Literature DB >> 17998656

In-air calibration of an HDR 192Ir brachytherapy source using therapy ion chambers.

Narayan Prasad Patel1, Bishnu Majumdar, V Vijiyan, Pradeep K Hota.   

Abstract

The Gammamed Plus 192Ir high dose rate brachytherapy sources were calibrated using the therapy level ionization chambers (0.1 and 0.6 cc) and the well-type chamber. The aim of the present study was to assess the accuracy and suitability of use of the therapy level chambers for in-air calibration of brachytherapy sources in routine clinical practice. In a calibration procedure using therapy ion chambers, the air kerma was measured at several distances from the source in a specially designed jig. The room scatter correction factor was determined by superimposition method based on the inverse square law. Various other correction factors were applied on measured air kerma values at multiple distances and mean value was taken to determine the air kerma strength of the source. The results from four sources, the overall mean deviation between measured and quoted source strength by manufacturers was found -2.04% (N = 18) for well-type chamber. The mean deviation for the 0.6 cc chamber with buildup cap was found -1.48 % (N = 19) and without buildup cap was 0.11% (N = 22). The mean deviation for the 0.1 cc chamber was found -0.24% (N = 27). Result shows that probably the excess ionization in case of 0.6 cc therapy ion chamber without buildup cap was estimated about 2.74% and 1.99% at 10 and 20 cm from the source respectively. Scattered radiation measured by the 0.1 cc and 0.6 cc chamber at 10 cm measurement distance was about 1.1% and 0.33% of the primary radiation respectively. The study concludes that the results obtained with therapy level ionization chambers were extremely reproducible and in good agreement with the results of the well-type ionization chamber and source supplier quoted value. The calibration procedure with therapy ionization chambers is equally competent and suitable for routine calibration of the brachytherapy sources.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 17998656     DOI: 10.4103/0973-1482.19590

Source DB:  PubMed          Journal:  J Cancer Res Ther        ISSN: 1998-4138            Impact factor:   1.805


  4 in total

1.  A Monte Carlo study on dose distribution evaluation of Flexisource (192)Ir brachytherapy source.

Authors:  Majid Alizadeh; Mahdi Ghorbani; Abbas Haghparast; Naser Zare; Toktam Ahmadi Moghaddas
Journal:  Rep Pract Oncol Radiother       Date:  2015-02-23

2.  Calibration of (192)Ir high dose rate brachytherapy source using different calibration procedures.

Authors:  Shwetha Bondel; Manickham Ravikumar; Sanjay Sudhakar Supe; Buchuppudi Rekha Reddy
Journal:  Rep Pract Oncol Radiother       Date:  2013-08-23

Review 3.  Ir-192 Calibration in Air with Farmer Chamber for HDR Brachytherapy.

Authors:  Liyun Chang; Sheng-Yow Ho; Tsair-Fwu Lee; Hueisch-Jy Ding; Pang-Yu Chen
Journal:  J Med Biol Eng       Date:  2016-03-31       Impact factor: 1.553

4.  Dosimetric evaluation of newly developed well-type ionization chamber for use in the calibration of brachytherapy sources.

Authors:  Sathiyan Saminathan; Henry Finlay Godson; Retna Ponmalar; Ravikumar Manickam; James Mazarello
Journal:  J Med Phys       Date:  2016 Oct-Dec
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.