Evaluation of template matching for tumor motion management with cine-MR images in lung cancer patients.

Xiutao Shi1, Tejan Diwanji1, Karen E Mooney1, Jolinta Lin1, Steven Feigenberg1, Warren D D'Souza1, Nilesh N Mistry1.   

Abstract

PURPOSE: Accurate determination of tumor position is crucial for successful application of motion compensated radiotherapy in lung cancer patients. This study tested the performance of an automated template matching algorithm in tracking the tumor position on cine-MR images by examining the tracking error and further comparing the tracking error to the interoperator variability of three human reviewers.
METHODS: Cine-MR images of 12 lung cancer patients were analyzed. Tumor positions were determined both automatically with template matching and manually by a radiation oncologist and two additional reviewers trained by the radiation oncologist. Performance of the automated template matching was compared against the ground truth established by the radiation oncologist. Additionally, the tracking error of template matching, defined as the difference in the tumor positions determined with template matching and the ground truth, was investigated and compared to the interoperator variability for all patients in the anterior-posterior (AP) and superior-inferior (SI) directions, respectively.
RESULTS: The median tracking error for ten out of the 12 patients studied in both the AP and SI directions was less than 1 pixel (= 1.95 mm). Furthermore, the median tracking error for seven patients in the AP direction and nine patients in the SI direction was less than half a pixel (= 0.975 mm). The median tracking error was positively correlated with the tumor motion magnitude in both the AP (R = 0.55, p = 0.06) and SI (R = 0.67, p = 0.02) directions. Also, a strong correlation was observed between tracking error and interoperator variability (y = 0.26 + 1.25x, R = 0.84, p < 0.001) with the latter larger.
CONCLUSIONS: Results from this study indicate that the performance of template matching is comparable with or better than that of manual tumor localization. This study serves as preliminary investigations towards developing online motion tracking techniques for hybrid MRI-Linac systems. Accuracy of template matching makes it a suitable candidate to replace the labor intensive manual tumor localization for obtaining the ground truth when testing other motion management techniques.

Entities:  

Mesh:

Year:  2014        PMID: 24784397      PMCID: PMC4008763          DOI: 10.1118/1.4870978

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  30 in total

1.  Prediction of respiratory tumour motion for real-time image-guided radiotherapy.

Authors:  Gregory C Sharp; Steve B Jiang; Shinichi Shimizu; Hiroki Shirato
Journal:  Phys Med Biol       Date:  2004-02-07       Impact factor: 3.609

2.  Evaluation of internal lung motion for respiratory-gated radiotherapy using MRI: Part I--correlating internal lung motion with skin fiducial motion.

Authors:  Nicholas Koch; H Helen Liu; George Starkschall; Marc Jacobson; Kenneth Forster; Zhongxing Liao; Ritsuko Komaki; Craig W Stevens
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-12-01       Impact factor: 7.038

3.  Real-time intra-fraction-motion tracking using the treatment couch: a feasibility study.

Authors:  Warren D D'Souza; Shahid A Naqvi; Cedric X Yu
Journal:  Phys Med Biol       Date:  2005-08-11       Impact factor: 3.609

4.  Towards fluoroscopic respiratory gating for lung tumours without radiopaque markers.

Authors:  Ross I Berbeco; Hassan Mostafavi; Gregory C Sharp; Steve B Jiang
Journal:  Phys Med Biol       Date:  2005-09-13       Impact factor: 3.609

5.  Correlation of lung tumor motion with external surrogate indicators of respiration.

Authors:  Jeremy D P Hoisak; Katharina E Sixel; Romeo Tirona; Patrick C F Cheung; Jean-Philippe Pignol
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-11-15       Impact factor: 7.038

6.  Residual motion of lung tumours in gated radiotherapy with external respiratory surrogates.

Authors:  Ross I Berbeco; Seiko Nishioka; Hiroki Shirato; George T Y Chen; Steve B Jiang
Journal:  Phys Med Biol       Date:  2005-07-28       Impact factor: 3.609

7.  Feasibility of synchronization of real-time tumor-tracking radiotherapy and intensity-modulated radiotherapy from viewpoint of excessive dose from fluoroscopy.

