Florian Streitner1, Nina Schoene1, Martin Borggrefe1, Jürgen Kuschyk1.
Abstract
Wireless video capsule endoscopy (CE) is a modern diagnostic tool. Because of its use of digital radiofrequency, it is still relatively contraindicated in patients with implanted cardiac devices. We report the case of a patient with an Optimizer III system delivering cardiac contractility modulating signals (CCM) for heart failure therapy and an implantable cardioverter defibrillator (ICD) who underwent CE. No interferences between the devices were found.Entities:
Year: 2011 PMID: 24940511 PMCID: PMC4008400 DOI: 10.1155/2011/275659
Source DB: PubMed Journal: Case Rep Cardiol ISSN: 2090-6404
Device settings.
| Optimizer settings | |
|---|---|
| Initiation time | 00:00 |
| Termination time | 24:00 |
| CCM (hours/day) | 10 |
| CCM signal delay (ms) | 50 |
| CCM signal amplitude (V) | 6.0 |
| CCM signal total duration (ms) | 5.14 |
| Sensitivity RA Lead (mV) | 2 |
| Sensitivity RV Lead (mV) | 2 |
| Sensitivity LS Lead (mV) | 4.1 |
| AV delay (ms) | 252 |
|
| |
|
| |
|
| |
| Pacer programming (DDD/min) | 60–130 |
| VT detection zone (bpm) | >150 |
| VF detection zone (bpm) | >231 |
| PVARB (ms) | 310 |
| Ventricular refractory period (ms) | 250 |
| Sensitivity RA Lead (mV) | 0.2 (auto) |
| Sensitivity RV Lead (mV) | 0.3 (auto) |
| A/V post-sensed decay delay (ms) | 0/60 |
| A/V post-sensed threshold (%) | 50/62.5 |
CCM: cardiac contractility modulation; RA: right atrium; RV: right ventricle; LS: lower septal; AV: atrioventricular; VT: ventricular tachycardia; VF: ventricular fibrillation; PVARB: postventricular atrial refractory period.
Figure 1ICD Printout: CE, ICD, and CCM active.
Figure 2ICD Printout: CE and ICD active, CCM off.