Development of a novel depth of interaction PET detector using highly multiplexed G-APD cross-strip encoding.

A Kolb1, C Parl1, F Mantlik2, C C Liu1, E Lorenz3, D Renker4, B J Pichler1.   

Abstract

PURPOSE: The aim of this study was to develop a prototype PET detector module for a combined small animal positron emission tomography and magnetic resonance imaging (PET/MRI) system. The most important factor for small animal imaging applications is the detection sensitivity of the PET camera, which can be optimized by utilizing longer scintillation crystals. At the same time, small animal PET systems must yield a high spatial resolution. The measured object is very close to the PET detector because the bore diameter of a high field animal MR scanner is limited. When used in combination with long scintillation crystals, these small-bore PET systems generate parallax errors that ultimately lead to a decreased spatial resolution. Thus, we developed a depth of interaction (DoI) encoding PET detector module that has a uniform spatial resolution across the whole field of view (FOV), high detection sensitivity, compactness, and insensitivity to magnetic fields.
METHODS: The approach was based on Geiger mode avalanche photodiode (G-APD) detectors with cross-strip encoding. The number of readout channels was reduced by a factor of 36 for the chosen block elements. Two 12 × 2 G-APD strip arrays (25 μm cells) were placed perpendicular on each face of a 12 × 12 lutetium oxyorthosilicate crystal block with a crystal size of 1.55 × 1.55 × 20 mm. The strip arrays were multiplexed into two channels and used to calculate the x, y coordinates for each array and the deposited energy. The DoI was measured in step sizes of 1.8 mm by a collimated (18)F source. The coincident resolved time (CRT) was analyzed at all DoI positions by acquiring the waveform for each event and applying a digital leading edge discriminator.
RESULTS: All 144 crystals were well resolved in the crystal flood map. The average full width half maximum (FWHM) energy resolution of the detector was 12.8% ± 1.5% with a FWHM CRT of 1.14 ± 0.02 ns. The average FWHM DoI resolution over 12 crystals was 2.90 ± 0.15 mm.
CONCLUSIONS: The novel DoI PET detector, which is based on strip G-APD arrays, yielded a DoI resolution of 2.9 mm and excellent timing and energy resolution. Its high multiplexing factor reduces the number of electronic channels. Thus, this cross-strip approach enables low-cost, high-performance PET detectors for dedicated small animal PET and PET/MRI and potentially clinical PET/MRI systems.

Entities:  

Mesh:

Year:  2014        PMID: 25086547     DOI: 10.1118/1.4890609

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Evaluation of Matrix9 silicon photomultiplier array for small-animal PET.

Authors:  Junwei Du; Jeffrey P Schmall; Yongfeng Yang; Kun Di; Emilie Roncali; Gregory S Mitchell; Steve Buckley; Carl Jackson; Simon R Cherry
Journal:  Med Phys       Date:  2015-02       Impact factor: 4.071

Review 2.  Quantitative Rodent Brain Receptor Imaging.

Authors:  Kristina Herfert; Julia G Mannheim; Laura Kuebler; Sabina Marciano; Mario Amend; Christoph Parl; Hanna Napieczynska; Florian M Maier; Salvador Castaneda Vega; Bernd J Pichler
Journal:  Mol Imaging Biol       Date:  2020-04       Impact factor: 3.488

3.  A depth-of-interaction encoding PET detector module with dual-ended readout using large-area silicon photomultiplier arrays.

Authors:  Junwei Du; Xiaowei Bai; Simon R Cherry
Journal:  Phys Med Biol       Date:  2018-12-14       Impact factor: 3.609

4.  H2RSPET: a 0.5 mm resolution high-sensitivity small-animal PET scanner, a simulation study.

Authors:  Youfang Lai; Qian Wang; Shiwei Zhou; Zhaoheng Xie; Jinyi Qi; Simon R Cherry; Mingwu Jin; Yujie Chi; Junwei Du
Journal:  Phys Med Biol       Date:  2021-03-09       Impact factor: 3.609

5.  Development and initial characterization of a high-resolution PET detector module with DOI.

Authors:  Mohan Li; Yuli Wang; Shiva Abbaszadeh
Journal:  Biomed Phys Eng Express       Date:  2020-10-14
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.