Christian Renaud1, Janet Crowley, Keith R Jerome, Jane Kuypers.
Abstract
The FilmArray Respiratory Panel (Idaho Technology) is a highly multiplexed respiratory virus real-time polymerase chain reaction (PCR) assay. Eighty-four respiratory viruses identified by laboratory-developed real-time reverse transcription-PCR assays (LDA) or by viral cultures were mixed and tested by FilmArray to assess its performance. FilmArray identified 72 (90%) of 80 viruses also detected by LDA. Six of the 8 viruses not detected by FilmArray had PCR cycle threshold values >35. Compared to LDA, FilmArray showed comparable sensitivity when used to test serial dilutions of virus mixtures and good agreement with negative samples. With the use of 1 FilmArray instrument, 7 clinical samples could be analyzed and reported in an 8-h shift compared to 20 using LDA and 1 real-time detection instrument. While the FilmArray was rapid and easy to use, its low throughput and qualitative results may be a disadvantage in some clinical settings.Entities:
Mesh:
Year: 2012 PMID: 22981482 PMCID: PMC7132734 DOI: 10.1016/j.diagmicrobio.2012.08.003
Source DB: PubMed Journal: Diagn Microbiol Infect Dis ISSN: 0732-8893 Impact factor: 2.803
Correlation between expected results from laboratory-developed real-time RT-PCR/viral cultures and FilmArray Respiratory Panel assay.
| Expected virus | Number tested | Number FilmArray positive |
|---|---|---|
| Adenovirus | 11 | 7 |
| A31 | 1 | 1 |
| B3 | 1 | 1 |
| B7 | 1 | 1 |
| B21 | 1 | 1 |
| C2 | 1 | 0 |
| C5 | 1 | 1 |
| C6 | 1 | 0 |
| D8 | 1 | 0 |
| E4 | 1 | 1 |
| F41 | 1 | 1 |
| Untyped | 1 | 0 |
| Bocavirus | 1 | 1 |
| Enterovirus | 5 | 5 |
| Coxsackievirus A16 | 1 | 1 |
| Coxsackievirus B3 | 1 | 1 |
| Coxsackiesvirus B5 | 1 | 1 |
| Echovirus 6 | 1 | 1 |
| Echovirus 9 | 1 | 1 |
| Human rhinovirus | 14 | 14 |
| A2 | 1 | 1 |
| A8 | 1 | 1 |
| A30 | 1 | 1 |
| A68 | 1 | 1 |
| A80 | 1 | 1 |
| B3 | 1 | 1 |
| B6 | 1 | 1 |
| B35 | 1 | 1 |
| C | 3 | 3 |
| Untyped | 3 | 3 |
| Influenza B | 5 | 5 |
| Influenza A | 14 | 11 |
| Pandemic H1, 2009 | 7 | 6 |
| Seasonal H1, 2008 | 1 | 1 |
| Seasonal H3, 2009 | 6 | 4 |
| Human metapneumovirus | 6 | 4 |
| Parainfluenza 1 | 4 | 3 |
| Parainfluenza 2 | 4 | 4 |
| Parainfluenza 3 | 5 | 5 |
| Parainfluenza 4 | 1 | 1 |
| Respiratory syncytial virus | 8 | 6 |
| A | 3 | 3 |
| B | 4 | 2 |
| Untyped | 1 | 1 |
| Coronavirus | 6 | 6 |
| 229E | 1 | 1 |
| HKU1 | 2 | 2 |
| NL63 | 2 | 2 |
| OC43 | 1 | 1 |
| Total | 84 | 72 |
Results of repeat laboratory-developed real-time RT-PCR assays for 12 samples with discrepant results by FilmArray.
| Expected virus | Ct value of repeated laboratory-developed real-time RT-PCR | Final sample resolution |
|---|---|---|
| Adenovirus C2 | 35.3 | Discordant |
| Adenovirus C6 | 24.9 | Discordant |
| Adenovirus D8 | 33.0 | Discordant |
| Human metapneumovirus | 39.0 | Discordant |
| Human metapneumovirus | 35.8 | Discordant |
| Influenza A pandemic 2009 H1 | 37.0 | Discordant |
| Influenza A seasonal H3, 2009 | 37.2 | Discordant |
| Respiratory syncytial virus B | 37.4 | Discordant |
| Adenovirus untyped | Negative | Concordant |
| Influenza A seasonal H3, 2009 | Negative | Concordant |
| Respiratory syncytial virus B | Negative | Concordant |
| Parainfluenza 1 | Negative | Concordant |
Results of laboratory-developed real-time RT-PCR assays and FilmArray Respiratory Panel assay for dilutions of the respiratory virus positive mix control.
| Virus target | Undiluted | 1:10 | 1:100 | 1:1000 | ||||
|---|---|---|---|---|---|---|---|---|
| Real-time CT value | FilmArray result | Real-time CT value | FilmArray result | Real-time CT value | FilmArray result | Real-time CT value | FilmArray result | |
| RSV | 28.3 | + | 33.6 | + | 38.1 | + | − | − |
| FLU A | 30.4 | + | 34.5 | + | 37.8 | + | − | − |
| PIV3 | 27.7 | + | 33.3 | + | 39.3 | + | − | − |
| MPV | 30.0 | + | 36.2 | − | − | − | − | − |
| ADV | 23.5 | + | 28.0 | + | 32.9 | + | 34.9 | − |
| CoV | 28.1 | + | 33.4 | + | 36.8 | + | 39.3 | − |
| RHV | 26.6 | + | 31.9 | + | 36.9 | + | 37.6 | − |
| BCV | 28.3 | + | 32.5 | + | 36.9 | + | − | − |
| PIV2 | 28.7 | + | 34.7 | − | 38.3 | − | − | − |
| FLU B | 30.3 | + | 34.4 | + | − | + | − | − |
| PIV4 | 28.5 | + | 34.7 | + | − | + | − | − |
| PIV1 | 29.3 | + | 33.6 | − | 38.0 | − | − | − |
Results of laboratory-developed real-time RT-PCR assays and FilmArray Respiratory Panel assay for dilutions of the influenza positive mix control.
| Virus target | Undiluted | 1:10 | 1:100 | |||
|---|---|---|---|---|---|---|
| Real-time Ct value | FilmArray result | Real-time Ct value | FilmArray result | Real-time Ct value | FilmArray result | |
| Pan FLU A | 30.1 | + | 33.0 | + | 36.8 | + |
| FLU A H3 | 28.9 | + | 32.1 | + | − | + |
| FLU A H1 seasonal | 32.8 | + | − | − | − | − |
| FLU A H1 2009 pandemic | 34.8 | − | 37.8 | − | − | − |
| FLUB | 31.9 | + | 35.7 | + | − | + |
Comparison of selected characteristics of direct fluorescent antibody, laboratory developed real-time PCR, and FDA-approved respiratory panel assays.
| Characteristic | DFA | LDA | XTAG | Prodesse | FilmArray |
|---|---|---|---|---|---|
| Rapid turnaround time for 1 sample | +++ | ++ | + | ++ | +++ |
| Ease of use | + | ++ | ++ | ++ | +++ |
| Throughput | ++ | +++ | +++ | +++ | + |
| Small laboratory footprint | +++ | ++ | + | ++ | +++ |
| Low cost | +++ | ++ | + | + | + |
| Sensitivity | + | +++ | +++ | +++ | +++ |
| Quantitative result | + | +++ | + | +++ | + |
DFA = Direct fluorescent antibody.
LDA = Laboratory developed assay.