| Literature DB >> 23755102 |
Praveen Srivastava1, Ganesh S Moorthy, Robert Gross, Jeffrey S Barrett.
Abstract
A selective and a highly sensitive method for the determination of the non-nucleoside reverse transcriptase inhibitor (NNRTI), efavirenz, in human plasma has been developed and fully validated based on high performance liquid chromatography tandem mass spectrometry (LC-MS/MS). Sample preparation involved protein precipitation followed by one to one dilution with water. The analyte, efavirenz was separated by high performance liquid chromatography and detected with tandem mass spectrometry in negative ionization mode with multiple reaction monitoring. Efavirenz and ¹³C₆-efavirenz (Internal Standard), respectively, were detected via the following MRM transitions: m/z 314.20243.90 and m/z 320.20249.90. A gradient program was used to elute the analytes using 0.1% formic acid in water and 0.1% formic acid in acetonitrile as mobile phase solvents, at a flow-rate of 0.3 mL/min. The total run time was 5 min and the retention times for the internal standard (¹³C₆-efavirenz) and efavirenz was approximately 2.6 min. The calibration curves showed linearity (coefficient of regression, r>0.99) over the concentration range of 1.0-2,500 ng/mL. The intraday precision based on the standard deviation of replicates of lower limit of quantification (LLOQ) was 9.24% and for quality control (QC) samples ranged from 2.41% to 6.42% and with accuracy from 112% and 100-111% for LLOQ and QC samples. The inter day precision was 12.3% and 3.03-9.18% for LLOQ and quality controls samples, and the accuracy was 108% and 95.2-108% for LLOQ and QC samples. Stability studies showed that efavirenz was stable during the expected conditions for sample preparation and storage. The lower limit of quantification for efavirenz was 1 ng/mL. The analytical method showed excellent sensitivity, precision, and accuracy. This method is robust and is being successfully applied for therapeutic drug monitoring and pharmacokinetic studies in HIV-infected patients.Entities:
Mesh:
Substances:
Year: 2013 PMID: 23755102 PMCID: PMC3673986 DOI: 10.1371/journal.pone.0063305
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Figure 1Representative chromatograms.
A) Blank human plasma, B) Efavirenz in human plasma at Lowest limit of quantification (LLOQ, 1.0 ng/mL) and C) 13C6-Efavirenz (Internal standard, 10 ng/mL) in human plasma.
Figure 2Calibration Curve of Efavirenz in human plasma (1.0 ng/mL, r = 0.9984).
HPLC-MS method for the quantification of nine anti-HIV drugs from dry plasma spot on glass filter and their long term stability in different conditions.
| Accuracy and Precision for Calibration Standards of Efavirenz in human plasma (n = 8) | |||
| Nominal Concentration (ng/mL) | (Mean±SD) (ng/mL) | CV (%) | Accuracy (%) |
| 1 | 0.995±0.111 | 11.1 | 100 |
| 2.5 | 2.50±0.259 | 10.4 | 100 |
| 5 | 5.14±0.415 | 8.07 | 103 |
| 10 | 10.3±0.618 | 6.01 | 103 |
| 50 | 51.9±3.30 | 6.37 | 104 |
| 100 | 99.6±6.52 | 6.55 | 100 |
| 250 | 249±10.9 | 4.38 | 100 |
| 500 | 505±24.5 | 4.85 | 101 |
| 1000 | 974±71.5 | 7.34 | 97.4 |
| 2500 | 2402±181 | 7.52 | 96.1 |
Accuracy and precision for Efavirenz Calibration standards in human plasma.
Summary of Validation outcomes of Efavirenz in human plasma.
| Intra Day and Inter Day Accuracy and Precision for Efavirenz in human plasma | Nominal Conc. (ng/mL) | Recovery (n = 6) | |||||||
| Nominal Conc. (ng/mL) | Intra Day (n = 6) | Inter Day (n = 3 days) | |||||||
| (Mean±SD) (ng/mL) | CV (%) | Accuracy (%) | (Mean±SD) (ng/mL) | CV (%) | Accuracy (%) | Recov. (%) | CV (%) | ||
|
| 1.12±0.103 | 9.24 | 112 | 1.07±0.214 | 12.3 | 108 |
| 88.6 | 4.53 |
|
| 4.42±0.284 | 6.42 | 111 | 4.18±0.383 | 9.18 | 104 |
| 95.4 | 8.22 |
|
| 86.0±5.12 | 5.95 | 107 | 86.3±4.94 | 5.72 | 108 |
| 93.5 | 4.64 |
|
| 816±23.5 | 2.88 | 102 | 819±24.8 | 3.03 | 102 |
| 102 | 3.40 |
|
| 1993±48.0 | 2.41 | 100 | 1903±103 | 5.42 | 95.2 |
| 107 | 5.94 |
Summary of stability outcomes of Efavirenz in human plasma.
| Nominal Conc. (ng/mL) | Auto-sampler stability 72hr (n = 6) (Mean±SD) (ng/mL) | CV (%) | Accuracy (%) | Stability Bench Top 24hr at RT (n = 6) (Mean±SD) (ng/mL) | CV (%) | Accuracy (%) | Stability 3 cycle freeze (-80°C) thaw (n = 6) (Mean±SD) (ng/mL) | CV (%) | Accuracy(%) | Long term Stability Day 30 (n = 6) (Mean±SD) (ng/mL) | CV (%) | Accuracy (%) |
| 4 | 4.20±0.418 | 9.94 | 105 | 4.14±0.27 | 6.59 | 103 | 4.43±0.438 | 9.87 | 111 | 3.86±0.553 | 14.3 | 96.4 |
| 80 | 88.6±3.74 | 4.22 | 111 | |||||||||
| 800 | 828±45.8 | 5.54 | 103 | |||||||||
| 2000 | 1993±37.2 | 1.87 | 99.7 | 1857±75.3 | 4.05 | 92.8 | 1787±86.2 | 4.82 | 89.3 | 2058±95.6 | 4.64 | 103 |
Figure 3Representative chromatograms of Efavirenz in human plasma clinical study samples with calculated concentrations.
A) 20100511_4016 GNVM (244 ng/mL) and B) 20100709_1021 NKUX (2220 ng/mL).