The 24-hour pulse wave velocity, aortic augmentation index, and central blood pressure in normotensive volunteers.

Tatyana Y Kuznetsova1, Viktoria A Korneva1, Evgeniya N Bryantseva2, Vitaliy S Barkan2, Artemy V Orlov3, Igor N Posokhov4, Anatoly N Rogoza5.   

Abstract

The purpose of this study was to examine the pulse wave velocity, aortic augmentation index corrected for heart rate 75 (AIx@75), and central systolic and diastolic blood pressure during 24-hour monitoring in normotensive volunteers. Overall, 467 subjects (206 men and 261 women) were recruited in this study. Participants were excluded from the study if they were less than 19 years of age, had blood test abnormalities, had a body mass index greater than 2 7.5 kg/m(2), had impaired glucose tolerance, or had hypotension or hypertension. Ambulatory blood pressure monitoring (ABPM) with the BPLab(®) device was performed in each subject. ABPM waveforms were analyzed using the special automatic Vasotens(®) algorithm, which allows the calculation of pulse wave velocity, AIx@75, central systolic and diastolic blood pressure for "24-hour", "awake", and "asleep" periods. Circadian rhythms and sex differences in these indexes were identified. Pending further validation in prospective outcome-based studies, our data may be used as preliminary diagnostic values for the BPLab ABPM additional index in adult subjects.

Entities:  

Keywords:  ambulatory; arterial stiffness; augmentation index; central blood pressure; monitoring; reference values

Mesh:

Year:  2014        PMID: 24812515      PMCID: PMC4011902          DOI: 10.2147/VHRM.S61978

Source DB:  PubMed          Journal:  Vasc Health Risk Manag        ISSN: 1176-6344


Introduction

Since the start of the new millennium, registries to assess the impact of ambulatory blood pressure monitoring (ABPM) on hypertension management and related outcomes have been established in several countries. ABPM registries are more commonly focused on patients with documented or possible hypertension who would benefit from ABPM.1 The BPLab-Vasotens registry that was established this year is the project of dozens of small groups aimed at joint solutions for various scientific problems. One of the challenges is research into new ABPM vascular indexes. Some of these indexes, such as pulse wave velocity (PWV), are recommended for practical clinical use, whereas others, such as central blood pressure (BP) and augmentation index (AIx), are indexes of great interest for mechanistic analyses in pathophysiology, pharmacology, and therapeutics, but more investigation is needed before recommending their routine clinical use.2 It is also necessary to note that at the present time, there are devices that provide not only a single pulse wave measurement but several pulse wave measurements over a period of 24 hours.3,4 The purpose of this study is to examine the PWV, aortic AIx corrected for heart rate 75 (AIx@75), and central systolic and diastolic blood pressure (SBPao and DBPao, respectively) during 24-hour monitoring in normotensive volunteers.

Materials and methods

Baseline characteristics

Participants were screened for medical history and submitted to physical exams, blood tests (including lipid profile and glucose), and echocardiography with 12-lead electrocardiography. Individuals were excluded from the study if they were less than 19 years of age, had blood test abnormalities, had a body mass index greater than 27.5 kg/m2, had impaired glucose tolerance, or had hypotension or hypertension. Overall, 467 Caucasian subjects (66 men and 67 women under the age of 34, 75 men and 83 women aged 35–49, 65 men and 76 women aged 50–64, and 35 women aged 65 or over) were recruited in this study after providing written informed consent for participation. The study was approved by the local ethical committee.

