Validation of the BPLab(®) 24-hour blood pressure monitoring system according to the European standard BS EN 1060-4:2004 and British Hypertension Society protocol.

Sergey A Koudryavtcev1, Vyacheslav M Lazarev.   

Abstract

INTRODUCTION: Automatic blood pressure (BP) measuring devices are more and more often used in BP self-checks and in 24-hour BP monitoring. Nowadays, 24-hour BP monitoring is a necessary procedure in arterial hypertension treatment. The aim of this study was to validate the BPLab(®) ambulatory blood pressure monitor according to the European standard BS EN 1060-4:2004 and the British Hypertension Society (BHS) protocol, as well as to work out solutions regarding the suitability of using this device in clinical practice.
METHODS: A group of 85 patients of both sexes and different ages, who voluntarily agreed to take part in the tests and were given detailed instructions on the measurement technique were recruited for this study. The results of the BP measurement obtained by a qualified operator using the BPLab(®) device were compared with the BP values measured using the Korotkov auscultatory method. Data were obtained simultaneously by two experts with experience of over 10 years and had completed a noninvasive BP measurement standardization training course. Discrepancies in the systolic and diastolic BP measurements (N = 510; 255 for each expert) were analyzed according to the criteria specified in the BHS-93 protocol.
RESULTS: The device passed the requirements of the European Standard BS EN 1060-4:2004 and was graded 'A' according to the criteria of the BHS protocol for both systolic BP and diastolic BP.
CONCLUSION: The BPLab(®) 24-hour ambulatory blood pressure monitoring device may be recommended for extensive clinical use.

Entities:  

Keywords:  24-hour blood pressure monitoring; BPLab®; device; validation

Year:  2011        PMID: 22915946      PMCID: PMC3417890          DOI: 10.2147/MDER.S25800

Source DB:  PubMed          Journal:  Med Devices (Auckl)        ISSN: 1179-1470


Introduction

Automatic blood pressure (BP) measuring devices are more and more often used for BP self-checks and in 24-hour BP monitoring. According to the experts of European Society of Hypertension, 24-hour BP monitoring has been used for research purposes and at present it is mandatorily used in clinical and pharmacological studies and in evaluation of the efficiency of new antihypertensive medical products. Nowadays, 24-hour BP monitoring becomes an obligatory procedure in arterial hypertension detection. Considering the abovementioned, it is essential to objectively evaluate the accuracy of measurements performed by 24-hour BP monitoring devices offered in the market. In 1990, the British Hypertension Society published a protocol, the purpose of which was to standardize the validation of noninvasive BP measuring instruments.1 This protocol was revised in 1993 and has gained international status.2 In 2004, the British Hypertension Society together with European Society of Hypertension (ESH) published an international standard for the validation of noninvasive sphygmomanometers.3 The principles recommended in the documents provided the basis for work aimed at the validation of the 24-hour BP monitoring device, BPLab® (Petr Telegin Ltd, Nizhny Novgorod, Russia). The results of this validation are given below.

Methods

Subject selection

Subject selection was based on the requirements of the Standard EN 1060-4:2004.3 A group of 85 patients of both sexes and different ages, who voluntarily agreed to take part in the tests and were given detailed instructions on the measurement technique, were recruited to participate in this study. All participants provided verbal informed consent of study participation. The statistical distribution of the patients’ clinical profile parameters is given in Table 1, and their distribution according to the BP level in Table 2.
Table 1

Patients’ clinical profile parameters

ParameterMeanSDMinMax
Age37151887
SBP, mmHg1342564198
DBP, mmHg851644122
Arm circumference, cm3172144
SexMaleFemale
4936

Abbreviations: SD, standard deviation; BP, blood pressure; SBP, systolic blood pressure; DBP, diastolic blood pressure.

Table 2

Distribution of patients according to BP level

SBPRange, mmHg<9090–129130–160161–180>180
No82223209
%9.425.927.123.510.6
DBPRange, mmHg<6060–7980–100101–110>110
No821262010
%9.424.730.623.511.8

Abbreviations: BP, blood pressure; SBP, systolic blood pressure; DBP, diastolic blood pressure.

Exclusion criteria include: heart failure; significant rhythm disturbances or presence of an artificial pacemaker; pregnancy; and body mass index (BMI) >30 kg/m2.

