Clinical outcomes after cataract surgery with implantation of the Tecnis ZMB00 multifocal intraocular lens.

Wojciech Lubiński1, Jolanta Gronkowska-Serafin1, Karolina Podborączyńska-Jodko1.   

Abstract

BACKGROUND: The aim of this study was to evaluate visual performance, contrast sensitivity, and patient satisfaction in patients undergoing cataract surgery with bilateral implantation of the Tecnis ZMB00 diffractive multifocal IOL (intraocular lens). MATERIAL/
METHODS: This was a prospective study of 40 eyes of 20 patients with an age range from 48 to 67 years and undergoing cataract surgery with implantation of the diffractive 1-piece IOL Tecnis ZMB00 (Abbott Medical Optics) in 1 eye and 3 weeks later in the other eye. The following parameters were evaluated at 3 and 6 months after the operation: binocular uncorrected distance, intermediate and near visual acuity (UDVA, UIVA, UNVA), uncorrected binocular photopic and mesopic distance and photopic near contrast sensitivity (CSV-1000), subjective symptoms, and patient satisfaction (VF-14).
RESULTS: No significant change was observed in logMAR UDVA between 3 and 6 months postoperatively (-0.11±0.14 vs. -0.10±0.13, p>0.05). In contrast, UNVA (0.06±0.12 vs. -0.02±0.12, p=0.004) and UIVA (0.12±0.15 vs. 0.07±0.11, p=0.005) in this period improved significantly. At 3 and 6 months after surgery, 85% of patients no longer needed to wear corrective lenses. Contrast sensitivity under different conditions was within normal age-matched limits, with significant improvements for some spatial frequencies at 3 and 6 months after surgery (p<0.04). Mean overall patient satisfaction was 9.39±1.06 and 9.19±1.20 (scale from 1 to 10, with 10 being the best score) at 3 and 6 months, respectively. Low level of halo perception was reported in 75% of patients.
CONCLUSIONS: The Tecnis ZMB00 IOL provides an effective restoration of the distance, intermediate, and near visual function, allowing patients to be totally free of need to wear corrective lenses and providing high levels of patient satisfaction.

Entities:  

Mesh:

Year:  2014        PMID: 25022700      PMCID: PMC4106930          DOI: 10.12659/MSM.890585

Source DB:  PubMed          Journal:  Med Sci Monit        ISSN: 1234-1010


Background

The goal of modern cataract surgery is not only to remove the opacified lens, but also to restore visual function at various distances. This can be achieved by implanting specific models of multifocal intraocular lenses (IOLs). Refractive, diffractive, and hybrid apodized (refractive/diffractive) multifocal IOLs are currently available for clinical use [1,2]. One relatively new design of multifocal IOL is the 1-piece IOL Tecnis ZMB00 (Abbott Medical Optics), which combines diffractive and aspheric optics. This IOL is CE-marked and obtained FDA approval in 2010. Specifically, the aspheric surface of this IOL induces a controlled amount of negative spherical aberration that compensates for the positive spherical aberration usually present in the cornea. Furthermore, chromatic dispersion induced by this relatively new IOL is low and can result in an improvement in contrast sensitivity of up to 12% in comparison with other IOLs made of hydrophobic acrylic materials (e.g., Zhao H, Piers PA, Mainster MA). The additive effects of different optical design elements contribute to contrast loss in pseudophakic eyes implanted with different aspheric IOLs (presented at the XXVII Congress of the ESCRS, Barcelona, 2009). To date, the results of 2 reported studies have shown that this type of multifocal IOL is able to provide excellent objective and subjective results, with effective restoration of near and distance visual function [3,4]. The aim of the current study was to evaluate binocular visual outcomes, including the analysis of intermediate visual function, with this diffractive multifocal IOL at 3 and 6 months after surgery.

Material and Methods

Patient selection

The study included 40 eyes of 20 patients (16 females, 4 males), with a mean age of 56.10±6.8 years (range 48–67 years) undergoing cataract surgery with implantation of the diffractive 1-piece IOL Tecnis ZMB00 in 1 eye and 3 weeks later in the other eye. Patients were included if they were between 40 and 70 years old and had bilateral cataracts, preoperative corneal astigmatism of less than 1.0 D, strong motivation for independence from wearing corrective lenses, and who agreed to attend the scheduled follow-up visits. Exclusion criteria included subjects under 40 or over 70 years of age, with unrealistic visual outcome expectations or with a profession demanding visual precision (e.g., an architect), psychiatric diseases, stroke, dyslexia, dissatisfaction with progressive glasses, or the need for an IOL power beyond the available diopter range (+5.0 to +34 D). This study was approved by the local ethics committee and was performed in accordance with the ethical standards laid down in the Declaration of Helsinki.