Measurements

ABPM with the BPLab® device (Petr Telegin, Nizhny Novgorod, Russia) was performed on each subject. The accuracy of this device has previously been validated for ABPM.5 There have been other validations for vascular indexes.6–8 The principle of the oscillometric method is based on plethysmography and registers pulsatile pressure changes in the brachial artery. The recordings were made in ambulatory conditions using a conventional brachial BP cuff for adults. During the BP measurement the pressure waveforms in the cuff are registered, digitalized, and stored in the device while performing a step-by-step deflation. The ABPM waveforms were analyzed by a special automatic mathematical algorithm using BPLab Vasotens® technology (Petr Telegin), which allows calculation of central pulse wave parameters from the peripheral pulse wave. An aortic pulse wave is generated with a generalized transfer function that uses a modification of a certain frequency range within the acquired pulse signal to derive the aortic pressure wave. The modulus and phase characteristics of the Vasotens transfer function have previously been published.7 The principle of the PWV measurement of the Vasotens has also previously been described.4,6 The difference in time between the first wave and the second (ie, the reflected wave) is related to the distance measure according to the manufacturer’s instructions, and the resulting PWV is in m/second. PWV, AIx@75, SBPao and DBPao were calculated for “24-hour”, “awake”, and “asleep” periods. Quality control consisted of a visual assessment of the curves on the Vasotens clinical report screen. There was an average of 49 (standard deviation =8.3) successful readings during the monitoring period.

Statistics

BPStat® software version 05.01 (Petr Telegin) enabled the automatic tabulation of all indexes of every measured waveform used. Statistica 8.0 (StatSoft Inc., Tulsa, OK, USA) and MedCalc (MedCalc Software, Ostend, Belgium) software were used for the analysis. Mean and standard deviation were used in the descriptive statistics. The association between age, sex, and time of day (awake or asleep) with PWV, AIx@75, and central BP of the subjects was determined using a multiple regression analysis. The strength of the association assessed by the multiple regression analysis was indicated by the standardized regression coefficient value (Beta). We also used the percentile method to estimate the reference intervals.

Results

Table 1 describes the PWV, AIx@75, SBPao, and DBPao in the male and female study participants. Men demonstrated significantly (P<0.01) higher 24-hour PWV in the 50–64-year-old age group (9.3 m/second) compared with women in the same age group (8.3 m/second). Furthermore, the 24-hour PWV in men increased with age significantly (7.8 versus 9.32 m/second, P<0.01), whereas women did not display a significant increase with age (7.9 versus 8.3 m/second, P>0.05). Both men and women in all age groups displayed a significant nocturnal fall of PWV.
Table 1

Distribution of Vasotens®-calculated additional ABPM parameters in normotensive volunteers according to sex and age

