A statistical approach to infer the minimum setup distance of a well chamber to the wall or to the floor for 192Ir HDR calibration.

Liyun Chang1, Sheng-Yow Ho, Chen-Shou Chui, Jeng-Hung Lee, Yi-Chun Du, Tainsong Chen.   

Abstract

We propose a new method based on statistical analysis technique to determine the minimum setup distance of a well chamber used in the calibration of 192Ir high dose rate (HDR). The chamber should be placed at least this distance away from any wall or from the floor in order to mitigate the effect of scatter. Three different chambers were included in this study, namely, Sun Nuclear Corporation, Nucletron, and Standard Imaging. The results from this study indicated that the minimum setup distance varies depending on the particular chamber and the room architecture in which the chamber was used. Our result differs from that of a previous study by Podgorsak et al. [Med. Phys. 19, 1311-1314 (1992)], in which 25 cm was suggested, and also differs from that of the International Atomic Energy Agency (IAEA)-TECDOC-1079 report, which suggested 30 cm. The new method proposed in this study may be considered as an alternative approach to determine the minimum setup distance of a well-type chamber used in the calibration of 192Ir HDR.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18649450     DOI: 10.1118/1.2911156

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  Calibration of (192)Ir high dose rate brachytherapy source using different calibration procedures.

Authors:  Shwetha Bondel; Manickham Ravikumar; Sanjay Sudhakar Supe; Buchuppudi Rekha Reddy
Journal:  Rep Pract Oncol Radiother       Date:  2013-08-23
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.