Test measurement conditions

In accordance with the requirements of the BHS-93 protocol, the results of the BP measurement obtained by a qualified operator using the BPLab device were compared with the BP values measured using the Korotkov auscultatory method simultaneously by two expert medical specialists with over 10 years of experience, who had completed a noninvasive BP measurement standardization training course. The measurements were taken in the morning in comfortable settings (ambient temperature 22°C–25°C, no stimulatory sights or sounds, etc), after the patients had been relaxing in a seated position for 10 minutes. The patients were prohibited to take alcohol <24 hours beforehand or to smoke or take stimulants (tea, coffee, etc) <8 hours before the measurement; they were only allowed to have a light breakfast not less than half an hour before the test. Additionally, taking medication that has a strong effect on the cardiovascular system was not allowed on the day of the measurements. Expert BP measurements were performed by two experts simultaneously with usage of a high-quality auscultoscope with one head and two headbands, and two individual calibrated sphygmomanometers. The collar dimension depended on the patient’s shoulder girth.

Measurement schedule

The BP of each patient was measured nine times – alternating between the experts and the tested device – according to the following schedule: Measurement A: expert Measurement B: instrument Measurement 1: expert Measurement 2: instrument Measurement 3: expert Measurement 4: instrument Measurement 5: expert Measurement 6: instrument Measurement 7: expert The BP was measured on the nondominant arm. The interval between measurements was 30–60 seconds. The BP values obtained by the experts were recorded by each of them according to their own protocol. Three measurement error values were calculated for each patient (for systolic BP [SBP] and diastolic BP [DBP], separately) according to the following formulae: where BPn is a BP measurement result corresponding to the measurement number n. Measurement A was used to refer a patient to this or that group according to their BP level; measurement B was treated as a calibration measurement. Discrepancies in the systolic and diastolic BP measurements (N = 510; 255 for each expert) were subsequently analyzed according to the criteria specified in the BHS-93 protocol.

Results

Detailed results of the data analysis are given in Table 3 in the form recommended by the BHS-93 protocol.2 The number of measurement error values which are within the ranges envisaged by the protocol is indicated in the form of a percentage ratio in relation to the total number of measurements.
Table 3

Study results

Difference between standard and test device (%)GradeAAMI criteria


≤5 mmHg≤10 mmHg≤15 mmHgMean difference, mmHgSD of difference, mmHgResult
Observer 1
SBP658998A1.26.5Passed
DBP618797A−1.27.1Passed
Observer 2
SBP669098A1.16.4Passed
DBP608696A−1.47.4Passed
Final grading
SBP669098A1.16.4Passed
DBP618797A−1.27.1Passed
Observer comparison
SBP84991000.22.2
DBP8195100−0.93.8

Abbreviations: SD, standard deviation; SBP, systolic blood pressure; DBP, diastolic blood pressure; AAMI, Association for the Advancement of Medical Instrumentation.

Figure 1 is a graphical illustration of discrepancies of each measurement depending on the relevant BP value (separately for SBP and DBP). As recommended in the BHS-93 protocol, the figures show only the results of comparison with the expert who obtained closer results (for SBP – observer 2, for DBP – observer 1).
Figure 1

Plot of pressure difference between the better observer and the test device, and mean pressure.

Abbreviations: SBP, systolic; DBP, diastolic blood pressure.

Discussion

Recently, the accuracy of conventional sphygmomanometers for the measurement of BP has been called into question, and much effort has been made to improve the quality of automatic device measurements.5 Ambulatory BP monitoring is being increasingly recommended for routine clinical practice because it provides valued prognostic data.6,7 The accuracy of BP measuring devices is of great importance and should be validated before devices are put into operation. The aim of this study was to validate the accuracy of the BPLab® 24-hour ambulatory BP monitoring device (Petr Telegin Ltd, Nizhny Novgorod, Russia). The results of this study suggest that the BPLab® 24-hour ambulatory BP monitoring device passes the test carried out in a mixed group of patients in accordance with the requirements of the European Standard BS EN 1060-4:2004 and the BHS-93 protocol, and may be categorized in the ‘A’ accuracy class according to the abovementioned protocol. The tested device may be recommended for extensive clinical use.
  5 in total

1.  Replacing the mercury sphygmomanometer. Requires clinicians to demand better automated devices.

Authors:  E O'Brien
Journal:  BMJ       Date:  2000-03-25

2.  Training and assessment of observers for blood pressure measurement in hypertension research.

Authors:  E O'Brien; F Mee; K S Tan; N Atkins; K O'Malley
Journal:  J Hum Hypertens       Date:  1991-02       Impact factor: 3.012

3.  The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems.

Authors:  E O'Brien; J Petrie; W Littler; M de Swiet; P L Padfield; K O'Malley; M Jamieson; D Altman; M Bland; N Atkins
Journal:  J Hypertens       Date:  1990-07       Impact factor: 4.844

4.  Call to action on use and reimbursement for home blood pressure monitoring: executive summary: a joint scientific statement from the American Heart Association, American Society Of Hypertension, and Preventive Cardiovascular Nurses Association.