Age in yearsMen
Women
24 hour
Awake
Asleep
24 hour
Awake
Asleep
M (SD)Me (5–95 pc)M (SD)Me (5–95 pc)M (SD)Me (5–95 pc)M (SD)Me (5–95 pc)M (SD)Me (5–95 pc)M (SD)Me (5–95 pc)
Pulse wave velocity, m/s
<347.8 (0.13)7.9 (7.7–8.2)8.5 (0.17)8.7 (8.3–9.0)7.1 (0.11)7.2 (6.9–7.4)7.9 (0.15)7.8 (7.5–8.1)8.4 (0.16)7.8 (7.6–8.2)6.7 (0.13)6.7 (6.5–6.7)
35–498.5 (0.14)8.6 (8.3–8.9)9.1 (0.21)9.2 (8.8–9.6)8.1 (0.15)8.2 (7.8–8.5)8.2 (0.13)8.3 (8.1–8.5)8.8 (0.16)8.3 (8.1–8.6)8.0 (0.12)8.0 (7.8–8.3)
50–649.3 (0.7)9.5 (8.1–10.8)9.5 (0.4)9.6 (8.8–10.5)8.6 (0.6)8.7 (7.4–10.0)8.3 (0.24)8.3 (7.8–8.8)8.9 (0.2)8.4 (7.9–8.8)8.2 (0.2)8.2 (7.9–8.6)
>658.3 (0.4)8.4 (7.6–9.2)8.9 (0.5)8.4 (7.6–9.1)8.1 (0.3)8.1 (7.5–8.7)
Aortic augmentation index corrected for heart rate 75, %
<349.4 (7.7)9 (−5.5–24.7)10.4 (8.1)10.5 (−5.7–26.7)7.8 (6.3)7.8 (−4.7–20.5)11.3 (3.7)11.4 (3.9–18.8)13.4 (4.8)12.4 (3.9–18.9)10.8 (3.2)10.8 (4.4–17.2)
35–4917.7 (4.4)17.8 (9.04–26.6)17.8 (6.2)17.9 (5.6–30.3)15.1 (3.1)15.2 (8.9–21.4)19.7 (5.2)19.7 (9.4–30.2)21.6 (6.4)20.9 (9.3–30.1)17.3 (5.4)17.3 (6.5–28.1)
50–6423.3 (12.0)23.5 (−0.5–47.5)25.2 (4.3)25.4 (16.8–33.9)22.9 (6.3)23.0 (10.4–35.6)27.0 (8.3)27.1 (10.5–43.7)27.6 (11.6)28.1 (10.5–44.7)26.2 (8.1)26.2 (10.2–42.4)
>6533.8 (12.5)33.9 (8.9–58.9)36.2 (15.7)34.9 (9.9–59.8)30.6 (11.3)30.6 (8.0–53.2)
Central systolic blood pressure, mmHg
<34107.5 (4.2)107 (99.2–116.0)112.5 (6.1)112.6 (100.4–124.8)97.3 (1.3)97.3 (94.7–99.9)103.1 (1.7)103.2 (99.8–106.8)109.3 (2.3)109.6 (105.0–114.2)94.5 (1.3)94.5 (91.8–97.1)
35–49104.7 (1.3)104.8 (102.4–107.2)109.3 (7.5)109.6 (94.4–124.5)95.3 (2.5)95.3 (90.3–100.4)103.5 (1.8)103.6 (99.9–107.2)110.6 (1.7)110.9 (107.5–114.3)97.3 (1.7)97.3 (93.9–100.7)
50–64114.5 (11.5)114.6 (91.6–137.7)118.2 (13.1)118.3 (92.2–144.5)106.1 (7.4)106.2 (91.4–121.0)112.5 (2.3)112.6 (108.0–117.2)116.9 (2.5)117.3 (112.3–122.3)101.2 (2.0)101.2 (97.2–105.2)
>65107.0 (7.1)107.1 (93.1–121.1)112.9 (9.3)113.3 (94.7–131.9)99.4 (4.8)99.4 (89.8–109.0)
Central diastolic blood pressure, mmHg
<3471.3 (1.3)71.4 (68.8–74.0)75.5 (1.6)75.6 (72.4–78.8)64.7 (1.5)64.7 (61.7–67.7)70.3 (1.2)70.4 (67.8–72.7)74.9 (1.5)75.2 (72.2–78.2)62.6 (1.4)62.6 (59.8–65.4)
35–4972.7 (1.3)72.8 (70.4–75.2)77.5 (3.8)77.6 (70.0–85.2)64.9 (2.1)64.8 (60.8–69.2)69.7 (1.9)69.8 (65.9–73.6)73.4 (1.7)73.6 (70.4–77.1)61.5 (1.7)61.5 (58.1–64.9)
50–6481.6 (10.2)81.1 (61.2–101.2)84.8 (12.3)84.6 (60.3–109.5)72.7 (6.4)72.8 (60.1–85.6)73.7 (2.3)73.9 (69.2–78.5)78.4 (2.4)78.8 (74.0–83.6)63.8 (2.7)63.82 (58.4–69.2)
>6572.5 (2.5)72.6 (67.6–77.6)77.6 (3.8)78 (70.4–85.6)61.8 (2.9)61.8 (56.0–67.6)

Abbreviations: 5 pc, the upper limit of the 5th percentile; 95 pc, the lower limit of the 95th percentile; ABPM, ambulatory blood pressure monitoring; M, mean; Me, median; SD, standard deviation.

