| Literature DB >> 2488152 |
G W Brauer1, P D Fisher, J A Hanley, G W Ritchie.
Abstract
Signal detection performance was evaluated on the basis of ROC analysis using both digital and conventional images of a humanoid chest phantom. Simulated focal (coin) lesions were the target pathology. Digital images were acquired using a 57-cm image intensifier, digitized to 1024 x 1024 x 10 bits, and compared, in both video and laser-printed film formats, with conventional 14 x 17-inch chest films. Signal detection using digital video and laser printed images, of the same image polarity as conventional images, was found not to differ significantly from that achieved using conventional images, despite the smaller size of the digital images.Entities:
Mesh:
Year: 1989 PMID: 2488152 DOI: 10.1007/bf03168015
Source DB: PubMed Journal: J Digit Imaging ISSN: 0897-1889 Impact factor: 4.056