Preoperative and postoperative examination

Before surgery, all patients had a comprehensive ophthalmological examination, including uncorrected and best corrected visual acuity, subjective refraction, corneal topography (Corneal Videokeratography Zeiss), slitlamp biomicroscopy, Goldman tonometry, biometry (IOLMaster500, Carl Zeiss Meditec AG), and binocular indirect ophthalmoscopy through a dilated pupil. At 3 and 6 months after surgery, the clinical evaluation included the following tests: binocular uncorrected distance visual acuity (UDVA) (logMAR ETDRS chart at 4 m), uncorrected near visual acuity (UNVA) (logMAR at 35 cm), uncorrected intermediate visual acuity (UIVA) (logMAR at 60 cm), manifest refraction, binocular photopic (85 cd/m2) and mesopic (3 cd/m2) distance (2.5 m), binocular photopic near (35 cm) contrast sensitivity (1.5, 3, 6, 12, and 18 cycles/degree, CSV-1000, FACT), screening stereoscopic test (Lang Stereotest II), subjective symptoms (halo, glare), and patient satisfaction evaluation (Visual Function questionnaire VF-14) [5].

Surgical technique

All surgeries were performed by the same surgeon (WL) under general anaesthesia through a 2.8-mm incision. Continuous curvilinear capsulorrhexis of approximately 5 mm of diameter was performed. After cataract removal, the Tecnis ZMB00 IOL was inserted into the capsular bag by using the Emerald AR Unfolder (Abbott Medical Optics, Santa Ana, CA). The IOL power was calculated using optical biometry (IOLMaster, Carl Zeiss-Meditec, Jena, Germany), the SRK-T formula, and the A-constant recommended by the manufacturer (118.8). Target refraction was emmetropia in all cases.

Statistical analysis

Statistica software (IBM, Armonk, NY, USA) was used for statistical analysis. All of the data samples analyzed followed a normal distribution according to the results of the Kolmogorov-Smirnov test. Therefore, non-parametric statistical tests were used. Specifically, the Wilcoxon rank sum test was used to assess the significance of differences between 3-month and 6-month postoperative visual, contrast sensitivity, and patient satisfaction data, using the same level of significance (p<0.05) in all cases.

Results

Preoperatively, 13 eyes were hyperopic with a spherical equivalent (SE) ranging from +1.00 to + 3.50 D, a mean value of +2.04±0.74 D, and a median of 2.00 D. Ten eyes were myopic with an SE range from −1.00 to −7.00 D, a mean value of −3.73±1.90 D, and a median of −4.0 D. The remaining 17 eyes presented a SE of 0.00 D. For the whole sample, mean and median preoperative SE were −0.26±2.40 D and 0.00 D, respectively. Mean preoperative binocular logMAR UDVA was 0.43±0.28 and mean binocular logMAR corrected distance visual acuity (CDVA) was 0.12±0.17. At 3 and 6 months after surgery, mean SE was 0.20 D for distance and near vision. No significant improvement of binocular UDVA was observed between 3 months and 6 months postoperatively (logMAR −0.11±0.14 vs. −0.10±0.13, p=NS). In contrast, a small but statistically significant improvement of binocular UNVA was detected (0.06±0.12 vs. −0.02±0.12, p= 0.004) (Table 1).
Table 1

Binocular 3 and 6-month mean UDVA, UNVA, UIVA, and CIVA (logMAR).

Parameter3 months6 monthsp-value
UDVA−0.11±0.14−0.10±0.13NS
UNVA0.06±0.12−0.02±0.120.004
UIVA0.12±0.150.07±0.110.004
CIVA0.00±0.05−0.06±0.060.005

NS – statistically not significant.

