Sergey A Koudryavtcev1, Vyacheslav M Lazarev.
Abstract
INTRODUCTION: Automatic blood pressure (BP) measuring devices are more and more often used in BP self-checks and in 24-hour BP monitoring. Nowadays, 24-hour BP monitoring is a necessary procedure in arterial hypertension treatment. The aim of this study was to validate the BPLab(®) ambulatory blood pressure monitor according to the European standard BS EN 1060-4:2004 and the British Hypertension Society (BHS) protocol, as well as to work out solutions regarding the suitability of using this device in clinical practice.Entities:
Keywords: 24-hour blood pressure monitoring; BPLab®; device; validation
Year: 2011 PMID: 22915946 PMCID: PMC3417890 DOI: 10.2147/MDER.S25800
Source DB: PubMed Journal: Med Devices (Auckl) ISSN: 1179-1470
Patients’ clinical profile parameters
| Parameter | Mean | SD | Min | Max |
|---|---|---|---|---|
| Age | 37 | 15 | 18 | 87 |
| SBP, mmHg | 134 | 25 | 64 | 198 |
| DBP, mmHg | 85 | 16 | 44 | 122 |
| Arm circumference, cm | 31 | 7 | 21 | 44 |
| Sex | Male | Female | ||
| 49 | 36 |
Abbreviations: SD, standard deviation; BP, blood pressure; SBP, systolic blood pressure; DBP, diastolic blood pressure.
Distribution of patients according to BP level
| SBP | Range, mmHg | <90 | 90–129 | 130–160 | 161–180 | >180 |
| No | 8 | 22 | 23 | 20 | 9 | |
| % | 9.4 | 25.9 | 27.1 | 23.5 | 10.6 | |
| DBP | Range, mmHg | <60 | 60–79 | 80–100 | 101–110 | >110 |
| No | 8 | 21 | 26 | 20 | 10 | |
| % | 9.4 | 24.7 | 30.6 | 23.5 | 11.8 |
Abbreviations: BP, blood pressure; SBP, systolic blood pressure; DBP, diastolic blood pressure.
Study results
| Difference between standard and test device (%) | Grade | AAMI criteria | |||||
|---|---|---|---|---|---|---|---|
|
|
| ||||||
| ≤5 mmHg | ≤10 mmHg | ≤15 mmHg | Mean difference, mmHg | SD of difference, mmHg | Result | ||
| SBP | 65 | 89 | 98 | A | 1.2 | 6.5 | Passed |
| DBP | 61 | 87 | 97 | A | −1.2 | 7.1 | Passed |
| SBP | 66 | 90 | 98 | A | 1.1 | 6.4 | Passed |
| DBP | 60 | 86 | 96 | A | −1.4 | 7.4 | Passed |
| SBP | 66 | 90 | 98 | A | 1.1 | 6.4 | Passed |
| DBP | 61 | 87 | 97 | A | −1.2 | 7.1 | Passed |
| SBP | 84 | 99 | 100 | 0.2 | 2.2 | ||
| DBP | 81 | 95 | 100 | −0.9 | 3.8 | ||
Abbreviations: SD, standard deviation; SBP, systolic blood pressure; DBP, diastolic blood pressure; AAMI, Association for the Advancement of Medical Instrumentation.
Figure 1Plot of pressure difference between the better observer and the test device, and mean pressure.
Abbreviations: SBP, systolic; DBP, diastolic blood pressure.