Sampling strategies for subsampled segmented EPI PRF thermometry in MR guided high intensity focused ultrasound.

Henrik Odéen1, Nick Todd2, Mahamadou Diakite1, Emilee Minalga2, Allison Payne2, Dennis L Parker2.   

Abstract

PURPOSE: To investigate k-space subsampling strategies to achieve fast, large field-of-view (FOV) temperature monitoring using segmented echo planar imaging (EPI) proton resonance frequency shift thermometry for MR guided high intensity focused ultrasound (MRgHIFU) applications.
METHODS: Five different k-space sampling approaches were investigated, varying sample spacing (equally vs nonequally spaced within the echo train), sampling density (variable sampling density in zero, one, and two dimensions), and utilizing sequential or centric sampling. Three of the schemes utilized sequential sampling with the sampling density varied in zero, one, and two dimensions, to investigate sampling the k-space center more frequently. Two of the schemes utilized centric sampling to acquire the k-space center with a longer echo time for improved phase measurements, and vary the sampling density in zero and two dimensions, respectively. Phantom experiments and a theoretical point spread function analysis were performed to investigate their performance. Variable density sampling in zero and two dimensions was also implemented in a non-EPI GRE pulse sequence for comparison. All subsampled data were reconstructed with a previously described temporally constrained reconstruction (TCR) algorithm.
RESULTS: The accuracy of each sampling strategy in measuring the temperature rise in the HIFU focal spot was measured in terms of the root-mean-square-error (RMSE) compared to fully sampled "truth." For the schemes utilizing sequential sampling, the accuracy was found to improve with the dimensionality of the variable density sampling, giving values of 0.65 °C, 0.49 °C, and 0.35 °C for density variation in zero, one, and two dimensions, respectively. The schemes utilizing centric sampling were found to underestimate the temperature rise, with RMSE values of 1.05 °C and 1.31 °C, for variable density sampling in zero and two dimensions, respectively. Similar subsampling schemes with variable density sampling implemented in zero and two dimensions in a non-EPI GRE pulse sequence both resulted in accurate temperature measurements (RMSE of 0.70 °C and 0.63 °C, respectively). With sequential sampling in the described EPI implementation, temperature monitoring over a 192×144×135 mm3 FOV with a temporal resolution of 3.6 s was achieved, while keeping the RMSE compared to fully sampled "truth" below 0.35 °C.
CONCLUSIONS: When segmented EPI readouts are used in conjunction with k-space subsampling for MR thermometry applications, sampling schemes with sequential sampling, with or without variable density sampling, obtain accurate phase and temperature measurements when using a TCR reconstruction algorithm. Improved temperature measurement accuracy can be achieved with variable density sampling. Centric sampling leads to phase bias, resulting in temperature underestimations.

Mesh:

Year:  2014        PMID: 25186406      PMCID: PMC4144162          DOI: 10.1118/1.4892171

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  56 in total

1.  Real-time adaptive methods for treatment of mobile organs by MRI-controlled high-intensity focused ultrasound.

Authors:  Baudouin Denis de Senneville; Charles Mougenot; Chrit T W Moonen
Journal:  Magn Reson Med       Date:  2007-02       Impact factor: 4.668

2.  k-space inherited parallel acquisition (KIPA): application on dynamic magnetic resonance imaging thermometry.

Authors:  Jun-Yu Guo; Eugene G Kholmovski; Ling Zhang; Eun-Kee Jeong; Dennis L Parker
Journal:  Magn Reson Imaging       Date:  2006-04-27       Impact factor: 2.546

3.  Temporally constrained reconstruction of dynamic cardiac perfusion MRI.

Authors:  Ganesh Adluru; Suyash P Awate; Tolga Tasdizen; Ross T Whitaker; Edward V R Dibella
Journal:  Magn Reson Med       Date:  2007-06       Impact factor: 4.668

Review 4.  Ultrasound microbubble contrast agents: fundamentals and application to gene and drug delivery.

Authors:  Katherine Ferrara; Rachel Pollard; Mark Borden
Journal:  Annu Rev Biomed Eng       Date:  2007       Impact factor: 9.590

5.  Multiecho time-resolved acquisition (META): a high spatiotemporal resolution Dixon imaging sequence for dynamic contrast-enhanced MRI.