For intermediate vision, 2 subjects needed to wear corrective lenses ranging from +1.00 D to +1.37 D at the 3-month follow-up visit, and 6 months after surgery 2 subjects needed to ear corrective lenses ranging from +1.37 D to +1.50 D. Seven subjects achieved a corrected intermediate visual acuity (CIVA) of 0.00 logMAR and 19 subjects achieved a logMAR CIVA of 0.10. A binocular logMAR CIVA of 0.00 was achieved in 35% (7/20) of patients. Mean logMAR CIVA was 0.00±0.05 and −0.06±0.06 at 3 and 6 months after surgery, respectively. This change was statistically significant (p=0.005). At 3 months postoperatively, mean binocular UIVA was 0.12±0.15. Binocular logMAR UIVA was 0.10 or better in 65% (13/20) of patients. At 6 months postoperatively, logMAR UIVA improved significantly (p=0.004), achieving a mean binocular value of 0.07±0.11. The percentage of eyes achieving a logMAR UIVA of 0.10 or better was the same as that obtained at 3 months postoperatively.

Corrective lenses independence outcomes

At 3 and 6 months after surgery, 85% of patients (17/20) were totally free of the need to wear corrective lenses. The best results of corrective lenses independence were obtained for distance (94%–95%), followed by near (85%–90%) and intermediate vision (88%–90%).

Contrast sensitivity (CS) outcomes

At 3 and 6 months postoperatively, photopic and mesopic CS for distance and photopic CS for near were found to be within normal limits for the normal population in the range of 50 to 75 years [6] (Figure 1A–1C, Table 2). At 6 months postoperatively, significant improvements of mesopic CS for distance (3 cycles/º: 1.76±0.09 vs. 1.83±0.08, p<0.03) and photopic CS for near (1.5 cycles/º: 1.76±0.07 vs. 1.82±0.07, p<0.04; 3 cycles/º: 1.80±0.06 vs. 1.86±0.05, p<0.02; 6 cycles/º: 1.69±0.12 vs. 1.78±0.12, p<0.01) were observed for some spatial frequencies (Table 2).
Figure 1

Mean uncorrected binocular distance photopic (A) and mesopic (B) CS, as well as mean photopic near CS (C), at 3 and 6 months after surgery in comparison to the normal values for a population aged 50–75 years (continuous line).

Table 2

Mean uncorrected binocular photopic, mesopic distance and photopic near contrast sensitivity. Comparison of results 3 and 6 month postoperatively.

3 months6 monthsp value
Photopic distance
 3 cpd1.84±0.101.87±0.07NS
 6 cpd1.89±0.121.93±0.13NS
 12 cpd1.56±0.181.53±0.15NS
 18 cpd0.99±0.191.04±0.17NS
Mesopic distance
 3 cpd1.76±0.091.83±0.08<0.03
 6 cpd1.84±0.121.89±0.14NS
 12 cpd1.42±0.181.47± 0.13NS
 18 cpd0.97±0.270.95±0.15NS
Photopic near
 1.5 cpd1.76±0.071.82±0.07<0.04
 3 cpd1.80±0.061.86±0.05<0.02
 6 cpd1.69±0.121.78±0.12<0.01
 12 cpd1.40±0.121.54±0.11NS
 18 cpd1.04±0.161.17±0.17NS

NS – statistically not significant.

Patient satisfaction outcomes

No significant changes were observed in the general vision satisfaction score between 3 and 6 months after surgery (9.39±1.06 vs. 9.19±1.20, p=NS, scale ranging from 0 [not satisfied at all] to 10 [completely satisfied]). Furthermore, no significant changes between 3 and 6 months postoperatively were detected in the scores for the different aspects of visual function evaluated with the VF-14 test (Table 3). Patients had mild or no difficulties in performing the different activities evaluated with the VF-14 questionnaire (Table 3). Driving at night was the only activity during which patients experienced little to moderate difficulties.
Table 3

Visual function test (VF-14); comparison of 3 and 6 month results.

Activity3 months6 monthsp value
1. Reading small print1.36±0.681.38±0.84NS
2. Reading a newspaper or book1.31±0.671.38±0.69NS
3. Reading a large-print book1.05±0.231.00±0.00NS
4. Recognizing people close by1.00±0.001.00±0.00NS
5. Seeing steps, stairs or curbs1.05±0.221.05±0.23NS
6. Reading traffic signs, street names or store signs1.16±0.501.17±0.51NS
7. Doing fine handicrafts1.26±0.561.17±0.51NS
8. Writing cheques or filling out forms1.16±0.371.11±0.47NS
9. Playing cards1.00±0.001.00±0.00NS
10. Doing sport activities1.00±0.001.00±0.00NS
11. Cooking1.10±0.311.00±0.00NS
12. Watching television1.26±0.451.00±0.00NS
13. Driving during the day1.18±0.401.22±0.44NS
14. Driving at night2.54±1.132.22±1.30NS

(Scale: no difficulties – 1, little difficulties – 2, moderate difficulties – 3, big difficulties – 4, unable to perform activity – 5).

