| Literature DB >> 24841248 |
Wei Zhang1, G Ruben H Regterschot2, Hana Schaabova3, Heribert Baldus4, Wiebren Zijlstra5.
Abstract
Chair rise performance is incorporated in clinical assessments to indicate fall risk status in older persons. This study investigated the test-retest reliability of a pendant-sensor-based assessment of chair rise performance. Forty-one older persons (28 females, 13 males, age: 72-94) were assessed in two sessions with 3 to 8 days in between. Repeated chair rise transfers were measured after different instructions. Relative and absolute test-retest reliability of chair rise measurements in individual tests and average over all tests were evaluated by means of intra-class correlation coefficients (ICCs) and standard error of measurement (SEM) as a percentage of the measurement mean. Systematic bias between the measurements in test and retest was examined with paired t-tests. Heteroscedasticity of the measurements was visually checked with Bland-Altman plots. In the different test conditions, the ICCs ranged between 0.63 and 0.93, and the SEM% ranged between 5.7% and 21.2%. The relative and absolute reliability of the average over all tests were ICC = 0.86 and SEM% = 9.5% for transfer duration, ICC = 0.93 and SEM% = 9.2% for maximum vertical acceleration, and ICC = 0.89 and SEM% = 10.0% for peak power. The results over all tests indicated that a fall risk assessment application based on pendant-worn-sensor measured chair rise performance in daily life might be feasible.Entities:
Mesh:
Year: 2014 PMID: 24841248 PMCID: PMC4063047 DOI: 10.3390/s140508705
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1.3D acceleration (color) and norm (black) of the acceleration of a chair rise transfer.
Test-retest reliability of the clinical assessments.
| TUG (sec) | 15.77 ± 9.36 | 15.27 ± 9.19 | −0.50 ± 1.68 | 0.08 | −1.05–0.05 | 0.99 | 5.4 |
| FTSS (sec) | 17.36 ± 4.87 | 16.29±4.68 | −1.07 ± 2.92 | 0.04 | −2.07–0.07 | 0.90 | 9.0 |
Figure 2.Bland-Altman plots of average measurements over all tests. The mean of differences between the test and the test scores is indicated by the dotted line in the middle of the plot. The upper and the lower boundaries of the 95% CI are indicated by the dotted lines above and below the mean. All measurement scores are dimensionless numbers.
Summary of evaluations of the relative and the absolute reliability in chair rise measurements in individual tests.
|
|
|
|
| |||||
|---|---|---|---|---|---|---|---|---|
|
|
|
|
| |||||
| Duration | 0.77 | 10.8 | 0.86 | 7.5 | 0.80 | 12.4 | 0.90 | 5.7 |
| Max Vert Acc | 0.86 | 13.0 | 0.82 | 11.5 | 0.91 | 10.1 | 0.93 | 7.2 |
| Peak power | 0.85 | 11.0 | 0.63 | 13.8 | 0.88 | 10.6 | 0.88 | 8.5 |
| Max Jerk | 0.78 | 20.7 | 0.66 | 21.2 | 0.85 | 15.2 | 0.86 | 12.6 |
Test-retest reliability analysis based on average scores of measurements over all tests. All measurement scores are dimensionless numbers.
| Duration | 4.88 ± 1.16 | 4.67 ± 1.27 | −0.21 ± 0.85 | 0.15 | −0.50–0.08 | 0.86 | 9.5 |
| Max Vert Acc | 0.37 ± 0.13 | 0.38 ± 0.13 | 0.01 ± 0.07 | 0.30 | −0.01–0.03 | 0.93 | 9.2 |
| Peak Power | 0.32 ± 0.10 | 0.35 ± 0.10 | 0.02 ± 0.06 | 0.03 | 0.00–0.05 | 0.89 | 10.0 |
| Maximum Jerk | 0.05 ± 0.02 | 0.05 ± 0.02 | 0.00 ± 0.01 | 0.43 | −0.00–0.01 | 0.86 | 14.8 |