Hsiang-Chi Kuo1, Keyur J Mehta1, Linda Hong1, Ravindra Yaparpalvi1, Leslie L Montgomery2, William Bodner1, Wolfgang A Tomé1, Shalom Kalnicki1.
Abstract
PURPOSE: To evaluate the variations of multi-lumen balloon (MLB)-based brachytherapy from simulation day to treatment day and their dosimetric impacts during accelerated partial breast irradiation (APBI).Entities:
Keywords: APBI; Contura; brachytherapy; breast cancer; multi-lumen balloon; treatment uncertainty
Year: 2014 PMID: 24790624 PMCID: PMC4003432 DOI: 10.5114/jcb.2014.42025
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
Summary of the initial anatomic parameters from the reference CT image set and its corresponding plan evaluated parameters (group A) for each patient in this study. Values within the parenthesis were plan parameters for group B
| Patient | dAP1 | dLat1 | Vballoon2 | VPTV_Eval2 | Vseroma2 | dskin1 | drib1 | V953 | Dskin3 | Drib3 |
|---|---|---|---|---|---|---|---|---|---|---|
| 1 | 51 | 53 | 69.4 | 120.2 | < 0.1 | 14 | 24 | 100 (90.9) | 80.7 (74.9) | 55.8 (47.6) |
| 2 | 35 | 40 | 27.3 | 75.6 | < 0.1 | 14.6 | 16.4 | 99.5 (90.1) | 75.5 (68) | 61.4 (57.1) |
| 3 | 38 | 43 | 31.1 | 79.7 | 0.1 | 33 | 15 | 97.9 (90.5) | 29.0 (27.4) | 63.1 (39.8) |
| 4 | 42 | 55 | 48.3 | 101.5 | < 0.1 | 22 | 4 | 95.0 (90.6) | 55.4 (52.2) | 116.3 (108.1) |
| 5 | 42 | 41 | 37.3 | 87.2 | < 0.1 | 13 | 38 | 97.7 (90.3) | 97.6 (88.5) | 29.7 (27.8) |
| 6 | 45 | 50 | 55.1 | 107.1 | < 0.1 | 12 | 5 | 95.1 (90.0) | 101.4 (95.0) | 121.9 (117.8) |
| 7 | 33 | 36 | 24 | 65 | 1.1 | 21 | 4 | 92.4 (90.1) | 61.6 (61.1) | 122.4 (117.3) |
| Mean | 40.9 | 45.4 | 41.8 | 90.9 | 0.2 | 18.5 | 15.2 | 96.8 | 71.6 | 81.5 |
Unit: 1 – mm; 2 – cm3; 3 –%.
Fig. 1Scatterplot of the variation of dskin due to INTR and the corresponding Dskin, where “% PD” was the percentage of the prescription dose and the larger symbols were the intial values taken from reference CT
Summary of internal variations due to the change of balloon and anatomy, geometrical variations due to the change of catheter length, size of systematic error, and their dosimetric impact on PTV_Eval
| ΔdAP1 | ΔdLat1 | ΔVballoon2 | ΔVseroma2 | Δdskin1 | Δdrib1 | ΔV903 | ΔV953 | ΔV1502 | ΔV2002 | RSQL3 | |
|---|---|---|---|---|---|---|---|---|---|---|---|
| –0.9 ± 2.1 | –0.2 ± 1.1 | –0.4 ± 1.6 | 0.2 ± 0.9 | –1.4 ± 9.6 | 3.1 ± 7.0 | –0.3 ± 2 | 0.1 ± 2 | 0.8 ± 2.7 | 0.7 ± 2 | ||
|
| –0.3 ± 2 | 0.0 ± 2 | 0.8 ± 2.7 | 0.7 ± 2 | |||||||
| –1.6 ± 2 | –1.2 ± 2 | 0.8 ± 2.7 | 1.0 ± 2.8 | 2.7 ± 1.6 | |||||||
|
| –1.0 ± 2 | –1.3 ± 2 | 0.7 ± 2.8 | 1.0 ± 3 | |||||||
| GEOS | S–4mm | S–3mm | S–2mm | S–1mm | S1mm | S2mm | S3mm | S4mm | |||
| –4.0 ± 8 | –2.1 ± 6 | –0.7 ± 4 | –0.0 ± 2 | –0.7 ± 2 | –1.9 ± 3 | –3.6 ± 5 | –5.6 ± 7 | ||||
|
| –5.6 ± 13 | –3.4 ± 11 | –1.6 ± 8 | –0.5 ± 4 | –0.9 ± 4 | –2.6 ± 7 | –4.9 ± 7 | –7.7 ± 8 | |||
| –5.2 ± 11 | –2.8 ± 9 | –1.1 ± 7 | 0.0 ± 4 | –0.8 ± 3 | –2.4 ± 5 | –4.4 ± 7 | 6.7 ± 9 | ||||
|
| –4.4 ± 13 | –2.1 ± 11 | –0.5 ± 8 | 0.5 ± 5 | 0.0 ± 4 | –1.6 ± 7 | –4.0 ± 8 | –6.8 ± 8 |
Unit: 1, mm; 2, cm3; 3,%. Value displays as Mean±2SD.
Group A
Group B
Fig. 2Frequency distribution of the change of 50% dose of sugical clips due to INTR and GEOR, in terms of the percentage of priscription dose
Fig. 3Boxplots of the dosimetric impacts (ΔV90, ΔV95, ΔV150, ΔV200) vs. different type of variations, where INT is INTR, GEO is GEOR, S represents GEOS and is followed with different size of systematic errors