia; Luigi Maruzzelli; Kelvin Cortis; Marcello Piazza; Roberta Gerasia; Simona Maggio; Fabio Tuzzolino; Angelo Luca
Journal:  Pediatr Radiol       Date:  2014-09-10

3.  Investigation of reference levels and radiation dose associated with abdominal EVAR (endovascular aneurysm repair) procedures across several European Centres.

Authors:  E Tuthill; L O'Hora; M O'Donohoe; S Panci; P Gilligan; D Campion; R Trenti; E Fox; D Catania; L Rainford
Journal:  Eur Radiol       Date:  2017-05-18       Impact factor: 5.315

4.  Implantability, Complications, and Follow-Up After Transjugular Intrahepatic Portosystemic Stent-Shunt Creation With the 6F Self-Expanding Sinus-SuperFlex-Visual Stent.

Authors:  Daniel Spira; Jakub Wiskirchen; Ulrich Lauer; Dominik Ketelsen; Konstantin Nikolaou; Benjamin Wiesinger
Journal:  Iran J Radiol       Date:  2016-02-17       Impact factor: 0.212

5.  Transition from image intensifier to flat panel detector in interventional cardiology: Impact of radiation dose.

Authors:  Roshan S Livingstone; David Chase; Anna Varghese; Paul V George; Oommen K George
Journal:  J Med Phys       Date:  2015 Jan-Mar