Alemseged Abdissa1, Lubbe Wiesner2, Helen McIlleron2, Henrik Friis3, Ase Bengård Andersen4, Pernille Kaestel3.
Abstract
INTRODUCTION: Therapeutic drug monitoring (TDM) may improve antiretroviral efficacy through adjustment of individual drug administration. This could result in reduced toxicity, prevent drug resistance, and aid management of drug-drug interactions. However, most measurement methods are too costly to be implemented in resource-limited settings. This study evaluated a commercially available immunoassay for measurement of plasma efavirenz.Entities:
Keywords: LC–MS/MS; TDM; antiretroviral therapy; efavirenz; immunoassay; method evaluation
Mesh:
Substances:
Year: 2014 PMID: 24909561 PMCID: PMC4049128 DOI: 10.7448/IAS.17.1.18979
Source DB: PubMed Journal: J Int AIDS Soc ISSN: 1758-2652 Impact factor: 5.396
Figure 1Bland–Altman plots of the agreement between the immunoassay and the LC–MS/MS. The solid line is the mean difference between the two methods (Immunoassay-LC–MS/MS); the dashed lines are 95% limits of agreement (±2 SD).
Mean difference (Immunoassay – LC–MS/MS) between the efavirenz concentrations obtained by immunoassay and LC–MS/MS in different categories
| Category from LC–MS/MS | Number of samples | Mean difference | 95% CI of difference |
|---|---|---|---|
| Negative | 85 | 0.31 | −0.97; 1.59 |
| Suboptimal (<1 µg/mL) | 22 | 0.08 | −0.4; 0.56 |
| Normal (1–4 µg/mL) | 136 | −0.15 | −1.29; 0.99 |
| High (>4 µg/mL) | 72 | −0.72 | −4.15; 1.97 |
Cross tabulation between the immunoassay and the LC–MS/MS in categorization of the efavirenz concentration, n (%)
| Immunoassay | ||||
|---|---|---|---|---|
|
| ||||
| LC–MS/MS | Suboptimal (<1 µg/mL) | Normal (1–4 µg/mL) | High (>4 µg/mL) | All |
| Negative | 81 (95.3) | 2 (2.3) | 2 (2.3) | 85 (100) |
| Suboptimal (<1 µg/mL) | 20 (90.9) | 2 (9.1) | 0 (0.0) | 22 (100) |
| Normal (1–4 µg/mL) | 2 (1.5) | 130 (95.6) | 4 (2.9) | 136 (100) |
| High (>4 µg/mL) | 0 (0.0) | 23 (31.9) | 49 (68.1) | 72 (100) |
Between-run precision, coefficients of variation and accuracy (% of value provided by ARK diagnostics) of efavirenz concentrations measured by the immunoassay (n=8 runs)
| Actual value in µg/ml | Mean ±SD µg/ml | CV% | Accuracy | |
|---|---|---|---|---|
| Control low | 1 | 1.42±0.14 | 9.9 | 142 |
| Control medium | 4 | 4.61±0.34 | 7.4 | 115 |
| Control high | 8 | 8.31±0.50 | 6.0 | 104 |
| Efavirenz negative sample | 0 | 0.15±0.03 | 20.0 |