Conversely, women displayed significantly (P<0.01) higher 24-hour AIx@75 in the 50–64-year-old age group (27.0%) compared with men of the same age (23.3%). It should be noted that the nocturnal shift of AIx@75 was not significant. Regarding the central aortic BP, both men and women displayed a characteristic nocturnal pattern similar to the “dipping” status of the peripheral pressure. There was an increase of 24-hour central aortic BP with age in men (systolic: 107.5 for age <34 years versus 114.5 for the 50–64-year-old age group, P<0.01; diastolic: 71.3 for ages <34-years-old versus 81.0 for ages 50–64, P<0.01) and in women (systolic: 103.1 for age <34-years-old versus 112.5 for 50–64-years-old, P<0.01; diastolic: 70.3 for age <34-years-old versus 73.7 for 50–64-years-old, P<0.01). Most of the studied indexes showed significant association with time of day after multiple adjustments for age and sex (Table 2).
Table 2

Multiple regression results for the relationship between pulse wave velocity, aortic augmentation index corrected for heart rate 75, and central systolic and diastolic blood pressure with variables studied

PWV
AIx@75
SBPao
DBPao
BetaP-levelBetaP-levelBetaP-levelBetaP-level
Time of day (awake or asleep)0.400.00030.110.07310.440.00040.420.0004
Age0.240.02030.190.04590.290.03090.230.0413
Sex0.230.04420.270.02730.200.04640.170.0623

Abbreviations: AIx@75, aortic augmentation index corrected for heart rate 75; Beta, standardized regression coefficient value; DBPao, central diastolic blood pressure; PWV, pulse wave velocity; SBPao, central systolic blood pressure.

Discussion

The main result of the present study is the establishment of normal and reference values for indexes of Vasotens technology, namely PWV, central BP, and AIx@75. These values are based on a dataset obtained from four centers distributed across Russia. From a practical viewpoint, the accurate determination of PWV cut-off value excesses is an important task. This task allows physicians to draw conclusions about the significant alterations of aortic function and, accordingly, to include a patient in one risk group or another. It is evident that the representation of PWV values according to age is similar to the representation of the data reported in other published reports.9–11 According to some studies, there is nocturnal fall in PWV.2 However, in our study, normal and reference values for the night PWV, according to sex and age, have been published for the first time. Interestingly, women in this study had a significantly lower PWV compared to men. It is well known that clinical and autopsy observations have suggested that women seem to develop vascular disease later than men, which could be related to factors specific to women that account, for example, for resistance to atheroma growth, etc.12,13 The circadian rhythms of central BP were also observed. Brachial and central BP both demonstrated typical diurnal patterns with lower pressures at night. Nighttime central BP may provide prognostic importance and warrants further investigation.14,15 Our study adds data to the above information on sex differences in central BP. The relationship of the AIx to sex has recently been reported.16 Sex-specific equations including age, heart rate, and height to calculate reference values for AIx have also been validated.17 The AIx@75 in our study is corrected for heart rate, but it still has the most pronounced sex differences. There were limitations associated with this study. The number of normotensive men over 65-years-old was too low, thus limiting our ability to detect significant differences in that age group. Studies involving a large number of patients are essential to generalize the results of the study. This study did not exclude smokers, and its findings may be slightly difficult to apply to general populations. In conclusion, pending further validation in prospective outcome-based studies, our data may be used as preliminary diagnostic values of BPLab ABPM additional index in adult subjects.
  16 in total

1.  European Society of Hypertension position paper on ambulatory blood pressure monitoring.

Authors:  Eoin O'Brien; Gianfranco Parati; George Stergiou; Roland Asmar; Laurie Beilin; Grzegorz Bilo; Denis Clement; Alejandro de la Sierra; Peter de Leeuw; Eamon Dolan; Robert Fagard; John Graves; Geoffrey A Head; Yutaka Imai; Kazuomi Kario; Empar Lurbe; Jean-Michel Mallion; Giuseppe Mancia; Thomas Mengden; Martin Myers; Gbenga Ogedegbe; Takayoshi Ohkubo; Stefano Omboni; Paolo Palatini; Josep Redon; Luis M Ruilope; Andrew Shennan; Jan A Staessen; Gert vanMontfrans; Paolo Verdecchia; Bernard Waeber; Jiguang Wang; Alberto Zanchetti; Yuqing Zhang
Journal:  J Hypertens       Date:  2013-09       Impact factor: 4.844