Authors:  Manojkumar Saranathan; Dan Rettmann; Ersin Bayram; Christine Lee; James Glockner
Journal:  J Magn Reson Imaging       Date:  2009-06       Impact factor: 4.813

6.  Analysis of tissue and arterial blood temperatures in the resting human forearm.

Authors:  H H PENNES
Journal:  J Appl Physiol       Date:  1948-08       Impact factor: 3.531

7.  MRI-compatible transurethral ultrasound system for the treatment of localized prostate cancer using rotational control.

Authors:  Rajiv Chopra; Nicole Baker; Vanessa Choy; Aaron Boyes; Kee Tang; David Bradwell; Michael J Bronskill
Journal:  Med Phys       Date:  2008-04       Impact factor: 4.071

8.  Optimal k-space sampling for dynamic contrast-enhanced MRI with an application to MR renography.

Authors:  Ting Song; Andrew F Laine; Qun Chen; Henry Rusinek; Louisa Bokacheva; Ruth P Lim; Gerhard Laub; Randall Kroeker; Vivian S Lee
Journal:  Magn Reson Med       Date:  2009-05       Impact factor: 4.668

9.  Temporally constrained reconstruction applied to MRI temperature data.

Authors:  Nick Todd; Ganesh Adluru; Allison Payne; Edward V R DiBella; Dennis Parker
Journal:  Magn Reson Med       Date:  2009-08       Impact factor: 4.668

10.  Validation of fast MR thermometry at 1.5 T with gradient-echo echo planar imaging sequences: phantom and clinical feasibility studies.

Authors:  Alexandru Cernicanu; Matthieu Lepetit-Coiffe; Joerg Roland; Christoph D Becker; Sylvain Terraz
Journal:  NMR Biomed       Date:  2008-10       Impact factor: 4.044

View more
  7 in total

1.  Respiration artifact correction in three-dimensional proton resonance frequency MR thermometry using phase navigators.

Authors:  Bryant T Svedin; Allison Payne; Dennis L Parker
Journal:  Magn Reson Med       Date:  2015-08-13       Impact factor: 4.668

2.  Simultaneous MR thermometry and acoustic radiation force imaging using interleaved acquisition.

Authors:  Joshua T de Bever; Henrik Odéen; Lorne W Hofstetter; Dennis L Parker
Journal:  Magn Reson Med       Date:  2017-08-10       Impact factor: 4.668

3.  Near-Infrared Light Increases Functional Connectivity with a Non-thermal Mechanism.

Authors:  Grzegorz M Dmochowski; Ahmed Duke Shereen; Destiny Berisha; Jacek P Dmochowski
Journal:  Cereb Cortex Commun       Date:  2020-03-19

4.  Correcting heat-induced chemical shift distortions in proton resonance frequency-shift thermometry.

Authors:  Pooja Gaur; Ari Partanen; Beat Werner; Pejman Ghanouni; Rachelle Bitton; Kim Butts Pauly; William A Grissom
Journal:  Magn Reson Med       Date:  2015-08-24       Impact factor: 4.668

5.  Evaluation of a three-dimensional MR acoustic radiation force imaging pulse sequence using a novel unbalanced bipolar motion encoding gradient.

Authors:  Joshua T de Bever; Henrik Odéen; Nick Todd; Alexis I Farrer; Dennis L Parker
Journal:  Magn Reson Med       Date:  2015-10-07       Impact factor: 4.668

6.  MR thermometry for focused ultrasound monitoring utilizing model predictive filtering and ultrasound beam modeling.

Authors:  Henrik Odéen; Scott Almquist; Joshua de Bever; Douglas A Christensen; Dennis L Parker
Journal:  J Ther Ultrasound       Date:  2016-09-22

7.  Spatially-segmented undersampled MRI temperature reconstruction for transcranial MR-guided focused ultrasound.

Authors:  Pooja Gaur; Beat Werner; Xue Feng; Samuel W Fielden; Craig H Meyer; William A Grissom
Journal:  J Ther Ultrasound       Date:  2017-05-30
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.