| Literature DB >> 24985763 |
Hee Jae Huh1, Kyung Sun Park1, Ji-Youn Kim2, Hyeon Jeong Kwon2, Jong-Won Kim1, Chang-Seok Ki3, Nam Yong Lee1.
Abstract
The Anyplex(TM) II RV16 detection kit (RV16; Seegene, Seoul, South Korea) is a multiplex real-time PCR assay based on tagging oligonucleotide cleavage extension. In this prospective study, we evaluated the RV16 assay by comparing with the Seeplex(®) RV12 ACE detection kit (RV12; Seegene), a multiplex end-point PCR kit. A total of 365 consecutive respiratory specimens were tested with both RV16 and RV12 assays in parallel and detected 140 (38.4%) and 89 (24.4%) positive cases, respectively. The positive percent agreement, negative percent agreement, and kappa values for the 2 assays were 95.6% (95% confidence interval [CI], 89.4-98.3%), 80.4% (95% CI, 75.3-84.6%), and 0.64 (95% CI, 0.56-0.72), respectively. The monoplex PCR and sequencing for the samples with discrepant results revealed that majority of the results were concordant with the results from RV16 assays. In conclusion, the RV16 assay produces results comparable to the RV12 assay.Entities:
Keywords: Korea; Multiplex real-time PCR; Respiratory virus; Validation
Mesh:
Year: 2014 PMID: 24985763 PMCID: PMC7132765 DOI: 10.1016/j.diagmicrobio.2014.01.025
Source DB: PubMed Journal: Diagn Microbiol Infect Dis ISSN: 0732-8893 Impact factor: 2.803
Detection of respiratory viruses by AnyplexTM II RV16 and Seeplex® RV12 ACE assays.
| Virus | Agreement | Kappa value | ||||
|---|---|---|---|---|---|---|
| Positive | Negative | Observed kappa | 95% CI | |||
| % | 95% CI | % | 95% CI | |||
| Total | 95.5 | 88.3–98.5 | 80.0 | 74.8–84.2 | 0.64 | 0.55–0.71 |
| ADV | 100 | 39.6–100 | 95.6 | 92.8–97.4 | 0.32 | 0.08–0.56 |
| INF A | 95.6 | 76.0–99.8 | 97.1 | 94.5–98.5 | 0.78 | 0.66–0.91 |
| INF B | NA | NA | 100 | 98.7–100 | NA | NA |
| PIV 1 | 0 | 0–94.5 | 100 | 98.7–100 | 0 | 0–0 |
| PIV 2 | 100 | 54.6 - 100 | 100 | 98.7–100 | 1.00 | 1.00–1.00 |
| PIV 3 | NA | NA | 100 | 98.7–100 | NA | NA |
| HRV | 90.0 | 59.6–98.2 | 93.6 | 93.4–97.8 | 0.53 | |
| RSV (A or B) | 97.1 | 83.4–99.8 | 98.2 | 95.9–99.3 | 0.90 | 0.82–0.97 |
| HMPV | 100 | 46.3–100 | 98.3 | 96.2–99.3 | 0.62 | 0.34–0.90 |
| CoV | 100 | 69.9–100 | 94.6 | 91.6–96.6 | 0.54 | 0.36–0.72 |
NA = not applicable.
Analysis for positive results in AnyplexTM II RV16 and Seeplex® RV12 ACE assays.
| Virus | RV16+/RV12+ | RV16+/RV12− | RV16−/RV12+ | ||
|---|---|---|---|---|---|
| No. of samples | No. of positive results in | No. of samples | No. of positive results in monoplex PCR and sequencing | ||
| ADV | 4 | 16 | 13 | 0 | 0 |
| INF A | 22 | 10 | 5 | 1 | 0 |
| PIV 1 | 0 | 0 | 0 | 1 | 0 |
| PIV 2 | 1 | 0 | 0 | 0 | 0 |
| HRV | 9 | 14 | 13 | 1 | 0 |
| RSV (A or B) | 34 | 6 | 4 | 1 | ND |
| HMPV | 5 | 6 | 6 | 0 | 0 |
| CoV | 12 | 19 | 17 | 0 | 0 |
ND = not performed due to lack of sample.