A theoretical comparison of tissue parameter extraction methods for dual energy computed tomography.

Jean-Étienne Tremblay1, Stéphane Bedwani1, Hugo Bouchard2.   

Abstract

PURPOSE: To evaluate the reliability of common sinogram-based DECT reconstruction methods for radiotherapy tissue characterization and to evaluate the advantage of combining them with a stoichiometric calibration.
METHODS: The sinogram-based DECT method defined byAlvarez and Macovski ["Energy-selective reconstructions in x-ray computerized tomography," Phys. Med. Biol. 21, 733-744 (1976)] is adapted to the XCOM photon cross sections database and also generalized to a two-material decomposition method. A theoretical framework is developed using a test phantom containing human tissue compositions for comparing the sinogram-based methods and the calibration-based method, being defined as the application of the stoichiometric calibration technique of Bourque et al. ["A stoichiometric calibration method for dual energy computed tomography," Phys. Med. Biol. 59, 2059-2088 (2014)] on monoenergetic images being generated with a sinogram-based method. Applying a bias correction to the sinogram-based method, its performance in extracting human tissue parameters in the presence of noise as well as by altering the photon energy spectrum is compared to the calibration-based method.
RESULTS: In the absence of noise and without spectrum alteration, the calibration-based method is found to have no benefit on the sinogram-based method. However, the calibration-based method is shown to be potentially more reliable than bias-corrected sinogram-based methods in situations comparable to the clinical environment, where noise is present and the photon energy spectra can differ from what is used during image reconstruction. In determining electron density, the performance of all methods is comparable in the presence of noise only. Moreover, combined with heavy spectrum alteration, the mean errors on electron density are found higher in sinogram-based methods in comparison with the calibration-based method, with 1.2% versus 0.2%. In the presence of significant noise, bias-corrected sinogram-based methods yield mean errors on effective atomic number of about 2.5%, as compared to 0.5% for the calibration-based method. When combined with heavy spectrum alteration, bias-corrected sinogram-based methods can lead to error of up to 4% on the effective atomic number versus 1.8% for the calibration-based method.
CONCLUSIONS: While sinogram-based methods have the advantage of eliminating beam hardening effects, results of this study suggest improvements in the accuracy and reliability of extracting tissue parameters by applying the DECT stoichiometric calibration of Bourqueet al. to monoenergetic images being generated with such DECT reconstruction methods.

Entities:  

Mesh:

Year:  2014        PMID: 25086536     DOI: 10.1118/1.4886055

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  The effect of iodine uptake on radiation dose absorbed by patient tissues in contrast enhanced CT imaging: Implications for CT dosimetry.

Authors:  Kostas Perisinakis; Antonis Tzedakis; Kostas Spanakis; Antonios E Papadakis; Adam Hatzidakis; John Damilakis
Journal:  Eur Radiol       Date:  2017-07-14       Impact factor: 5.315

Review 2.  Status and innovations in pre-treatment CT imaging for proton therapy.

Authors:  Patrick Wohlfahrt; Christian Richter
Journal:  Br J Radiol       Date:  2019-11-11       Impact factor: 3.039

3.  Accuracy of virtual monochromatic images generated by the decomposition of photoelectric absorption and Compton scatter in dual-energy computed tomography.

Authors:  Kazuhiro Sato; Chifumi Sato; Ayami Takahashi; Hirokazu Takano; Shingo Kayano; Ayana Ishiguro; Yumi Takane; Tomohiro Kaneta
Journal:  Phys Eng Sci Med       Date:  2022-01-28

4.  One-step iterative reconstruction approach based on eigentissue decomposition for spectral photon-counting computed tomography.

Authors:  Mikaël Simard; Hugo Bouchard
Journal:  J Med Imaging (Bellingham)       Date:  2022-07-27

5.  Projection decomposition via univariate optimization for dual-energy CT.

Authors:  Wenxiang Cong; Bruno De Man; Ge Wang
Journal:  J Xray Sci Technol       Date:  2022       Impact factor: 2.442

6.  On the equivalence of image-based dual-energy CT methods for the determination of electron density and effective atomic number in radiotherapy.

Authors:  Christian Möhler; Patrick Wohlfahrt; Christian Richter; Steffen Greilich
Journal:  Phys Imaging Radiat Oncol       Date:  2018-04-11
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.