Authors:  Thomas G Pickering; Nancy Houston Miller; Gbenga Ogedegbe; Lawrence R Krakoff; Nancy T Artinian; David Goff
Journal:  Hypertension       Date:  2008-05-22       Impact factor: 10.190

5.  Ambulatory or home measurement of blood pressure?

Authors:  Philippe Gosse; Paul Coulon
Journal:  J Clin Hypertens (Greenwich)       Date:  2009-04       Impact factor: 3.738

  5 in total
  13 in total

1.  Tendency for age-specific mortality with hypertension in the European Union from 1980 to 2011.

Authors:  Lichan Tao; Cunying Pu; Shutong Shen; Hongyi Fang; Xiuzhi Wang; Qinkao Xuan; Junjie Xiao; Xinli Li
Journal:  Int J Clin Exp Med       Date:  2015-02-15

2.  Relationships between 24-h blood pressure variability and 24-h central arterial pressure, pulse wave velocity and augmentation index in hypertensive patients.

Authors:  Stefano Omboni; Igor N Posokhov; Anatoly N Rogoza
Journal:  Hypertens Res       Date:  2016-11-24       Impact factor: 3.872

Review 3.  Twenty-Four-Hour Ambulatory Pulse Wave Analysis in Hypertension Management: Current Evidence and Perspectives.

Authors:  Stefano Omboni; Igor N Posokhov; Yulia V Kotovskaya; Athanase D Protogerou; Jacques Blacher
Journal:  Curr Hypertens Rep       Date:  2016-10       Impact factor: 5.369

4.  Ambulatory blood pressure and arterial stiffness web-based telemonitoring in patients at cardiovascular risk. First results of the VASOTENS (Vascular health ASsessment Of The hypertENSive patients) Registry.

Authors:  Stefano Omboni; Igor Posokhov; Gianfranco Parati; Anatoly Rogoza; Yulia Kotovskaya; Ayana Arystan; Alberto Avolio; Vitaliy Barkan; Natalia Bulanova; Ernesto Cardona Muñoz; Elena Grigoricheva; Alexandra Konradi; Irina Minyukhina; Maria Lorenza Muiesan; Giuseppe Mulè; Iana Orlova; Telmo Pereira; João Manuel Peixoto Maldonado; Mikhail E Statsenko; Ioan Tilea; Gabriel Waisman
Journal:  J Clin Hypertens (Greenwich)       Date:  2019-07-11       Impact factor: 3.738

5.  Comparison of the BPLab® sphygmomanometer for ambulatory blood pressure monitoring with mercury sphygmomanometry in pregnant women: validation study according to the British Hypertension Society protocol.

Authors:  Inna V Dorogova; Elena S Panina
Journal:  Vasc Health Risk Manag       Date:  2015-04-13

6.  The "Pulse Time Index of Norm" highly correlates with the left ventricular mass index in patients with arterial hypertension.

Authors:  Igor N Posokhov; Natalya N Kulikova; Irina V Starchenkova; Elena A Grigoricheva; Vitaly V Evdokimov; Artemy V Orlov; Anatoly N Rogoza
Journal:  Vasc Health Risk Manag       Date:  2014-03-19

7.  Day-to-day repeatability of the Pulse Time Index of Norm.

Authors:  Igor N Posokhov; Aleksandra O Konradi; Eugeny V Shlyakhto; Oleg V Mamontov; Artemy V Orlov; Anatoly N Rogoza
Journal:  Med Devices (Auckl)       Date:  2014-02-27

8.  Evaluation of 24-Hour Arterial Stiffness Indices and Central Hemodynamics in Healthy Normotensive Subjects versus Treated or Untreated Hypertensive Patients: A Feasibility Study.

Authors:  Stefano Omboni; Igor N Posokhov; Anatoly N Rogoza
Journal:  Int J Hypertens       Date:  2015-01-26       Impact factor: 2.420

9.  Validation of the integration of technology that measures additional "vascular" indices into an ambulatory blood pressure monitoring system.

Authors:  Yulia V Kotovskaya; Zhanna D Kobalava; Artemy V Orlov
Journal:  Med Devices (Auckl)       Date:  2014-05-02

10.  The 24-hour pulse wave velocity, aortic augmentation index, and central blood pressure in normotensive volunteers.

Authors:  Tatyana Y Kuznetsova; Viktoria A Korneva; Evgeniya N Bryantseva; Vitaliy S Barkan; Artemy V Orlov; Igor N Posokhov; Anatoly N Rogoza
Journal:  Vasc Health Risk Manag       Date:  2014-04-28
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.