2.  Reference values for the augmentation index and pulse pressure in apparently healthy korean subjects.

Authors:  Jin Wook Chung; Young Soo Lee; Jeong Hyun Kim; Myung Jun Seong; So Yeon Kim; Jin Bae Lee; Jae Kean Ryu; Ji Yong Choi; Kee Sik Kim; Sung Gug Chang; Geon Ho Lee; Sung Hi Kim
Journal:  Korean Circ J       Date:  2010-04-22       Impact factor: 3.243

3.  Aortic augmentation index: reference values in a large unselected population by means of the SphygmoCor device.

Authors:  Julie H Janner; Nina S Godtfredsen; Steen Ladelund; Jørgen Vestbo; Eva Prescott
Journal:  Am J Hypertens       Date:  2009-12-03       Impact factor: 2.689

4.  Gender gap, inflammation and acute coronary disease: are women resistant to atheroma growth? Observations at autopsy.

Authors:  Richard J Frink
Journal:  J Invasive Cardiol       Date:  2009-06       Impact factor: 2.022

5.  Age- and gender-specific reference values of pulse wave velocity for African adults: preliminary results.

Authors:  Pedro Magalhães; Daniel P Capingana; Amílcar B T Silva; Albano V L Ferreira; Roberto de Sá Cunha; Sérgio L Rodrigues; José G Mill
Journal:  Age (Dordr)       Date:  2013-01-15

6.  Pulse wave velocity as marker of preclinical arterial disease: reference levels in a uruguayan population considering wave detection algorithms, path lengths, aging, and blood pressure.

Authors:  Ignacio Farro; Daniel Bia; Yanina Zócalo; Juan Torrado; Federico Farro; Lucía Florio; Alicia Olascoaga; Walter Alallón; Ricardo Lluberas; Ricardo L Armentano
Journal:  Int J Hypertens       Date:  2012-05-16       Impact factor: 2.420

7.  Effect of age on aortic atherosclerosis.

Authors:  Michael A Chen; Miwa Kawakubo; Patrick M Colletti; Dongxiang Xu; Laurie Labree Dustin; Robert Detrano; Stanley P Azen; Nathan D Wong; Xue-Qiao Zhao
Journal:  J Geriatr Cardiol       Date:  2013-06       Impact factor: 3.327

8.  Day-to-day repeatability of the Pulse Time Index of Norm.

Authors:  Igor N Posokhov; Aleksandra O Konradi; Eugeny V Shlyakhto; Oleg V Mamontov; Artemy V Orlov; Anatoly N Rogoza
Journal:  Med Devices (Auckl)       Date:  2014-02-27

9.  Novel description of the 24-hour circadian rhythms of brachial versus central aortic blood pressure and the impact of blood pressure treatment in a randomized controlled clinical trial: The Ambulatory Central Aortic Pressure (AmCAP) Study.

Authors:  Bryan Williams; Peter S Lacy; Fabio Baschiera; Patrick Brunel; Rainer Düsing
Journal:  Hypertension       Date:  2013-04-29       Impact factor: 10.190

10.  Validation of the BPLab(®) 24-hour blood pressure monitoring system according to the European standard BS EN 1060-4:2004 and British Hypertension Society protocol.

Authors:  Sergey A Koudryavtcev; Vyacheslav M Lazarev
Journal:  Med Devices (Auckl)       Date:  2011-10-26
View more
  9 in total

Review 1.  Twenty-Four-Hour Ambulatory Pulse Wave Analysis in Hypertension Management: Current Evidence and Perspectives.

