Insertion torque versus mechanical resistance of mini-implants inserted in different cortical thickness.

Renata Faria Santos, Antonio Carlos de Oliveira Ruellas, Daniel Jogaib Fernandes, Carlos Nelson Elias.   

Abstract

OBJECTIVE: This study aimed to measure insertion torque, tip mechanical resistance to fracture and transmucosal neck of mini-implants (MI) (Conexão Sistemas de PróteseT), as well as to analyze surface morphology.
METHODS: Mechanical tests were carried out to measure the insertion torque of MIs in different cortical thicknesses, and tip mechanical resistance to fracture as well as transmucosal neck of MIs. Surface morphology was assessed by scanning electron microscopy (SEM) before and after the mechanical tests.
RESULTS: Values of mechanical resistance to fracture (22.14 N.cm and 54.95 N.cm) were higher and statistically different (P < 0.05) from values of insertion torque for 1-mm (7.60 N.cm) and 2-mm (13.27 N.cm) cortical thicknesses. Insertion torque was statistically similar (P > 0.05) to torsional fracture in the tip of MI (22.14 N.cm) when 3 mm cortical thickness (16.11 N.cm) and dense bone (23.95 N.cm) were used. Torsional fracture of the transmucosal neck (54.95 N.cm) was higher and statistically different (P < 0.05) from insertion torsional strength in all tested situations. SEM analysis showed that the MIs had the same smooth surface when received from the manufacturer and after the mechanical tests were performed. Additionally, no significant marks resulting from the manufacturing process were observed.
CONCLUSION: All mini-implants tested presented adequate surface morphology. The resistance of mini-implants to fracture safely allows placement in 1 and 2-mm cortical thickness. However, in 3-mm cortical thickness and dense bones, pre-drilling with a bur is recommended before insertion.

Entities:  

Keywords:  Orthodontic anchorage procedures; Orthodontics; Torque

Mesh:

Substances:

Year:  2014        PMID: 25162571      PMCID: PMC4296615          DOI: 10.1590/2176-9451.19.3.090-094.oar

Source DB:  PubMed          Journal:  Dental Press J Orthod        ISSN: 2176-9451


  12 in total

1.  Re-use of implant coverscrews changes their surface properties but not clinical outcome.

Authors:  Z Schwartz; C H Lohmann; G Blau; C R Blanchard; A W Soskolne; Y Liu; D L Cochran; D D Dean; B D Boyan
Journal:  Clin Oral Implants Res       Date:  2000-06       Impact factor: 5.977

2.  Load-related implant reaction of mini-implants used for orthodontic anchorage.

Authors:  André Büchter; Dirk Wiechmann; Stefan Koerdt; Hans Peter Wiesmann; Josef Piffko; Ulrich Meyer
Journal:  Clin Oral Implants Res       Date:  2005-08       Impact factor: 5.977

3.  Factors affecting the clinical success of screw implants used as orthodontic anchorage.

Authors:  Hyo-Sang Park; Seong-Hwa Jeong; Oh-Won Kwon
Journal:  Am J Orthod Dentofacial Orthop       Date:  2006-07       Impact factor: 2.650

4.  The use of microimplants in orthodontic anchorage.

Authors:  Chung-Ho Chen; Chao-San Chang; Chi-Hsin Hsieh; Yu-Chuan Tseng; Yee-Shyong Shen; I-Yueh Huang; Chia-Fu Yang; Chun-Ming Chen
Journal:  J Oral Maxillofac Surg       Date:  2006-08       Impact factor: 1.895

5.  Success rate of mini- and micro-implants used for orthodontic anchorage: a prospective clinical study.

Authors:  Dirk Wiechmann; Ulrich Meyer; André Büchter
Journal:  Clin Oral Implants Res       Date:  2007-04       Impact factor: 5.977

6.  Factors associated with the success rate of orthodontic miniscrews placed in the upper and lower posterior buccal region.

Authors:  Cheol-Hyun Moon; Dong-Gun Lee; Hyun-Sun Lee; Jeong-Soo Im; Seung-Hak Baek
Journal:  Angle Orthod       Date:  2008-01       Impact factor: 2.079

Review 7.  Mini-implants in orthodontics: a systematic review of the literature.

Authors:  Reint Reynders; Laura Ronchi; Shandra Bipat
Journal:  Am J Orthod Dentofacial Orthop       Date:  2009-05       Impact factor: 2.650

8.  2008 survey of AAO members on miniscrew usage.

Authors:  Peter H Buschang; Roberto Carrillo; Brian Ozenbaugh; P Emile Rossouw
Journal:  J Clin Orthod       Date:  2008-09

Review 9.  Factors influencing efficiency of sliding mechanics to close extraction space: a systematic review.

Authors:  M Barlow; K Kula
Journal:  Orthod Craniofac Res       Date:  2008-05       Impact factor: 1.826

10.  Comparison of the fracture torque of different Brazilian mini-implants.

Authors:  Guilherme Machado Alvares de Lima; Mário Sérgio Soares; Sibele Sarti Penha; Marcelo Munhóes Romano
Journal:  Braz Oral Res       Date:  2011 Mar-Apr
View more
  1 in total

1.  Effect of vertical placement angle on the insertion torque of mini-implants in human alveolar bone.

Authors:  Rafael Ribeiro Maya; Célia Regina Maio Pinzan-Vercelino; Julio de Araujo Gurgel
Journal:  Dental Press J Orthod       Date:  2016 Sep-Oct
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.