Regarding photic phenomena, a significant reduction in halo-related difficulties at work, as well as in the level of halo perception, was noted at 6 months after surgery (Table 4). Low levels of halo perception were observed in 55% (11/20) and 60% (12/20) of patients at 3 and 6 months after surgery, respectively. No patients complained about any other photic adverse effects such as glare.
Table 4

Photic adverse effects (halo) – 3 and 6-month comparison (range 0–4; 0 = none, 4 = strong/severe).

Question3 months6 monthsp value
Work difficulty regarding “halo”1.55±0.980.94±1.11<0.003
Level of “halo” perception1.33±1.080.76±1.03<0.005

Complications

No intra- or postoperative complications were observed. At 6 months, no cases of posterior capsular opacification requiring YAG capsulotomy were detected.

Discussion

Mean binocular logMAR UDVA in the current series was −0.10 at 6 months after surgery, which confirms the ability of this multifocal IOL to successfully restore distance vision, consistent with reports by other authors using the same type of multifocal IOL [3,4]. Specifically, Bautista et al. [3] found a mean postoperative logMAR UDVA of 0.08 at 2 months postoperatively, and Friedrich [4] reported that 94.7% of patients implanted with the same IOL used in our study had a binocular UDVA of 0.1 LogMAR or better. The good distance vision outcome obtained in the current and previous studies is accompanied by a predictable correction of ocular refraction, resulting in minimum residual refractive errors. In comparison with other models of diffractive and zonal refractive multifocal IOLs, the level of binocular distance visual acuity is similar or even better [7-12]. Mean binocular logMAR UNVA at 6 months after surgery in the current series was −0.02, which is consistent with the results of previous series evaluating the same type of multifocal IOL [3,4]. Bautista et al. [3] found that 94.3% of eyes from a sample implanted with the Tecnis ZMB00 IOL could read 0.00 logMAR without correction at 2 months postoperatively. Similarly, in a sample of patients implanted bilaterally with the same type of multifocal IOL, Friedrich4 reported that 67.7% of eyes could read Jaeger 1+ (0.0 LogMAR) and 93.6% could read Jaeger 1 (0.1 LogMAR) or better at 6 months after surgery. These results indicate that this multifocal IOL is also able to restore the near vision function successfully. These outcomes are similar to or better than those reported for other modalities of aspheric diffractive and zonal refractive multifocal IOLs [7-12]. Tsaousis et al. [7] found a mean postoperative (mean follow-up: 26±6 months) logMAR binocular UNVA of 0.11±0.10 in a sample of eyes implanted with the hybrid apodized diffractive/refractive IOL AcrySof IQ ReSTOR IOL (Alcon, Inc. USA). Alfonso et al. [12] found a mean 6-month postoperative logMAR UNVA of −0.05±0.07 in a sample of eyes implanted with the fully diffractive IOL Acri.Lisa 366D. Besides distance and near vision outcomes, the current study is the first reporting on intermediate visual outcomes achievable with the Tecnis ZMB00 multifocal IOL. Mean logMAR UIVA was 0.12, a very similar value to those reported by other authors using other types of multifocal IOLs [7-12] (Tsaousis et al. [7] 0.11±0.10 with AcrySof IQ ReSTOR and Alfonso et al. [12] 0.19±0.14 with Acri.Lisa 366D). Likewise, the UIVA outcome obtained in the current series is comparable to that reported for a previous model of the Tecnis multifocal lens (ZM900) [13]. Our results show that the IOL evaluated also provides a functional level of intermediate vision. Furthermore, in the current series, UIVA and UNVA improved significantly from 3 to 6 month postoperatively. Several factors may have contributed to this finding, but patient neuroadaptation to the multifocality induced by the IOL optics seems to have played a major role. Indeed, a similar finding has been reported with other diffractive multifocal IOLs [14,15] and it has even been demonstrated that visual performance after multifocal IOL implantation can be significantly accelerated by training programs [16]. Via the neuroadaptation process (synaptogenesis, neurogenesis) [17], the brain has the ability to select an image related to the object that is being looked at, suppressing other images. Regarding contrast sensitivity, the results obtained in the current series were well within the normal limits defined for the 50–75 years age range [6], which corresponds to the age range of the sample of patients of the current series. However, the values obtained for higher spatial frequencies were close to the lower limit of the normality range. This outcome is similar to or even better than those reported for other designs of multifocal IOLs, including diffractive and zonal refractive models [9,18,19]. Furthermore, some significant improvements were detected in distance and near contrast sensitivity between the 3-month and 6-month postoperative visits. As with UNVA and UIVA, neuroadaptation may also have played a role in this improvement of visual performance. As a result of the ability of the IOL to restore the visual acuity and contrast sensitivity, independence from wearing corrective lenses was high, with a total of 85% of patients achieving complete independence. Similar results have also been reported with other models of multifocal IOLs [13,20-22]. Cillino et al. [22], in a comparative study of the clinical outcomes obtained with 4 types of IOL, found that independence from wearing corrective lenses was achieved in 20% of cases implanted with a monofocal IOL (AR40 from AMO), in 43.7% and 53.3% of cases implanted with the multifocal refractive IOLs Array SA40N and ReZoom from AMO, respectively, and in 87.5% of cases implanted with the diffractive multifocal IOL Tecnis ZM900. Finally, patient satisfaction with the outcome of surgery was evaluated using the VF-14 questionnaire. General patient satisfaction was very high and stable, with most of patients scoring their level of satisfaction at between 8 and 10 (0 = not satisfied at all and 10 = completely satisfied), as reported for a previous model of the Tecnis diffractive IOL [21]. The VF-14 test detected only slight to moderate difficulty in driving at night. This effect is mainly attributed to the photic phenomena induced by diffractive optics. This effect is reduced with the introduction of an aspheric optic, minimizing the level of spherical aberration. In contrast, the effect is increased if the multifocal IOL has an additional refractive component [23]. In the current series, a low level of halo perception was reported in 60% of patients at 6 months postoperatively, with no severe complaints of halos. Future studies should confirm if this subjective symptomatology disappears with time, as suggested by many authors based on the short- and medium-term outcomes [15] and according to the significant reduction of the intensity of photic phenomena with time observed in the current series.