Authors:  Stefano Omboni; Igor N Posokhov; Yulia V Kotovskaya; Athanase D Protogerou; Jacques Blacher
Journal:  Curr Hypertens Rep       Date:  2016-10       Impact factor: 5.369

Review 2.  Ambulatory arterial stiffness in chronic kidney disease: a methodological review.

Authors:  Andrea László; György Reusz; János Nemcsik
Journal:  Hypertens Res       Date:  2015-12-03       Impact factor: 3.872

3.  Ambulatory blood pressure and arterial stiffness web-based telemonitoring in patients at cardiovascular risk. First results of the VASOTENS (Vascular health ASsessment Of The hypertENSive patients) Registry.

Authors:  Stefano Omboni; Igor Posokhov; Gianfranco Parati; Anatoly Rogoza; Yulia Kotovskaya; Ayana Arystan; Alberto Avolio; Vitaliy Barkan; Natalia Bulanova; Ernesto Cardona Muñoz; Elena Grigoricheva; Alexandra Konradi; Irina Minyukhina; Maria Lorenza Muiesan; Giuseppe Mulè; Iana Orlova; Telmo Pereira; João Manuel Peixoto Maldonado; Mikhail E Statsenko; Ioan Tilea; Gabriel Waisman
Journal:  J Clin Hypertens (Greenwich)       Date:  2019-07-11       Impact factor: 3.738

4.  Evaluating daily profile of central aortic pressure and reflected pulse wave parameters in climacteric women.

Authors:  Sergey N Tolstov; Igor A Salov; Anton R Kiselev; Andrey P Rebrov
Journal:  BMC Cardiovasc Disord       Date:  2021-05-19       Impact factor: 2.298

5.  Comparison of the BPLab® sphygmomanometer for ambulatory blood pressure monitoring with mercury sphygmomanometry in pregnant women: validation study according to the British Hypertension Society protocol.

Authors:  Inna V Dorogova; Elena S Panina
Journal:  Vasc Health Risk Manag       Date:  2015-04-13

6.  Evaluation of 24-Hour Arterial Stiffness Indices and Central Hemodynamics in Healthy Normotensive Subjects versus Treated or Untreated Hypertensive Patients: A Feasibility Study.

Authors:  Stefano Omboni; Igor N Posokhov; Anatoly N Rogoza
Journal:  Int J Hypertens       Date:  2015-01-26       Impact factor: 2.420

7.  Validation of the BPLab(®) 24-hour blood pressure monitoring system in a pediatric population according to the 1993 British Hypertension Society protocol.

Authors:  Mikhail Y Ledyaev; Olga V Stepanova; Anastasia M Ledyaeva
Journal:  Med Devices (Auckl)       Date:  2015-02-02

8.  Vascular Health Assessment of The Hypertensive Patients (VASOTENS) Registry: Study Protocol of an International, Web-Based Telemonitoring Registry for Ambulatory Blood Pressure and Arterial Stiffness.

Authors:  Stefano Omboni; Igor N Posokhov; Gianfranco Parati; Alberto Avolio; Anatoly N Rogoza; Yulia V Kotovskaya; Giuseppe Mulè; Maria Lorenza Muiesan; Iana A Orlova; Elena A Grigoricheva; Ernesto Cardona Muñoz; Parounak H Zelveian; Telmo Pereira; João Manuel Peixoto Maldonado
Journal:  JMIR Res Protoc       Date:  2016-06-29

9.  Six-Week Exercise Training With Dietary Restriction Improves Central Hemodynamics Associated With Altered Gut Microbiota in Adolescents With Obesity.

Authors:  Junhao Huang; Jingwen Liao; Yang Fang; Hailin Deng; Honggang Yin; Bing Shen; Min Hu
Journal:  Front Endocrinol (Lausanne)       Date:  2020-12-07       Impact factor: 5.555

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.