Literature DB >> 25086796

Radiation exposure to hand surgeons' hands: a practical comparison of large and mini C-arm fluoroscopy.

Michael M Vosbikian1, Asif M Ilyas2, Derek D Watson2, Charles F Leinberry2.   

Abstract

PURPOSE: To determine and compare the radiation exposure to surgeons' hands with large and mini C-arm fluoroscopy in a practical, clinically based model.
METHODS: Two hand surgeons monitored radiation exposure to their hands with a ring dosimeter over a 14-month period using large and mini C-arm fluoroscopic units. One surgeon performed all cases with a large C-arm unit in a hospital setting, and the other performed all cases with mini C-arms in surgical centers. For each case, fluoroscopic time, the output displayed by the unit, radiation by time, and ring dosimeter absorption were recorded and analyzed.
RESULTS: A total of 160 consecutive cases were reviewed with 71 cases and 89 cases in the large and mini C-arm groups, respectively. The median output displayed by the large C-arm was 0.7 mGy/case, and the median output displayed by the mini C-arm was 10.0 mGy/case. With output as a product of time, the median calculated values were 0.02 mGy/s for the large C-arm group and 0.28 mGy/s for the mini C-arm group. Cumulative ring dosimeter absorption to the surgeons' hands was found to be 380 mrem for 71 cases in the large C-arm group versus 1,000 mrem for 89 cases in the mini C-arm group.
CONCLUSIONS: In our model, the use of the mini C-arm resulted in more than a 10-fold increase in the rate of output and approximately double the dosimeter absorption to the surgeon's hand compared with the large C-arm. Although it has been shown that the mini C-arm produces less radiation scatter, in a practical model, it may not be a safer alternative with respect to the surgeon's hands. Based on these findings, we recommend that surgeons be more aware of radiation exposure risk, know their C-arm unit's specifications, and try to minimize radiation exposure. TYPE OF STUDY/LEVEL OF EVIDENCE: Prognostic II.
Copyright © 2014 American Society for Surgery of the Hand. Published by Elsevier Inc. All rights reserved.

Keywords:  Fluoroscopy; large C-arm; mini C-arm; radiation; radiation exposure

Mesh:

Year:  2014        PMID: 25086796     DOI: 10.1016/j.jhsa.2014.06.133

Source DB:  PubMed          Journal:  J Hand Surg Am        ISSN: 0363-5023            Impact factor:   2.230


  11 in total

1.  Mini C-Arm Fluoroscopy: Does Its Configuration Matter for Radiation Exposure to the Surgeon?

Authors:  Talia Chapman; Dennis P Martin; Christopher Williamson; Brian Tinsley; Mark L Wang; Asif M Ilyas
Journal:  Hand (N Y)       Date:  2017-06-23

2.  Hand Surgery and Fluoroscopic Eye Radiation Dosage: A Prospective Pilot Comparison of Large Versus Mini C-Arm Fluoroscopy Use.

Authors:  Mark L Wang; C Edward Hoffler; Asif M Ilyas; William H Kirkpatrick; Pedro K Beredjiklian; Charles F Leinberry
Journal:  Hand (N Y)       Date:  2016-04-06

3.  Elevated Radiation Exposure Associated With Above Surface Flat Detector Mini C-Arm Use.

Authors:  Dennis P Martin; Talia Chapman; Christopher Williamson; Brian Tinsley; Asif M Ilyas; Mark L Wang
Journal:  Hand (N Y)       Date:  2017-11-22

4.  Are Indian orthopaedic surgeons aware of the health hazards of radiation exposure? A survey and review on awareness and ways to mitigate them.

Authors:  Binoti Sheth; Arzan Jesia; Rohan Parihar; Rohit Somani; Pankaj Pawar
Journal:  J Clin Orthop Trauma       Date:  2022-08-07

Review 5.  Are we putting ourselves in danger? Occupational hazards and job safety for orthopaedic surgeons.

Authors:  Robert C Ryu; Phillip H Behrens; Azeem T Malik; Jonathan D Lester; Christopher S Ahmad
Journal:  J Orthop       Date:  2021-02-20

6.  Assessment of Radiation Protection in Hand-Shielding Products With Mini C-Arm Fluoroscopy.

Authors:  Matthew B Cantlon; Asif M Ilyas
Journal:  Hand (N Y)       Date:  2019-08-12

Review 7.  Intraoperative radiation safety in orthopaedics: a review of the ALARA (As low as reasonably achievable) principle.

Authors:  Daniel J Kaplan; Jay N Patel; Frank A Liporace; Richard S Yoon
Journal:  Patient Saf Surg       Date:  2016-12-12

8.  Reduction of operator radiation exposure using a passive robotic device during fluoroscopy-guided arterial puncture: an experimental study in a swine model.

Authors:  Muhammad Umair Ahmad Khan; Chang-Hwan Yoon; Byung-Ju Yi
Journal:  Eur Radiol Exp       Date:  2019-05-29

9.  Evaluation of radiation exposure dose at double-balloon endoscopy for the patients with small bowel disease.

Authors:  Asuka Nagura; Masanao Nakamura; Osamu Watanabe; Takeshi Yamamura; Kohei Funasaka; Eizaburo Ohno; Ryoji Miyahara; Hiroki Kawashima; Shuji Koyama; Tomoki Hinami; Hidemi Goto; Yoshiki Hirooka
Journal:  Nagoya J Med Sci       Date:  2016-08       Impact factor: 1.131

10.  Reduction of radiation exposure to operating physician and assistant using a real-time auditory feedback dosimeter during femoral artery puncturing: a study on swine model.

Authors:  Muhammad Umair Ahmad Khan; Byung-Ju Yi
Journal:  Eur Radiol Exp       Date:  2019-09-23
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.