Twenty-year followup of an uncemented stem in primary THA.

John B Meding1, Merrill A Ritter, E Michael Keating, Michael E Berend.   

Abstract

BACKGROUND: Uncemented stems have been used in THA for well over two decades, but there are relatively few studies reporting on the results after 20 years. QUESTIONS/PURPOSES: The purpose of this study was to evaluate at a minimum followup of 20 years (1) hip scores; (2) radiographic findings, including stem fixation and osteolysis; (3) reoperations; and (4) survivorship free from aseptic loosening in a group of patients who underwent primary THA using a proximally porous-coated, plasma-sprayed, straight-stemmed, titanium-alloy femoral component.
METHODS: Between 1987 and 1993, we performed 1517 primary THAs, of which 447 were cementless, and 157 used the implant under study here (representing 10% of the THAs during the period in question). General indications for the use of the study stem included (1) younger age with relatively stronger bone (average age 55 years in this study); (2) patients not involved in another study protocol (as were the 278 other cementless hips implanted during this time); and (3) patients who, at the surgeon's discretion, would be compliant with protected weightbearing. Of those, 111 were available for followup at a minimum of 20 years (mean, 20 years), whereas 36 had died and nine (6%) were lost to followup or declined participation. The primary diagnosis was osteoarthritis in 77 hips (69%). The mean age at operation was 55 years. The clinical result was evaluated on the basis of the Harris hip score. Radiographic analysis was performed at each followup visit for distal cortical hypertrophy, spot welds, radiolucencies, stem subsidence, varus or valgus shift, osteolysis, and femoral component loosening. Kaplan-Meier analysis was performed to evaluate the survival of the femoral component.
RESULTS: The mean Harris hip score improved from 46 points to 87 points (of a possible 100 points) at the most recent followup. All hips had evidence of proximal femoral remodeling consistent with osseous ingrowth. Femoral osteolyis was seen in 22 hips (20%). One stem was revised as a result of a periprosthetic fracture. No femoral component had evidence of loosening, and none was revised. Thirty-three acetabular components (30%) were revised.
CONCLUSIONS: This femoral component provided durable long-term fixation for over two decades after THA. The porous stem geometry is still in use today and will continue to be studied into the third decade of use. LEVEL OF EVIDENCE: Level IV, therapeutic study. See Instructions for Authors for a complete description of levels of evidence.

Entities:  

Mesh:

Year:  2015        PMID: 24993142      PMCID: PMC4294900          DOI: 10.1007/s11999-014-3763-y

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  25 in total

1.  Long-term followup of anatomic graduated components posterior cruciate-retaining total knee replacement.

Authors:  M A Ritter; M E Berend; J B Meding; E M Keating; P M Faris; B M Crites
Journal:  Clin Orthop Relat Res       Date:  2001-07       Impact factor: 4.176

2.  Survivorship of 2000 tapered titanium porous plasma-sprayed femoral components.

Authors:  Adolph V Lombardi; Keith R Berend; Thomas H Mallory; Michael D Skeels; Joanne B Adams
Journal:  Clin Orthop Relat Res       Date:  2008-10-31       Impact factor: 4.176

3.  Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components.

Authors:  C A Engh; P Massin; K E Suthers
Journal:  Clin Orthop Relat Res       Date:  1990-08       Impact factor: 4.176

4.  Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation.

Authors:  W H Harris
Journal:  J Bone Joint Surg Am       Date:  1969-06       Impact factor: 5.284

5.  "Modes of failure" of cemented stem-type femoral components: a radiographic analysis of loosening.

Authors:  T A Gruen; G M McNeice; H C Amstutz
Journal:  Clin Orthop Relat Res       Date:  1979-06       Impact factor: 4.176

6.  21-year results of the uncemented fully textured lord hip prosthesis.

Authors:  Javier S Martínez de Aragón; Kjell S Keisu
Journal:  Clin Orthop Relat Res       Date:  2007-01       Impact factor: 4.176

7.  Comparison of collared and collarless femoral components in primary uncemented total hip arthroplasty.

Authors:  J B Meding; M A Ritter; E M Keating; P M Faris
Journal:  J Arthroplasty       Date:  1997-04       Impact factor: 4.757

8.  Results of the anatomic medullary locking total hip arthroplasty at a minimum of twenty years. A concise follow-up of previous reports.

Authors:  Philip J Belmont; Cara C Powers; Sarah E Beykirch; Robert H Hopper; C Anderson Engh; Charles A Engh
Journal:  J Bone Joint Surg Am       Date:  2008-07       Impact factor: 5.284

9.  Total hip arthroplasty with an uncemented tapered femoral component.

Authors:  Jeffrey R McLaughlin; Kyla R Lee
Journal:  J Bone Joint Surg Am       Date:  2008-06       Impact factor: 5.284

10.  Polyethylene wear in uncemented acetabular components.

Authors:  J R Hernandez; E M Keating; P M Faris; J B Meding; M A Ritter
Journal:  J Bone Joint Surg Br       Date:  1994-03
View more
  5 in total

1.  Densitometric evaluation of bone-prosthetic counterface in hip and knee arthroplasty with modern implants.

Authors:  Andrea Cozzi Lepri; Marco Giorgini; Carla Signorini; Christian Carulli; Roberto Civinini; Maria Luisa Brandi; Massimo Innocenti
Journal:  Clin Cases Miner Bone Metab       Date:  2016-10-05

2.  Effect of stem alignment on long-term outcomes of total hip arthroplasty with cementless Bi-Metric femoral components.

Authors:  Takaaki Shishido; Toshiyuki Tateiwa; Yasuhito Takahashi; Toshinori Masaoka; Tsunehito Ishida; Kengo Yamamoto
Journal:  J Orthop       Date:  2018-01-16

3.  The outcome of the partial resurfacing arthroplasty of the hip shows high numbers of failures and conversion to total arthroplasty.

Authors:  Thilo Floerkemeier; Stefan Budde; Nils Wirries; Gabriela von Lewinski; Henning Windhagen; Marco Ezechieli
Journal:  Int Orthop       Date:  2017-08-17       Impact factor: 3.075

4.  Diagnosis of and Early Revision Surgery for Biological Fixation Failure Due to Proximal-Distal Mismatch of Proximally Coated Tapered Cementless Stem.

Authors:  Yasuhiro Homma; So Kawakita; Tomonori Baba; Taiji Watari; Kazuo Kaneko
Journal:  Arthroplast Today       Date:  2020-11-04

5.  Ten-year comparison of two different techniques for femoral bone cavity preparation-broaching versus compaction in patients with cementless total hip arthroplasty : a randomized radiostereometric study of 30 total hip arthroplasties in 15 patients operated bilaterally.

Authors:  Maciej Okowinski; Mette Holm Hjorth; Sebastian Breddam Mosegaard; Jonathan Hugo Jürgens-Lahnstein; Stig Storgaard Jakobsen; Poul Hedevang Christensen; Søren Kold; Maiken Stilling
Journal:  Bone Jt Open       Date:  2021-12
  5 in total