Conclusions

The diffractive multifocal IOL Tecnis ZMB00 provides an effective restoration of distance, intermediate, and near vision, promoting a very high level of independence from wearing corrective lenses, as well as high patient satisfaction. Studies should be conducted to evaluate the long-term outcomes obtained with this modality of diffractive multifocal IOL.
  23 in total

1.  Intermediate visual function with different multifocal intraocular lens models.

Authors:  José F Alfonso; Luis Fernández-Vega; Cristina Puchades; Robert Montés-Micó
Journal:  J Cataract Refract Surg       Date:  2010-05       Impact factor: 3.351

Review 2.  Intraocular lenses for presbyopia correction: past, present, and future.

Authors:  Alejandro Lichtinger; David S Rootman
Journal:  Curr Opin Ophthalmol       Date:  2012-01       Impact factor: 3.761

Review 3.  Review of presbyopic IOLs: multifocal and accommodating IOLs.

Authors:  Patrick Gooi; Ike K Ahmed
Journal:  Int Ophthalmol Clin       Date:  2012

4.  Double-pass system analysis of the visual outcomes and optical performance of an apodized diffractive multifocal intraocular lens.

Authors:  Luis J Moreno; David P Piñero; Jorge L Alió; Antonio Fimia; Ana B Plaza
Journal:  J Cataract Refract Surg       Date:  2010-12       Impact factor: 3.351

5.  [IOL multifocality combined with the compensation for corneal spherical aberration: a new concept of presbyopia correcting IOL].

Authors:  R Friedrich
Journal:  Klin Monbl Augenheilkd       Date:  2012-09-19       Impact factor: 0.700

6.  Visual outcomes after cataract surgery with implantation of a +3.00 D or +4.00 D aspheric diffractive multifocal intraocular lens: Comparative study.

Authors:  Niels E de Vries; Carroll A B Webers; Robert Montés-Micó; Teresa Ferrer-Blasco; Rudy M M A Nuijts
Journal:  J Cataract Refract Surg       Date:  2010-08       Impact factor: 3.351

7.  The VF-14. An index of functional impairment in patients with cataract.

Authors:  E P Steinberg; J M Tielsch; O D Schein; J C Javitt; P Sharkey; S D Cassard; M W Legro; M Diener-West; E B Bass; A M Damiano
Journal:  Arch Ophthalmol       Date:  1994-05

8.  Evolution of visual performance in 250 eyes implanted with the Tecnis ZM900 multifocal IOL.

Authors:  Carlos Palomino Bautista; David Carmona González; Alfredo Castillo Gómez; José Angel Cristóbal Bescos
Journal:  Eur J Ophthalmol       Date:  2009 Sep-Oct       Impact factor: 2.597

