Towards integration of PET/MR hybrid imaging into radiation therapy treatment planning.

Daniel H Paulus1, Daniela Thorwath2, Holger Schmidt3, Harald H Quick4.   

Abstract

PURPOSE: Multimodality imaging has become an important adjunct of state-of-the-art radiation therapy (RT) treatment planning. Recently, simultaneous PET/MR hybrid imaging has become clinically available and may also contribute to target volume delineation and biological individualization in RT planning. For integration of PET/MR hybrid imaging into RT treatment planning, compatible dedicated RT devices are required for accurate patient positioning. In this study, prototype RT positioning devices intended for PET/MR hybrid imaging are introduced and tested toward PET/MR compatibility and image quality.
METHODS: A prototype flat RT table overlay and two radiofrequency (RF) coil holders that each fix one flexible body matrix RF coil for RT head/neck imaging have been evaluated within this study. MR image quality with the RT head setup was compared to the actual PET/MR setup with a dedicated head RF coil. PET photon attenuation and CT-based attenuation correction (AC) of the hardware components has been quantitatively evaluated by phantom scans. Clinical application of the new RT setup in PET/MR imaging was evaluated in anin vivo study.
RESULTS: The RT table overlay and RF coil holders are fully PET/MR compatible. MR phantom and volunteer imaging with the RT head setup revealed high image quality, comparable to images acquired with the dedicated PET/MR head RF coil, albeit with 25% reduced SNR. Repositioning accuracy of the RF coil holders was below 1 mm. PET photon attenuation of the RT table overlay was calculated to be 3.8% and 13.8% for the RF coil holders. With CT-based AC of the devices, the underestimation error was reduced to 0.6% and 0.8%, respectively. Comparable results were found within the patient study.
CONCLUSIONS: The newly designed RT devices for hybrid PET/MR imaging are PET and MR compatible. The mechanically rigid design and the reproducible positioning allow for straightforward CT-based AC. The systematic evaluation within this study provides the technical basis for the clinical integration of PET/MR hybrid imaging into RT treatment planning.

Entities:  

Mesh:

Year:  2014        PMID: 24989408     DOI: 10.1118/1.4881317

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  13 in total

Review 1.  Combined PET/MRI: Multi-modality Multi-parametric Imaging Is Here: Summary Report of the 4th International Workshop on PET/MR Imaging; February 23-27, 2015, Tübingen, Germany.

Authors:  D L Bailey; B J Pichler; B Gückel; H Barthel; A J Beer; J Bremerich; J Czernin; A Drzezga; C Franzius; V Goh; M Hartenbach; H Iida; A Kjaer; C la Fougère; C N Ladefoged; I Law; K Nikolaou; H H Quick; O Sabri; J Schäfer; M Schäfers; H F Wehrl; T Beyer
Journal:  Mol Imaging Biol       Date:  2015-10       Impact factor: 3.488

2.  How PET/MR Can Add Value For Children With Cancer.

Authors:  Heike Daldrup-Link
Journal:  Curr Radiol Rep       Date:  2017-02-21

3.  Voxel-wise correlation of functional imaging parameters in HNSCC patients receiving PET/MRI in an irradiation setup.

Authors:  Kerstin Zwirner; Daniela Thorwarth; René M Winter; Stefan Welz; Jakob Weiss; Nina F Schwenzer; Holger Schmidt; Christian la Fougère; Konstantin Nikolaou; Daniel Zips; Sergios Gatidis
Journal:  Strahlenther Onkol       Date:  2018-03-21       Impact factor: 3.621

Review 4.  Attenuation Correction for Magnetic Resonance Coils in Combined PET/MR Imaging: A Review.

Authors:  Mootaz Eldib; Jason Bini; David D Faul; Niels Oesingmann; Charalampos Tsoumpas; Zahi A Fayad
Journal:  PET Clin       Date:  2015-11-27

5.  Developing quality assurance tests for simultaneous Positron Emission Tomography - Magnetic Resonance imaging for radiotherapy planning.

Authors:  Jonathan J Wyatt; Hazel M McCallum; Ross J Maxwell
Journal:  Phys Imaging Radiat Oncol       Date:  2022-04-20

6.  Analysis of pairwise correlations in multi-parametric PET/MR data for biological tumor characterization and treatment individualization strategies.

Authors:  Sara Leibfarth; Urban Simoncic; David Mönnich; Stefan Welz; Holger Schmidt; Nina Schwenzer; Daniel Zips; Daniela Thorwarth
Journal:  Eur J Nucl Med Mol Imaging       Date:  2016-02-13       Impact factor: 9.236

Review 7.  Functional imaging for radiotherapy treatment planning: current status and future directions-a review.

Authors:  D Thorwarth
Journal:  Br J Radiol       Date:  2015-04-01       Impact factor: 3.039

8.  NEMA image quality phantom measurements and attenuation correction in integrated PET/MR hybrid imaging.

Authors:  Susanne Ziegler; Bjoern W Jakoby; Harald Braun; Daniel H Paulus; Harald H Quick
Journal:  EJNMMI Phys       Date:  2015-08-20

Review 9.  Target Definition in Salvage Radiotherapy for Recurrent Prostate Cancer: The Role of Advanced Molecular Imaging.

Authors:  Gaël Amzalag; Olivier Rager; Claire Tabouret-Viaud; Michael Wissmeyer; Electra Sfakianaki; Thomas de Perrot; Osman Ratib; Raymond Miralbell; Giampiero Giovacchini; Valentina Garibotto; Thomas Zilli
Journal:  Front Oncol       Date:  2016-03-31       Impact factor: 6.244

10.  Assessment of image quality of a radiotherapy-specific hardware solution for PET/MRI in head and neck cancer patients.

Authors:  René M Winter; Sara Leibfarth; Holger Schmidt; Kerstin Zwirner; David Mönnich; Stefan Welz; Nina F Schwenzer; Christian la Fougère; Konstantin Nikolaou; Sergios Gatidis; Daniel Zips; Daniela Thorwarth
Journal:  Radiother Oncol       Date:  2018-05-07       Impact factor: 6.280

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.