Literature DB >> 25157966

Selecting the correct weighting factors for linear and quadratic calibration curves with least-squares regression algorithm in bioanalytical LC-MS/MS assays and impacts of using incorrect weighting factors on curve stability, data quality, and assay performance.

Huidong Gu1, Guowen Liu, Jian Wang, Anne-Françoise Aubry, Mark E Arnold.   

Abstract

A simple procedure for selecting the correct weighting factors for linear and quadratic calibration curves with least-squares regression algorithm in bioanalytical LC-MS/MS assays is reported. The correct weighting factor is determined by the relationship between the standard deviation of instrument responses (σ) and the concentrations (x). The weighting factor of 1, 1/x, or 1/x(2) should be selected if, over the entire concentration range, σ is a constant, σ(2) is proportional to x, or σ is proportional to x, respectively. For the first time, we demonstrated with detailed scientific reasoning, solid historical data, and convincing justification that 1/x(2) should always be used as the weighting factor for all bioanalytical LC-MS/MS assays. The impacts of using incorrect weighting factors on curve stability, data quality, and assay performance were thoroughly investigated. It was found that the most stable curve could be obtained when the correct weighting factor was used, whereas other curves using incorrect weighting factors were unstable. It was also found that there was a very insignificant impact on the concentrations reported with calibration curves using incorrect weighting factors as the concentrations were always reported with the passing curves which actually overlapped with or were very close to the curves using the correct weighting factor. However, the use of incorrect weighting factors did impact the assay performance significantly. Finally, the difference between the weighting factors of 1/x(2) and 1/y(2) was discussed. All of the findings can be generalized and applied into other quantitative analysis techniques using calibration curves with weighted least-squares regression algorithm.

Mesh:

Substances:

Year:  2014        PMID: 25157966     DOI: 10.1021/ac5018265

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  30 in total

1.  Development and validation of a UPLC-MS/MS method for the simultaneous determination of paritaprevir and ritonavir in rat liver.

Authors:  Andrew J Ocque; Colleen E Hagler; Robin Difrancesco; Yvonne Woolwine-Cunningham; Cindy J Bednasz; Gene D Morse; Andrew H Talal
Journal:  Bioanalysis       Date:  2016-06-09       Impact factor: 2.681

2.  Quantitative and Qualitative Method for Sphingomyelin by LC-MS Using Two Stable Isotopically Labeled Sphingomyelin Species.

Authors:  Kotaro Hama; Yuko Fujiwara; Kazuaki Yokoyama
Journal:  J Vis Exp       Date:  2018-05-07       Impact factor: 1.355

3.  Development of cGAMP-Luc, a sensitive and precise coupled enzyme assay to measure cGAMP in complex biological samples.

Authors:  Rachel E Mardjuki; Jacqueline A Carozza; Lingyin Li
Journal:  J Biol Chem       Date:  2020-03-03       Impact factor: 5.157

4.  Interference filter tilting to detect a polycyclic aromatic hydrocarbon at the second harmonic of wavelength modulation frequency.

Authors:  Dmitri Lanevski; Koit Mauring; Eric Tkaczyk
Journal:  Appl Opt       Date:  2017-04-10       Impact factor: 1.980

5.  Is it possible to detect PEth 16:0/18:1 and PEth 18:1/18:1 in red blood cells after 20 years of storage in liquid nitrogen?

Authors:  Robert Boll; Theron Johnson; Rudolf Kaaks; Tilman Kühn; Gisela Skopp
Journal:  Int J Legal Med       Date:  2017-06-24       Impact factor: 2.686

6.  Untargeted and targeted analysis of sarin poisoning biomarkers in rat urine by liquid chromatography and tandem mass spectrometry.

Authors:  M F Vokuev; Т М Baygildiev; I V Plyushchenko; Y A Ikhalaynen; R L Ogorodnikov; I K Solontsov; А V Braun; E I Savelieva; I V Rуbalchenko; I A Rodin
Journal:  Anal Bioanal Chem       Date:  2021-09-21       Impact factor: 4.142

7.  Validation and Application of a Simple UHPLC-MS-MS Method for the Enantiospecific Determination of Warfarin in Human Urine.

Authors:  Osama Y Alshogran; Andrew J Ocque; François A Leblond; Vincent Pichette; Thomas D Nolin
Journal:  J Chromatogr Sci       Date:  2015-12-11       Impact factor: 1.618

8.  Comparison of endogenous GHB concentrations in blood and hair in death cases with emphasis on the post mortem interval.

Authors:  André L Castro; Sónia Tarelho; Mário Dias; Flávio Reis; Helena M Teixeira
Journal:  Int J Legal Med       Date:  2016-02-04       Impact factor: 2.686

9.  Pre-clinical pharmacokinetic and pharmacodynamic modelling study of 4-hydroxyisoleucine using validated ultra-performance liquid chromatography-tandem mass spectrometry.

Authors:  Geetika Wadhwa; Kowthavarapu Venkata Krishna; Rajeev Taliyan; Neeraj Tandon; Satyapal Singh Yadav; Chandrakant Katiyar; Sunil Kumar Dubey
Journal:  RSC Adv       Date:  2020-02-04       Impact factor: 4.036

10.  Quantification of Nonpersistent Pesticides in Small Volumes of Human Breast Milk with Ultrahigh Performance Liquid Chromatography Coupled to Tandem Mass Spectrometry.

Authors:  Theresa L Pedersen; Jennifer T Smilowitz; Carl K Winter; Shiva Emami; Rebecca J Schmidt; Deborah H Bennett; Irva Hertz-Picciotto; Ameer Y Taha
Journal:  J Agric Food Chem       Date:  2021-06-08       Impact factor: 5.895

View more