Calibration of (192)Ir high dose rate brachytherapy source using different calibration procedures.

Shwetha Bondel1, Manickham Ravikumar1, Sanjay Sudhakar Supe1, Buchuppudi Rekha Reddy1.   

Abstract

AIM: To calibrate Ir-192 high dose rate (HDR) brachytherapy source using different calibration methods and to determine the accuracy and suitability of each method for routine calibrations.
BACKGROUND: The source calibration is an essential part of the quality assurance programme for dosimetry of brachytherapy sources. The clinical use of brachytherapy source requires an independent measurement of the air kerma strength according to the recommendations of medical physics societies.
MATERIALS AND METHODS: The Ir-192 HDR brachytherapy source from Gammamed plus machine (Varian Medical Systems, Palo Alto, CA) was calibrated using three different procedures, one using the well-type ionization chamber, second by the in-air calibration method and third using solid water phantoms. The reference air kerma rate (RAKR) of the source was determined using Deutsche Gesellschaft fur Medizinische Physik (DGMP) recommendations.
RESULTS: The RAKR determined using different calibration methods are in good agreement with the manufacturer stated value. The mean percentage variations of 0.21, -0.94, -0.62 and 0.58 in RAKR values with respect to the manufacturer quoted values were observed with the well-type chamber, in-air calibration, cylindrical phantom and slab phantom measurements, respectively.
CONCLUSION: Measurements with a well-type chamber are relatively simple to perform. For in-air measurements, the indigenously designed calibration jig provides an accurate positioning of the source and chamber with minimum scatter contribution. The slab phantom system has an advantage that no additional phantom and chamber are required other than those used for external beam therapy dosimetry. All the methods of calibration discussed in this study are effective to be used for routine calibration purposes.

Entities:  

Keywords:  High dose rate brachytherapy; In-air calibration; Reference air kerma rate; Well-type chamber

Year:  2013        PMID: 24944818      PMCID: PMC4059024          DOI: 10.1016/j.rpor.2013.07.014

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  11 in total

1.  Comparison of calibration procedures for 192Ir high-dose-rate brachytherapy sources.

Authors:  D Baltas; K Geramani; G T Ioannidis; K Hierholz; B Rogge; C Kolotas; K Müller-Sievers; N Milickovic; B Kober; N Zamboglou
Journal:  Int J Radiat Oncol Biol Phys       Date:  1999-02-01       Impact factor: 7.038

2.  Experimental derivation of wall correction factors for ionization chambers used in high dose rate 192Ir source calibration.

Authors:  M H Maréchal; C E de Almeida; I H Ferreira; C H Sibata
Journal:  Med Phys       Date:  2002-01       Impact factor: 4.071

3.  Calibration of new high dose rate 192Ir sources.

Authors:  K E Stump; L A Dewerd; J A Micka; D R Anderson
Journal:  Med Phys       Date:  2002-07       Impact factor: 4.071

4.  Determination of the reference air kerma rate for 192Ir brachytherapy sources and the related uncertainty.

Authors:  Eduard van Dijk; Inger-Karine K Kolkman-Deurloo; Patricia M G Damen
Journal:  Med Phys       Date:  2004-10       Impact factor: 4.071

5.  In-air calibration of an HDR 192Ir brachytherapy source using therapy ion chambers.

Authors:  Narayan Prasad Patel; Bishnu Majumdar; V Vijiyan; Pradeep K Hota
Journal:  J Cancer Res Ther       Date:  2005 Oct-Dec       Impact factor: 1.805

6.  A statistical approach to infer the minimum setup distance of a well chamber to the wall or to the floor for 192Ir HDR calibration.

Authors:  Liyun Chang; Sheng-Yow Ho; Chen-Shou Chui; Jeng-Hung Lee; Yi-Chun Du; Tainsong Chen
Journal:  Med Phys       Date:  2008-06       Impact factor: 4.071

7.  Calibration of 192Ir high-dose-rate afterloading systems.

Authors:  S J Goetsch; F H Attix; D W Pearson; B R Thomadsen
Journal:  Med Phys       Date:  1991 May-Jun       Impact factor: 4.071

Review 8.  Code of practice for brachytherapy physics: report of the AAPM Radiation Therapy Committee Task Group No. 56. American Association of Physicists in Medicine.

Authors:  R Nath; L L Anderson; J A Meli; A J Olch; J A Stitt; J F Williamson
Journal:  Med Phys       Date:  1997-10       Impact factor: 4.071

9.  Evaluation of a new sealed reentrant well chamber for HDR and LDR brachytherapy calibrations.

Authors:  S Pai; L E Reinstein
Journal:  Med Phys       Date:  1998-05       Impact factor: 4.071

10.  Comprehensive QA for radiation oncology: report of AAPM Radiation Therapy Committee Task Group 40.

Authors:  G J Kutcher; L Coia; M Gillin; W F Hanson; S Leibel; R J Morton; J R Palta; J A Purdy; L E Reinstein; G K Svensson
Journal:  Med Phys       Date:  1994-04       Impact factor: 4.071

View more
  2 in total

1.  Evaluating the performance of TG-43 protocol in esophageal HDR brachytherapy viewpoint to trachea inhomogeneity.

Authors:  Seyed Mohsen Hosseini Daghigh; Hamid Reza Baghani; Seyed Mahmoud Reza Aghamiri; Seyed Rabi Mahdavi
Journal:  Rep Pract Oncol Radiother       Date:  2017-05-05

Review 2.  Ir-192 Calibration in Air with Farmer Chamber for HDR Brachytherapy.

Authors:  Liyun Chang; Sheng-Yow Ho; Tsair-Fwu Lee; Hueisch-Jy Ding; Pang-Yu Chen
Journal:  J Med Biol Eng       Date:  2016-03-31       Impact factor: 1.553

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.