| Literature DB >> 24724036 |
Subhendra N Sarkar1, Pooja R Sarkar2, Efstathios Papavassiliou3, Rafael R Rojas1.
Abstract
Introduction. In medically refractory Parkinson's disease (PD) deep-brain stimulation (DBS) is an effective therapeutic tool. Postimplantation MRI is important in assessing tissue damage and DBS lead placement accuracy. We wanted to identify which MRI sequence can detect DBS leads with smallest artifactual signal void, allowing better tissue/electrode edge conspicuity. Methods. Using an IRB approved protocol 8 advanced PD patients were imaged within MR conditional safety guidelines at low RF power (SAR ≤ 0.1 W/kg) in coronal plane at 1.5T by various sequences. The image slices were subjectively evaluated for diagnostic quality and the lead contact diameters were compared to identify a sequence least affected by metallic leads. Results and Discussion. Spin echo and fast spin echo based low SAR sequences provided acceptable image quality with comparable image blooming (enlargement) of stimulator leads. The mean lead diameters were 2.2 ± 0.1 mm for 2D, 2.1 ± 0.1 mm for 3D, and 4.0 ± 0.2 mm for 3D MPRAGE sequence. Conclusion. Low RF power spin echo and fast spin echo based 2D and 3D FSE sequences provide acceptable image quality adjacent to DBS leads. The smallest artifactual blooming of stimulator leads is present on 3D FSE while the largest signal void appears in the 3D MPRAGE sequence.Entities:
Year: 2014 PMID: 24724036 PMCID: PMC3956280 DOI: 10.1155/2014/508576
Source DB: PubMed Journal: Parkinsons Dis ISSN: 2042-0080
Figure 1Typical coronal MR images, (a) a section from 3D MPRAGE; measured contact diameter (measurement plan is shown in panel (c) by straight arrows) = 4 mm, more than that from (b) a section from 2D SE T1 (2.2 mm) or (c) 2D FSE PD (2.3 mm) and (d) 2D FSE T2 (2.3 mm) sequences. The location of the lead at left subthalamic nucleus is indirectly estimated as 11-12 mm lateral to midline across the superior-anterior border of the red nucleus (curved arrow, panel (d)).
The range of lead contact measurements from various MR sequences.
| Lead tip-size measurements |
|
| D2D FSE T2 (mm) | D3D FSE T2 (mm) |
|---|---|---|---|---|
| 1 | 3.8 | 2.1 | 2.3 | 2.2 |
| 2 | 3.9 | 2.3 | 2.1 | 2.2 |
| 3 | 4.1 | 2 | 2.2 | 2.1 |
| 4 | 4.2 | 2.2 | 2 | 2.1 |
| 5 | 3.9 | 2.2 | 2.1 | 2 |
| 6 | 3.8 | 2.2 | 2.1 | 2 |
| 7 | 3.7 | 2.1 | 2.2 | 2 |
| 8 | 4.3 | 2.2 | 2.3 | 2.1 |
| Mean diameter (mm) | 4 ± 0.2 | 2.2 ± 0.1 | 2.2 ± 0.1 | 2.1 ± 0.1 |
Figure 2(a) Coronal 3D long echo train T2 image showing 2.2 mm tip diameter (ETL 80, TE 85 ms) with visualization of white matter tissue adequately; (b) and (c) 3D long echo train FLAIR MR signal for an infected DBS lead (arrow showing infection prior to removal (b) and after removal and treatment (c)). Note the artifactual bright dots in T2 images are not prominently present on FLAIR while the size of the implant lead is similar to those in T2 images (Figure 1).
Wilcoxon signed-rank test results for visualized DBS lead diameters from images using various MR sequences for DBS patients (C.I. α = 0.05).
| Test number | Null hypotheses |
| Test results |
|---|---|---|---|
| I. |
| 0, 3, | Reject |
| 4.0 ± 0.2; 2.2 ± 0.1 | |||
|
| |||
| II. |
| 0, 3, | Reject |
| 4.0 ± 0.2; 2.1 ± 0.1 | |||
|
| |||
| III. |
| 0, 3, | Reject |
| 4.0 ± 0.2; 2.2 ± 0.1 | |||
|
| |||
| IV. |
| 17, 3, | Accept |
| 2.1 ± 0.1; 2.2 ± 0.1 | |||
|
| |||
| V. |
| 8, 3, | Accept |
| 2.1 ± 0.1; 2.2 ± 0.1 | |||
|
| |||
| VI. |
| 8, 3, | Accept |
| 2.1 ± 0.1; 2.2 ± 0.1 | |||