Evaluation of Beckman Coulter DxI 800 immunoassay system using clinically oriented performance goals.

Neval Akbas1, Patricia G Schryver2, Alicia Algeciras-Schimnich1, Nikola A Baumann1, Darci R Block1, Jeffrey R Budd3, S J Stephen Gaston3, George G Klee4.   

Abstract

OBJECTIVES: We evaluated the analytical performance of 24 immunoassays using the Beckman Coulter DxI 800 immunoassay systems at Mayo Clinic, Rochester, MN for trueness, precision, detection limits, linearity, and consistency (across instruments and reagent lots).
METHODS: Clinically oriented performance goals were defined using the following methods: trueness-published desirable accuracy limits, precision-published desirable biologic variation; detection limits - 0.1 percentile of patient test values, linearity - 50% of total error, and consistency-percentage test values crossing key decision points. Local data were collected for precision, linearity, and consistency. Data were provided by Beckman Coulter, Inc. for trueness and detection limits.
RESULTS: All evaluated assays except total thyroxine were within the proposed goals for trueness. Most of the assays met the proposed goals for precision (86% of intra-assay results and 75% of inter-assay results). Five assays had more than 15% of the test results below the minimum detection limits. Carcinoembryonic antigen, total thyroxine and free triiodothyronine exceeded the proposed goals of ±6.3%, ±5% and ±5.7% for dilution linearity. All evaluated assays were within the proposed goals for instrument consistency. Lot-to-lot consistency results for cortisol, ferritin and total thyroxine exceeded the proposed goals of 3.3%, 11.4% and 7% at one medical decision level, while vitamin B12 exceeded the proposed goals of 5.2% and 3.8% at two decision levels.
CONCLUSIONS: The Beckman Coulter DxI 800 immunoassay system meets most of these proposed goals, even though these clinically focused performance goals represent relatively stringent limits.
Copyright © 2014 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Analytical performance; Beckman Coulter DxI 800; Immunoassay; Precision; Quality control

Mesh:

Substances:

Year:  2014        PMID: 25130957     DOI: 10.1016/j.clinbiochem.2014.08.005

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  3 in total

1.  Evaluation of Analytical Performances of Magnetic Force-Assisted Electrochemical Sandwich Immunoassay for the Quantification of Carcinoembryonic Antigen.

Authors:  Boo Young Hwang; Eunsoo Kim; Seung-Ha Kim; Hyundoo Hwang
Journal:  Front Bioeng Biotechnol       Date:  2022-01-03

Review 2.  Performance and comparability of laboratory methods for measuring ferritin concentrations in human serum or plasma: A systematic review and meta-analysis.

Authors:  Maria N Garcia-Casal; Juan P Peña-Rosas; Eloisa Urrechaga; Jesus F Escanero; Junsheng Huo; Ricardo X Martinez; Lucero Lopez-Perez
Journal:  PLoS One       Date:  2018-05-03       Impact factor: 3.240

3.  Comparison of Four Automated Carcinoembryonic Antigen Immunoassays: ADVIA Centaur XP, ARCHITECT I2000sr, Elecsys E170, and Unicel Dxi800.

Authors:  Joonhong Park; Seungok Lee; Yeongsic Kim; Aeran Choi; Hyeyoung Lee; Jihyang Lim; Yonggoo Kim; Kyungja Han; Eun Jee Oh
Journal:  Ann Lab Med       Date:  2018-07       Impact factor: 3.464

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.