Mark Packer1, E Valas Teuma2, Adrian Glasser3, Steven Bott2.
Abstract
PURPOSE: We define the ideal anterior capsulotomy through consideration of capsular histology and biomechanics. Desirable qualities include preventing posterior capsular opacification (PCO), maintaining effective lens position (ELP) and optimising capsular strength.Entities:
Keywords: Treatment Lasers
Mesh:
Year: 2015 PMID: 25829488 PMCID: PMC4518749 DOI: 10.1136/bjophthalmol-2014-306065
Source DB: PubMed Journal: Br J Ophthalmol ISSN: 0007-1161 Impact factor: 4.638
Figure 1Diagram of the capsule after Fincham, 1937,40 scaled to an equatorial diameter of 9.8 mm with one half of the capsule divided into 200 divisions in 10 division steps.25 The thickest part of the capsule, in accordance with the results from Barraquer, is between divisions 40 and 50 and occurs at a diameter of approximately 5 mm or greater.
Input parameters for the laser capsulotomy treatment
| Test # | 1 | 2 | 3 | 4 | 5 |
|---|---|---|---|---|---|
| Capsulotomy technique | CCC | Laser 1 | Laser 2 | Laser 3 | Laser 4 |
| Number of samples | 14 | 14 | 7 | 7 | 7 |
| Capsulotomy diameter (mm) | 5 | 5.0 | 5.5 | 5.5 | 4.0 |
| Shot spacing X-Y (µm) | N/A | 0.5 | 2 | 2 | 2 |
| Z-spacing (µm) | N/A | 25 | 25 | 20 | 20 |
CCC, continuous curvilinear capsulorhexis.
Figure 2Testing apparatus.
Figure 3Breakdown of the capsulotomy elongation (El): First, the stretcher pins are inserted into the capsular bag and positioned apart with a distance necessary for the pins to enter into contact with the capsular rim (bottom diagram). This position is considered as the test initial position. Then, the test starts with the moving probe elongating the edge until it ruptures. During the elongation of the capsular rim, the capsular edge changes shape, from circular (initial shape) to oblong shape (Xo) and then stretches until rupture (Xs, top image.)
Figure 4Break force and maximum extension for laser capsulotomies of diameters 4.0, 5.0 and 5.5 mm. The larger diameter capsulotomies exhibit higher average break forces and maximum extensibilities. Error bars are one SD.