Authors:  Hiroki Shirato; Masataka Oita; Katsuhisa Fujita; Yoshiharu Watanabe; Kazuo Miyasaka
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-09-01       Impact factor: 7.038

8.  Prediction of lung tumour position based on spirometry and on abdominal displacement: accuracy and reproducibility.

Authors:  Jeremy D P Hoisak; Katharina E Sixel; Romeo Tirona; Patrick C F Cheung; Jean-Philippe Pignol
Journal:  Radiother Oncol       Date:  2006-03-14       Impact factor: 6.280

9.  Real-time implementation of a dual-mode ultrasound array system: in vivo results.

Authors:  Andrew J Casper; Dalong Liu; John R Ballard; Emad S Ebbini
Journal:  IEEE Trans Biomed Eng       Date:  2013-05-21       Impact factor: 4.538

10.  Feasibility of insertion/implantation of 2.0-mm-diameter gold internal fiducial markers for precise setup and real-time tumor tracking in radiotherapy.

Authors:  Hiroki Shirato; Toshiyuki Harada; Tooru Harabayashi; Kazutoshi Hida; Hideho Endo; Kei Kitamura; Rikiya Onimaru; Koichi Yamazaki; Nobuaki Kurauchi; Tadashi Shimizu; Nobuo Shinohara; Michiaki Matsushita; Hirotoshi Dosaka-Akita; Kazuo Miyasaka
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-05-01       Impact factor: 7.038

View more
  6 in total

1.  High-quality t2-weighted 4-dimensional magnetic resonance imaging for radiation therapy applications.

Authors:  Dongsu Du; Shelton D Caruthers; Carri Glide-Hurst; Daniel A Low; H Harold Li; Sasa Mutic; Yanle Hu
Journal:  Int J Radiat Oncol Biol Phys       Date:  2015-03-30       Impact factor: 7.038

2.  Initial clinical observations of intra- and interfractional motion variation in MR-guided lung SBRT.

Authors:  David H Thomas; Anand Santhanam; Amar U Kishan; Minsong Cao; James Lamb; Yugang Min; Dylan O'Connell; Yingli Yang; Nzhde Agazaryan; Percy Lee; Daniel Low
Journal:  Br J Radiol       Date:  2018-01-22       Impact factor: 3.039

3.  Four-Dimensional Magnetic Resonance Imaging With 3-Dimensional Radial Sampling and Self-Gating-Based K-Space Sorting: Early Clinical Experience on Pancreatic Cancer Patients.

Authors:  Wensha Yang; Zhaoyang Fan; Richard Tuli; Zixin Deng; Jianing Pang; Ashley Wachsman; Robert Reznik; Howard Sandler; Debiao Li; Benedick A Fraass
Journal:  Int J Radiat Oncol Biol Phys       Date:  2015-08-21       Impact factor: 7.038

Review 4.  Image-guided radiotherapy and motion management in lung cancer.

Authors:  S S Korreman
Journal:  Br J Radiol       Date:  2015-05-08       Impact factor: 3.039

5.  The impact of 2D cine MR imaging parameters on automated tumor and organ localization for MR-guided real-time adaptive radiotherapy.

Authors:  Martin J Menten; Martin F Fast; Andreas Wetscherek; Christopher M Rank; Marc Kachelrieß; David J Collins; Simeon Nill; Uwe Oelfke
Journal:  Phys Med Biol       Date:  2018-11-22       Impact factor: 3.609

6.  Tumour auto-contouring on 2d cine MRI for locally advanced lung cancer: A comparative study.

Authors:  Martin F Fast; Björn Eiben; Martin J Menten; Andreas Wetscherek; David J Hawkes; Jamie R McClelland; Uwe Oelfke
Journal:  Radiother Oncol       Date:  2017-10-10       Impact factor: 6.280