Lead externalization - a potential source of lead-related infective endocarditis?

Maciej Polewczyk1, Dorota Nowosielecka2, Konrad Tomków3, Łukasz Tułecki3, Andrzej Kutarski4, Anna Polewczyk1.   

Abstract

Entities:  

Year:  2018        PMID: 30302112      PMCID: PMC6173088          DOI: 10.5114/aic.2018.78340

Source DB:  PubMed          Journal:  Postepy Kardiol Interwencyjnej        ISSN: 1734-9338            Impact factor:   1.426


× No keyword cloud information.
A 32-year-old patient with a single chamber implantable cardioverter-defibrillator (ICD) was admitted to a cardiological ward due to recurrent episodes of fever since 1 month before. Blood tests showed a slightly elevated C-reactive protein (CRP) level without leukocytosis or procalcitonin increase. The transthoracic echocardiography (TTE) examination revealed a vegetation (29 mm × 6 mm) in the lower right atrium connected with the lead; the image was confirmed in transesophageal echocardiography (TEE). Negative blood culture was only probably negative due to the antibiotic therapy. The patient was diagnosed with lead-related infective endocarditis (LRIE), received modified antibiotic therapy and was initially qualified for the surgical extraction of the ICD system, but he did not agree to surgery. After consultation with the referential transvenous lead extraction (TLE) center the patient was qualified for TLE. Preoperative fluoroscopy revealed an interesting finding – ICD lead externalization was observed at the same location as the previously described vegetation. The TLE procedure was performed under general anesthesia in a hybrid operating room with intraoperative TEE monitoring and cardiac surgical standby. Due to the presence of a relatively large vegetation, a special nitinol basket was introduced into the main pulmonary artery for pulmonary circulation protection. The lead was removed using a Byrd dilator sheath (Cook, USA). Although during the procedure the vegetation was fragmented (only small parts were caught by the basket) there were no clinical signs of pulmonary embolism. There were no periprocedural complications. Postoperative examination of the extracted Riata (St. Jude Medical, USA) defibrillating lead showed significant cable externalization with remnants of the vegetation and relevant lead degeneration (Figure 1).
Figure 1

A – Transvenous lead extraction procedure with pulmonary vascular bed protection, B – TEE image showing a large vegetation, C – externalization of a lead

A – Transvenous lead extraction procedure with pulmonary vascular bed protection, B – TEE image showing a large vegetation, C – externalization of a lead Lead externalization is relatively common among Riata leads. However, in the literature we may find cases of such a phenomenon in different leads produced by different manufacturers [1]. The mechanism of externalization of Riata leads is related to their specific construction. Inside there are several conductors which may move relative to the insulation, leading to its further damage – ‘inside-out abrasion’. In some research there was no evident correlation between lead externalization and electrical failure [2], but a large Canadian registry confirmed the impact of externalization on lead failure [3]; in our case the electrical parameters of the ICD system remained stable. The influence of lead externalization on LRIE development has not been investigated yet; there are only a few similar reports [4]. An externalized cable might be thrombogenic. A thrombus formation around the location of externalization may become the origin of a further infection. According to the recent reports, some patients with vegetations even exceeding 4 cm in diameter underwent TLE, avoiding cardiac surgery. However, the decision of the extraction method was made individually considering the total vegetation volume, the degree of infiltration of heart structures and the real risk of severe pulmonary embolism [5]. In clinical practice the decision of removal of a functional lead with present externalization remains problematic. Apart from possible electrical failure one should have in mind the fact that exposed internal lead conductors might be the chief sites of serious infection.

Conflict of interest

The authors declare no conflict of interest.
  5 in total

1.  Canadian Registry of Implantable Electronic Device Outcomes: Surveillance of the Riata Lead Under Advisory.

Authors:  Ratika Parkash; Bernard Thibault; Iqwal Mangat; Benoit Coutu; Matthew Bennett; Jeffrey Healey; Atul Verma; Francois Philippon; Roopinder Sandhu; Douglas Cameron; Felix Ayala-Paredes; Laurence Sterns; Vidal Essebag; Teresa Kus; Pablo Nery; Elizabeth Stephenson; Raymond Yee; Derek Exner; Satish Toal; David Birnie; George Wells; Andrew Krahn
Journal:  Circ Arrhythm Electrophysiol       Date:  2016-10

2.  Mechanical and electrical dysfunction of Riata implantable cardioverter-defibrillator leads.

Authors:  Fatma Demirel; Ahmet Adiyaman; Peter Paul H M Delnoy; Jaap Jan J Smit; Anand R Ramdat Misier; Arif Elvan
Journal:  Europace       Date:  2014-05-19       Impact factor: 5.214

3.  Lead-related infective endocarditis: factors influencing the formation of large vegetations.

Authors:  Anna Polewczyk; Wojciech Jachec; Andrzej Tomaszewski; Wojciech Brzozowski; Marek Czajkowski; Aneta Maria Polewczyk; Marianna Janion; Andrzej Kutarski
Journal:  Europace       Date:  2017-06-01       Impact factor: 5.214

4.  Could Externalized St. Jude Medical Riata® Lead Be a Culture Medium of a Polymicrobial Endocarditis? A Clinical Case.

Authors:  Zefferino Palamà; Roberta Trotta; Cosimo Mandurino; Mariangela Pinto; Giovanni Luzzi; Stefano Favale
Journal:  Case Rep Cardiol       Date:  2017-01-15

Review 5.  Externalized conductors and insulation failure in Biotronik defibrillator leads: History repeating or a false alarm?

Authors:  Elia De Maria; Ambra Borghi; Lorenzo Bonetti; Pier Luigi Fontana; Stefano Cappelli
Journal:  World J Clin Cases       Date:  2017-02-16       Impact factor: 1.337