9.  One-year outcomes with new-generation multifocal intraocular lenses.

Authors:  Salvatore Cillino; Alessandra Casuccio; Francesco Di Pace; Raffaella Morreale; Francesco Pillitteri; Giovanni Cillino; Gaetano Lodato
Journal:  Ophthalmology       Date:  2008-06-05       Impact factor: 12.079

10.  Evolution of visual performance in 70 eyes implanted with the Tecnis(®) ZMB00 multifocal intraocular lens.

Authors:  Carlos Palomino Bautista; David Carmona González; Alfredo Castillo Gómez
Journal:  Clin Ophthalmol       Date:  2012-03-21
View more
  12 in total

1.  Diffractive multifocal intraocular lens implantation in patients with monofocal intraocular lens in the contralateral eye.

Authors:  Jae Yong Kim; Yunhan Lee; Hun Jae Won; Hyerin Jeong; Jin Hyoung Park; Myoung Joon Kim; Hungwon Tchah
Journal:  Int J Ophthalmol       Date:  2020-05-18       Impact factor: 1.779

2.  Comparison of long-term decentration and tilt in two types of multifocal intraocular lenses with OPD-Scan III aberrometer.

Authors:  Wenwen He; Xiaodi Qiu; Shaohua Zhang; Yu Du; Yinglei Zhang; Yi Lu; Xiangjia Zhu
Journal:  Eye (Lond)       Date:  2018-03-09       Impact factor: 3.775

3.  Clinical outcomes of a new diffractive multifocal intraocular lens.

Authors:  Baha Toygar; Ozge Yabas Kiziloglu; Okan Toygar; Ali Murat Hacimustafaoglu
Journal:  Int J Ophthalmol       Date:  2017-12-18       Impact factor: 1.779

4.  Bilateral Implantation of a New Refractive Multi-Segmented Multifocal Intraocular Lens in Cataract or Refractive Lens Exchange Patients.

Authors:  Mike P Holzer; Rudy M M A Nuijts; Soraya M R Jonker; Erik L J G Mertens; Asim Bozkurt Sener; Jorge A O Cazal; Antoni Salvador Playa; Ramon Ruiz Mesa; Bettina C Thomas
Journal:  Clin Ophthalmol       Date:  2021-05-21

5.  Visual Outcomes at 12 Months in Patients Following Implantation of a Diffractive Multifocal Intraocular Lens.

Authors:  Mehmet Baykara; Yonca Aydin Akova; Osman Sevki Arslan; Necdet Cinhuseyinoglu; Tamer Takmaz; Ahmet Gucukoglu; Yildirim Beyazit Usta
Journal:  Ophthalmol Ther       Date:  2015-03-31

6.  Visual Outcomes, Patient Satisfaction and Spectacle Independence with a Trifocal Diffractive Intraocular Lens.

Authors:  Florian Tobias Alwin Kretz; Chul Young Choi; Matthias Müller; Matthias Gerl; Ralf Helmar Gerl; Gerd Uwe Auffarth
Journal:  Korean J Ophthalmol       Date:  2016-05-18

7.  Influence of the level of monovision on visual outcome with an extended range of vision intraocular lens.

Authors:  Béatrice Cochener
Journal:  Clin Ophthalmol       Date:  2018-11-12

8.  Prospective comparative study of tolerance to refractive errors after implantation of extended depth of focus and monofocal intraocular lenses with identical aspheric platform in Korean population.

Authors:  Hyeck-Soo Son; Seong Ho Kim; Gerd U Auffarth; Chul Young Choi
Journal:  BMC Ophthalmol       Date:  2019-08-19       Impact factor: 2.209

9.  Visual outcomes of patients bilaterally implanted with the extended range of vision intraocular lens: A prospective study.

Authors:  Gitansha Shreyas Sachdev; Shreyas Ramamurthy; Umesh Sharma; Ramamurthy Dandapani
Journal:  Indian J Ophthalmol       Date:  2018-03       Impact factor: 1.848

10.  Visual Performances of a New Extended Depth-of-Focus Intraocular Lens with a Refractive Design: A Prospective Study After Bilateral Implantation.

Authors:  Leopoldo Spadea; Maria Ilaria Giannico; Martina Formisano; Ludovico Alisi
Journal:  Ther Clin Risk Manag       Date:  2021-07-16       Impact factor: 2.423

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.