New device for creating a continuous curvilinear capsulorhexis.

Matthew A Powers1, Malik Y Kahook2.   

Abstract

PURPOSE: To describe the evolution of a new device to facilitate continuous curvilinear capsulorhexis (CCC) creation.
SETTING: Department of Ophthalmology, University of Colorado School of Medicine, Aurora, Colorado, USA.
DESIGN: Experimental study.
METHODS: Bench-side ex vivo testing of unique prototypes for guidance and assistance of CCC in bovine and human eyes was performed. Five designs were sequentially tested as follows: a flexible circular blade of nickel-titanium alloy (nitinol), a flexible nitinol guide wire, a flexible elastomeric suction device, a combination approach of a nitinol guide wire and flexible silicone ring, and a freestanding micropatterned silicone ring.
RESULTS: The first 3 designs were not amenable to insertion through a sub-2.4 mm corneal incision and failed to maintain adequate downward force to cut the capsule and/or prevent radial tears. The fourth design was successfully inserted through a 2.4 mm incision and maintained adequate downward pressure and contact to guide a manual CCC without radial tears. The final design was insertable through a 2.4 mm incision and exhibited self-adhesive characteristics after placement on the anterior capsule of an ophthalmic viscosurgical device-filled anterior chamber.
CONCLUSIONS: Given the steep learning curve of manual capsulorhexis and the high cost of capsulotomy-assistive devices, such as the femtosecond laser, an alternative approach for creating a CCC is desirable. Performance of a highly precise manual CCC through a small incision using a medical-grade silicone device with an adhesive micropatterned design is a viable and cost-effective option for use in cataract surgery across a wide range of user experience. FINANCIAL DISCLOSURE: All authors are named as the inventors in a patent filed by the University of Colorado covering the details in this report.
Copyright © 2014 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24767915     DOI: 10.1016/j.jcrs.2013.10.041

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

1.  Defining the ideal femtosecond laser capsulotomy.

Authors:  Mark Packer; E Valas Teuma; Adrian Glasser; Steven Bott
Journal:  Br J Ophthalmol       Date:  2015-03-31       Impact factor: 4.638

Review 2.  Techniques of anterior capsulotomy in cataract surgery.

Authors:  Bhavana Sharma; Robin G Abell; Tarun Arora; Tom Antony; Rasik B Vajpayee
Journal:  Indian J Ophthalmol       Date:  2019-04       Impact factor: 1.848

3.  Real-Time Extraction of Important Surgical Phases in Cataract Surgery Videos.

Authors:  Shoji Morita; Hitoshi Tabuchi; Hiroki Masumoto; Tomofusa Yamauchi; Naotake Kamiura
Journal:  Sci Rep       Date:  2019-11-12       Impact factor: 4.379

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.